The subject devices have the same intended use as those specified in the 510(k) submissions for the predicate devices listed. In general, these devices are intended for use in primary or revision knee arthroplasty.
Device Story
Stryker UHMWPE knee patella components and hinge knee bumpers; previously cleared devices; subject to packaging modification only. Devices packaged in nitrogen vacuum (N2Vac) environment using impermeable foil pouch within PETG blister sealed with Tyvek lid. Used by orthopedic surgeons in clinical/OR settings during primary or revision knee arthroplasty. Output is the physical implant component; provides structural replacement/stabilization for knee joint. Benefits include restoration of joint function and stability in patients with degenerative or traumatic knee conditions.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing only. Testing included material property evaluation (small punch, oxygen content, oxidation index), packaging integrity (ship testing, seal strength, bubble leak), accelerated aging for 5-year shelf-life validation, and biocompatibility (endotoxin per AAMI ST72, cytotoxicity per ISO 10993-5).
Indicated for patients with painful, disabling knee joint disease (degenerative, rheumatoid, or post-traumatic arthritis, avascular necrosis); post-traumatic loss of joint configuration/function; moderate varus, valgus, or flexion deformity requiring ligamentous stabilization; revision of unsuccessful knee replacement; or fracture of distal femur/proximal tibia not manageable by standard techniques. Triathlon Hinge Bumper is indicated for joint surface destruction, ligamentous instability, or revision of failed prosthesis with gross instability/bone loss, including use with MRH or GMRS components.
Regulatory Classification
Identification
A knee joint femorotibial metal/polymer constrained cemented prosthesis is a device intended to be implanted to replace part of a knee joint. The device limits translation or rotation in one or more planes and has components that are linked together or affined. This generic type of device includes prostheses composed of a ball-and-socket joint located between a stemmed femoral and a stemmed tibial component and a runner and track joint between each pair of femoral and tibial condyles. The ball-and-socket joint is composed of a ball at the head of a column rising from the stemmed tibial component. The ball, the column, the tibial plateau, and the stem for fixation of the tibial component are made of an alloy, such as cobalt-chromium-molybdenum. The ball of the tibial component is held within the socket of the femoral component by the femoral component's flat outer surface. The flat outer surface of the tibial component abuts both a reciprocal flat surface within the cavity of the femoral component and flanges on the femoral component designed to prevent distal displacement. The stem of the femoral component is made of an alloy, such as cobalt-chromium-molybdenum, but the socket of the component is made of ultra-high molecular weight polyethylene. The femoral component has metallic runners which align with the ultra-high molecular weight polyethylene tracks that press-fit into the metallic tibial component. The generic class also includes devices whose upper and lower components are linked with a solid bolt passing through a journal bearing of greater radius, permitting some rotation in the transverse plane, a minimal arc of abduction/adduction. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
K061882 — OXYGENLESS PACKAGING CONVERSION OF LEGACY CENTERPULSE STANDARD POLYETHYLENE DEVICES · Zimmer, Inc. · Aug 1, 2006
K213389 — Identity Imprint Cruciate Retaining (CR) Knee Replacement System and Identity Imprint Posterior Stabilized (PS) Knee Replacement System · Conformis, Inc. · Nov 10, 2021
K113756 — HIGHLY CROSS LINKED VE CENTRAL PEG PATELLA HIGHLY CROSS LINKED VE TRI-PEG PATELLA HIGHLY CROSS LINKED VE METAL BACKED PA · Encore Medical L.P. · Mar 14, 2012
K221059 — Identity Imprint Knee Replacement System (including Identity Imprint Posterior Stabilized (PS) KRS and Identity Imprint Cruciate Retaining (CR) KRS) · Conformis, Inc. · May 10, 2022
K150105 — Klassic Knee System-Sombrero Patella · Total Joint Orthopedics, Inc. · Feb 17, 2015
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
December 15, 2025
Howmedica Osteonics Corp. Dba Stryker Orthopaedics
