K111911 · Corin USA · KWL · Aug 4, 2011 · Orthopedic
Device Facts
Record ID
K111911
Device Name
CORIN OPTIMOM MODULAR HEAD
Applicant
Corin USA
Product Code
KWL · Orthopedic
Decision Date
Aug 4, 2011
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3360
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Corin Optimom Modular Head is indicated for hemi-arthroplasty in cemented and uncemented primary or revision femoral stem applications whose indications include procedures for patients suffering pain and disability due to osteoarthritis, rheumatoid arthritis, avascular necrosis of the femoral head, femoral neck fracture and abnormalities where the major pathology affects the femoral head, the acetabular cavity is normal and acetabular replacement is either undesirable or not required. The device is intended for prescription use only.
Device Story
Corin Optimom Modular Head is a polished, truncated sphere femoral head prosthesis; functions as a hemi-arthroplasty component. Device features internal female taper modified for compatibility with 12/14 male taper femoral hip stems. Used in orthopedic surgery for patients with femoral head pathology and normal acetabulum. Implanted by surgeons; provides articulation for hip joint. Benefits include pain relief and restoration of function for patients with arthritis or femoral neck fractures. Substantial equivalence established via design, material, and mechanical testing comparisons to predicate modular heads.
Clinical Evidence
No clinical data. Substantial equivalence determined via non-clinical bench testing including fatigue, rotational torque, axial pull-off, and fretting corrosion analysis.
Technological Characteristics
Polished, truncated sphere femoral head; Cobalt-Chrome alloy (ASTM F75); diameters 40-56mm; internal female taper for 12/14 male taper compatibility. Mechanical device; no software or electronics.
Indications for Use
Indicated for hemi-arthroplasty in patients requiring primary or revision femoral stem procedures due to osteoarthritis, rheumatoid arthritis, avascular necrosis of the femoral head, or femoral neck fracture, where the acetabulum is normal and acetabular replacement is not required.
Regulatory Classification
Identification
A hip joint femoral (hemi-hip) metallic cemented or uncemented prosthesis is a device intended to be implanted to replace a portion of the hip joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum. This generic type of device includes designs which are intended to be fixed to the bone with bone cement (§ 888.3027) as well as designs which have large window-like holes in the stem of the device and which are intended for use without bone cement. However, in these latter designs, fixation of the device is not achieved by means of bone ingrowth.
K153725 — Corin TaperFit Hip Stem · Corin U.S.A. Limited · Mar 25, 2016
K143184 — Evolve UniPolar Head · Signature Orthopaedics Pty, Ltd. · Mar 17, 2015
K063791 — CORIN UNIPOLAR MODULAR HEAD · Corin USA · Apr 3, 2007
K142761 — Corin TaperFit Hip Stem · Corin USA · Apr 3, 2015
K962646 — PRIME MODULAR ENDO HEAD · Orthopaedic Innovations, Inc. · Sep 23, 1996
Submission Summary (Full Text)
{0}------------------------------------------------
K||||9||#1/2
AUG - 4 2011
| 3. 510(K) SUMMARY | |
|--------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------|
| 1. Applicant/Sponsor: | Corin USA<br>10500 University Center Drive<br>Suite 190<br>Tampa, Florida 33612<br>Establishment Registration No.: 1056629 |
| 2. Contact Person: | Lucinda Gerber<br>Regulatory Affairs Associate<br>Corin USA<br>813-977-4469<br>lucinda.gerber@coringroup.com |
| 3. Proprietary Name: | Corin Optimom Modular Head |
| 4. Common Name: | Optimom Modular Head |
| 5. Classification Name: | 21CFR 888.3360 - Hip joint femoral (hemi-hip) metal<br>cemented or uncemented prosthesis |
| 6. Legally Marketed Devices to which Substantial Equivalence is claimed: | |
| | • Corin Unipolar Modular Head (K063791) |
| | • Corin Taperfit Total Hip System (K003666) |
| | • Corin Tri-Fit Femoral Stem (K010243) |
- (K110087)
# 7. Device Description:
The Corin Optimom Modular Head is a polished, truncated sphere with a high tolerance internal female taper. It is a device modification of the internal female taper of the Corin Unipolar Modular Heads (K063791) to provide for compatibility with an additional range of Corin femoral hip stem designs with a 12/14 male taper. The materials and dimensions of the internal female taper are substantially equivalent to the CoCr Modular Heads cleared in K003666 and K010243 and the Trinity Modular Heads cleared in K110087. The Optimom Modular Head is manufactured from Cobalt-Chrome alloy conforming to
{1}------------------------------------------------
ASTM F75 and is available in diameters ranging from 40-56mm with a variety of offsets.
