← Product Code [KWQ](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWQ) · K252611

# aprevo® cervical plating system (K252611)

_Carlsmed, Inc. · KWQ · Dec 22, 2025 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWQ/K252611

## Device Facts

- **Applicant:** Carlsmed, Inc.
- **Product Code:** [KWQ](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWQ.md)
- **Decision Date:** Dec 22, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3060
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The aprevo® cervical plating system is intended for anterior cervical fixation (C2-T1) for the following indications: • Degenerative disc disease (DDD) (defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), • Spondylolisthesis, • Trauma (i.e., fracture or dislocation), • Spinal stenosis, • Deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis), • Tumor, • Pseudoarthrosis, and • Failed previous fusion.

## Device Story

The aprevo® cervical plating system provides anterior cervical spine fixation (C2-T1). The system consists of segmental and multilevel plates and fixation screws. Plates are additively manufactured from titanium alloy (Ti-4Al-6V ELI); screws are manufactured from titanium alloy (Ti-4Al-6V ELI). Associated instruments are stainless steel. The device is provided sterile for single-patient use. It is used by surgeons in a clinical/OR setting to stabilize the cervical spine. The surgeon implants the system to facilitate fusion and provide structural support. The device benefits patients by addressing spinal instability or deformity, potentially reducing pain and restoring alignment.

## Clinical Evidence

No clinical data. Substantial equivalence is based on non-clinical bench testing, including static and dynamic compression bending and static torsion testing per ASTM F1717, and cadaveric evaluation.

## Technological Characteristics

System includes segmental and multilevel plates (Ti-4Al-6V ELI per ASTM F3001) and fixation screws (Ti-4Al-6V ELI per ASTM F136). Instruments are stainless steel (ASTM A564). Additively manufactured. Sterile, single-use. No software or electronic components.

## Regulatory Identification

A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.

## Predicate Devices

- Admiral ACP System ([K212139](/device/K212139.md))
- Quantum Anterior Cervical Plate ([K210424](/device/K210424.md))
- Struxxure MCS Anterior Cervical Plate System ([K240690](/device/K240690.md))

## Reference Devices

- aprevo® Cervical Interbody System ([K250827](/device/K250827.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

December 22, 2025

Carlsmed, Inc.
Jesse Albright
Sr. Manager, Regulatory Affairs
1800 Aston Ave Ste 100
Carlsbad, California 92130

Re: K252611
Trade/Device Name: aprevo® cervical plating system
Regulation Number: 21 CFR 888.3060
Regulation Name: Spinal Intervertebral Body Fixation Orthosis
Regulatory Class: Class II
Product Code: KWQ
Dated: December 9, 2025
Received: December 9, 2025

Dear Jesse Albright:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

U.S. Food &amp; Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K252611 - Jesse Albright
Page 2

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

{2}

K252611 - Jesse Albright
Page 3

Sincerely,

MAZIAR SHAH-MOHAMMADI-S

[For] Brent Showalter, Ph.D.
Assistant Director
DHT6B: Division of Spinal Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

|  Indications for Use  |   |   |
| --- | --- | --- |
|  Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K252611 | ?  |
|  Please provide the device trade name(s). |   | ?  |
|  aprevo® cervical plating system  |   |   |
|  Please provide your Indications for Use below. |   | ?  |
|  The aprevo® cervical plating system is intended for anterior cervical fixation (C2-T1) for the following indications:
• Degenerative disc disease (DDD) (defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies),
• Spondylolisthesis,
• Trauma (i.e., fracture or dislocation),
• Spinal stenosis,
• Deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis),
• Tumor,
• Pseudoarthrosis, and
• Failed previous fusion.  |   |   |
|  Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ?  |

{4}

K252611- Page 1 of 2

510(k) Summary

Contact Details
Applicant: Carlsmed, Inc.
Address: 1800 Aston Ave Ste 100
Carlsbad, CA 92008
Phone number: (760) 766-1923
Contact person: Jesse Albright
Sr. Manager, Regulatory Affairs
jalbright@carlsmed.com
Date prepared: August 18, 2025

Device Name
Trade name: aprevo® cervical plating system
Common name: Appliance, Fixation, Spinal Intervertebral Body
Classification name: Spinal Intervertebral Body Fixation Orthosis (21 CFR 888.3060)
Class: II
Product code: KWQ

Legally Marketed Predicate/Reference Devices
|  510(k) Number | Product Code | Trade Name | Manufacturer  |
| --- | --- | --- | --- |
|  Primary Predicate Device  |   |   |   |
|  K212139 | KWQ | Admiral ACP System | SeaSpine Orthopedics Corporation  |
|  Additional Predicate Device(s)  |   |   |   |
|  K210424 | KWQ | Quantum Anterior Cervical Plate | Nvision Biomedical Technologies, Inc.  |
|  K240690 | KWQ | Struxxure MCS Anterior Cervical Plate System | Nexxt Spine, LLC  |
|  Reference Device  |   |   |   |
|  K250827 | ODP, OVE | aprevo® Cervical Interbody System | Carlsmed, Inc.  |

Device Description
The aprevo® cervical plating system, which is comprised of the aprevo® cervical segmental plating system and the aprevo® cervical multilevel plating system configurations, is intended for anterior

{5}

K252611- Page 2 of 2

fixation of the cervical spine. The system consists of a variety of segmental and multilevel plates that are additively manufactured from titanium alloy (Ti-4Al-6V ELI) per ASTM F3001 as well as a range of fixation screws manufactured from titanium alloy (Ti-4Al-6V ELI) per ASTM F136. The associated instruments, which facilitate the placement, adjustment, and removal, if necessary, of the implants, are manufactured from stainless steel per ASTM A564. The plates, screws, and instruments are provided sterile packaged for single patient use.

## Indications for Use

The aprevo® cervical plating system is intended for anterior cervical fixation (C2-T1) for the following indications:

- Degenerative disc disease (DDD) (defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies),
- Spondylolisthesis,
- Trauma (i.e., fracture or dislocation),
- Spinal stenosis,
- Deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis),
- Tumor,
- Pseudoarthrosis, and
- Failed previous fusion.

## Summary of Technological Characteristics

The aprevo® cervical plating system technological characteristics are substantially equivalent to the cited predicate devices. The equivalence determination was based on comparison of intended use/indications for use, operating principle, design, components, materials, biocompatibility, manufacturing, packaging, labeling, sterility, and non-clinical testing.

## Non-Clinical Testing

The aprevo® cervical plating system demonstrated substantially equivalent mechanical performance to the cited predicate devices through static and dynamic compression bending and static torsion per ASTM F1717. Cadaveric evaluation demonstrated that the subject device performs as intended.

## Clinical Testing

Not applicable. The determination of substantial equivalence is not based on an assessment of clinical performance data.

## Conclusion

The submitted data demonstrates that the subject aprevo® cervical plating system is substantially equivalent to the cited legally marketed predicate devices.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWQ/K252611](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KWQ/K252611)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
