Indications commonly found in the interphalangeal joints; semi-rigid or rigid hammertoe deformity; angular deformity; impaired function and stability; pain; impaired toe length ratio. Implants are single use only.
Device Story
Stemmed flexible implant for replacement of lesser toe interphalangeal joints; constructed of medical grade silicone elastomer. System includes rods (2.0mm-2.5mm diameters), drills, and sizers. Used by surgeons in clinical settings to stabilize joints and correct deformities. Implant provides structural support to restore toe function and alignment; reduces pain. Single-use device.
Clinical Evidence
No clinical data provided; substantial equivalence based on design and material similarity to predicate.
Technological Characteristics
Medical grade silicone elastomer; stemmed flexible rod design; diameters 2.0mm to 2.5mm; includes manual surgical instruments (drills, sizers).
Indications for Use
Indicated for patients with interphalangeal joint conditions including semi-rigid or rigid hammertoe deformity, angular deformity, impaired function/stability, pain, and impaired toe length ratio.
Regulatory Classification
Identification
A toe joint polymer constrained prosthesis is a device made of silicone elastomer or polyester reinforced silicone elastomer intended to be implanted to replace the first metatarsophalangeal (big toe) joint. This generic type of device consists of a single flexible across-the-joint component that prevents dislocation in more than one anatomic plane.
Predicate Devices
Sgarlato Laboratories Shaw-Ship Rod, Hammer Toe Implant
Related Devices
K022886 — METATARSOPHALANGEAL FLEXIBLE STABILIZING ROD SYSTEM · Osteomed Corp. · Nov 25, 2002
K173491 — RTS Lesser MTP Implant System · In2bones USA, LLC · Feb 23, 2018
K190136 — Fusion Silastic System · Fusion Orthopedics, LLC · Oct 21, 2019
K102601 — ORTHOFLEX ROD · Ortho-Pro, LLC · Jan 14, 2011
K153065 — The Hammertoe Correction System · Life Spine, Inc. · Aug 1, 2016
Submission Summary (Full Text)
{0}------------------------------------------------
NOV 2 6 2002
K022887
page lotl
## 510(k) Summary
| Device Proprietary Name: | OsteoMed Interphalangeal<br>Flexible Stabilizing Rod System |
|--------------------------|-------------------------------------------------------------------------------------------------------------------|
| Device Common Name: | Toe Implants, Silicone |
| Classification Name: | Prosthesis, Toe Joint,<br>Constrained, Polymer |
| Name of Submitter: | OsteoMed Corporation<br>3750 Realty Road<br>Addison, Texas 75001<br>Phone: (972) 241-3401<br>Fax: (972) 241-3449 |
| Contact Person: | Dawn T. Holdeman |
| Date Prepared: | August 23, 2002 |
Summary:
This submission describes the OsteoMed Interphalangeal Flexible Stabilizing Rod System intended for indications commonly found in the interphalangeal joints; semi-rigid or rigid hammertoe deformity; anqular deformity; impaired function and stability; pain; impaired toe length ratio. OsteoMed Interphalangeal Flexible Stabilizing Rod Implants are intended for single use only.
The OsteoMed Interphalangeal Flexible Stabilizing Rod System is a stemmed flexible implant specifically designed for replacement of the interphalangeal joints of the lesser toes. It is constructed of medical grade silicone elastomer. The OsteoMed Interphalangeal Flexible Stabilizing Rods are offered in diameters of 2.0mm through 2.5mm. Drills and sizers will also be a part of the system.
Equivalence for this device is based on similarities in intended use, material, design and operational principle to the Sgarlato Laboratories Shaw-Ship Rod, Hammer Toe Implant.
Due to the similarity of materials and design to both pre-enactment and postenactment devices, OsteoMed believes that the OsteoMed Interphalangeal Flexible Stabilizing Rod System does not raise any new safety or effectiveness issues.
Image /page/0/Picture/11 description: The image shows the logo for Osteomed Corporation. The text "OSTEOMED" is in large, bold, sans-serif font. Below that, in a smaller font, is the word "CORPORATION". The logo is black and white.
{1}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird symbol, composed of three curved lines, representing the department's mission. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV 2 6 2002
Ms. Dawn T. Holdeman Regulatory Affairs / Document Control OsteoMed Corporation 3750 Realty Road Addison, Texas 75001-4311
Re: K022887
Trade/Device Name: Interphalangeal Flexible Stabilizing Rod System Regulation Numbers: 21 CFR 888.3720 Regulation Names: Toe joint polymer constrained prosthesis Regulatory Class: II Product Code: KWH Dated: August 29, 2002 Received: August 30, 2002
Dear Ms. Holdeman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{2}------------------------------------------------
Page 2 – Ms. Dawn T. Holdeman
forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely vours,
Mark A. Mulberson
Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## OsteoMed "Indications for Use" Submission
| 510(k) Number: | K022887 |
|---------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name: | OsteoMed Interphalangeal Flexible Stabilizing Rod System |
| Indication for Use: | Indications commonly found in the interphalangeal joints; semi-rigid or rigid hammertoe deformity; angular deformity; impaired function and stability; pain; impaired toe length ratio. Implants are single use only. |
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 810.109) Over-The Counter-Use (Optical Format 1-)
for Mark A. Milkersm
(Division
ﺍﻟﻤﺴﺎﻋﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘ Cestor Tive 11. 4. 1. and in a
K022887
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.