- 8. Intended Use / Indications:
The Corin Optimom Modular Head is indicated for hemi-arthroplasty in cemented and uncemented primary or revision femoral stem applications whose indications include procedures for patients suffering pain and disability due to osteoarthritis, rheumatoid arthritis, avascular necrosis of the femoral head, femoral neck fracture and abnormalities where the major pathology affects the femoral head, the acetabular cavity is normal and acetabular replacement is either undesirable or not required.
The device is intended for prescription use only.
- 9. Summary of Technologies/Substantial Equivalence:
The Optimom Modular Heads have the same intended use and indications as the Corin Unipolar Modular Heads (K063791). The range of diameters available for the Optimom Modular Heads is within the range cleared for the predicate Unipolar Modular Heads and the device design is similar to the Unipolar Heads with the only modification of the internal female taper. The Optimom Modular Heads are manufactured from the same material as the predicates Unipolar Modular Heads, CoCr Modular Heads (K003666, K010243) and Trinity Modular Heads (K110087). The design and dimensions of the internal female taper of the Optimom Modular Heads are the same as the internal female taper of the predicates CoCr Modular Heads and Trinity CoCr Modular heads. Based upon these similarities, the Optimom Modular Heads are believed to be substantially equivalent to the predicate devices.
### 10. Non-Clinical Testing:
Non-clinical testing conducted to demonstrate substantial equivalence includes a comparison of the designs, materials, intended use, indications and dimensions of the Corin Optimom Modular Head to the predicate devices Fatigue, rotational torque, axial pull off, and fretting corrosion were conducted.
- 11. Clinical Testing:
Clinical testing was not necessary to determine substantial equivalence between the Corin Optimom Modular Heads and the predicate devices.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health and Human Services (HHS) in the United States. The logo features a stylized graphic of an eagle or bird-like figure with three curved lines representing its wings or feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the graphic.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring. MID 20993-0002
Corin USA % Ms. Lucinda Gerber Regulatory Affairs Associate 10500 University Center Drive. Suite 190 Tampa. Florida 33612
AUG - 4 2011
Re: K111911 Trade/Device Name: Corin Optimom Modular Head Regulation Number: 21 CFR 888.3360 Regulation Name: Hip joint femoral (hemi-hip) metallic cemented or uncemented prosthesis Regulatory Class: Class II Product Code: KWL Dated: July 01, 2011 Received: July 05, 2011
Dear Ms. Gerber:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{3}------------------------------------------------
## Page 2 - Ms. Lucinda Gerber
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vour
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# 2. INDICATIONS FOR USE
510(k) Number (if known): _K1119 | |
Device Name: Corin Optimom Modular Head
Indications for Use:
The Corin Optimom Modular Head is indicated for hemi-arthroplasty in cemented and uncemented primary or revision femoral stem applications whose indications include procedures for patients suffering pain and disability due to osteoarthritis, theumatoid arthritis, avascular necrosis of the femoral head, femoral neck fracture and abnormalities where the major pathology affects the femoral head, the acetabular cavity is normal and acetabular replacement is either undesirable or not required.
The device is intended for prescription use only.
Prescription Use Over-The-Counter Use X - AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Igor M. Melkersen
<//19/1
Page 1 of 1
(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.