Richa Sharma
Principal Regulatory Affairs Specialist
325 Corporate Drive
Mahwah, New Jersey 07430
Re: K252898
Trade/Device Name: Triathlon® Tritanium® Asymmetric patella; Triathlon® Tritanium® Symmetric Patella; Triathlon® Symmetric Patella; Triathlon® Asymmetric Patella; Scorpio® Universal Dome Patella; Triathlon® Hinge Bumper
Regulation Number: 21 CFR 888.3510
Regulation Name: Knee Joint Femorotibial Metal/Polymer Constrained Cemented Prosthesis
Regulatory Class: Class II
Primary Product Code: KRO
Additional Product Code(s): JWH, MBH
Dated: September 11, 2025
Received: September 11, 2025
Dear Richa Sharma:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252898 - Richa Sharma
Page 2
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{2}
K252898 - Richa Sharma
Page 3
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Lixin Liu -S
Lixin Liu, Ph.D
Assistant Director
DHT6A: Division of Joint Arthroplasty Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K252898
Device Name
Scorpio® Universal Dome Patella; Triathlon® Hinge Bumper;
Triathlon® Symmetric Patella; Triathlon® Asymmetric Patella;
Triathlon® Tritanium® Asymmetric Patella; Triathlon® Tritanium® Symmetric Patella
Indications for Use (Describe)
Indications for Scorpio® Universal Dome Patella
- Painful, disabling joint disease of the knee resulting from: degenerative arthritis, rheumatoid arthritis or post-traumatic arthritis.
- Post-traumatic loss of knee joint configuration and function.
- Moderate varus, valgus, or flexion deformity in which the ligamentous structures can be returned to adequate function and stability.
- Revision of previous unsuccessful knee replacement or other procedure.
Indications for Triathlon® Hinge Bumper
Triathlon® Hinge Knee System is intended to be implanted with bone cement for the following condition(s):
- There is destruction of the joint surfaces, with or without significant bone deformity.
- The cruciate and/or collateral ligaments do not stabilize the knee joint.
- The ligaments are inadequate and/or the musculature is weak. And/or
- Revision is required of a failed prosthesis where there has been gross instability, with or without bone loss or inadequate soft tissue.
When used with MRH femur and/or MRH tibial baseplate replacement indicated in revision of an existing prosthesis:
- Revision is required of a failed prosthesis where there has been gross instability, with or without bone loss or inadequate soft tissue.
When used with compatible GMRS components:
- Where segmental resection and/or replacement of femur and/or proximal tibia is required.
Indications for Triathlon® Symmetric Patella and Triathlon® Asymmetric Patella
- Painful, disabling joint disease of the knee resulting from: noninflammatory degenerative joint disease (including osteoarthritis, traumatic arthritis, or avascular necrosis), rheumatoid arthritis or post-traumatic arthritis.
- Post-traumatic loss of knee joint configuration and function.
- Moderate varus, valgus, or flexion deformity in which the ligamentous structures can be returned to adequate function and stability.
- Revision of previous unsuccessful knee replacement or other procedure.
- Fracture of the distal femur and/or proximal tibia that cannot be stabilized by standard fracture management techniques.
Indications for Triathlon® Tritanium® Asymmetric Patella and Triathlon® Tritanium® Symmetric Patella
- Painful, disabling joint disease of the knee resulting from: noninflammatory degenerative joint disease (including osteoarthritis, traumatic arthritis, or avascular necrosis), rheumatoid arthritis or post-traumatic arthritis.
- Post-traumatic loss of knee joint configuration and function.
- Moderate varus, valgus, or flexion deformity in which the ligamentous structures can be returned to adequate function and stability.
- Revision of previous unsuccessful knee replacement or other procedure.
- Fracture of the distal femur and/or proximal tibia that cannot be stabilized by standard fracture management techniques.
FORM FDA 3881 (8/23)
PEC Publishing Services (301) 443-6740
{4}
The Triathlon Tritanium Tibial Baseplate and Tritanium Metal-Backed Patella components are indicated for both uncemented and cemented use.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
PSC Publishing Services (301) 443-6740
{5}
K252898 - Page 1 of 4
# 510(k) Summary
Sponsor
Howmedica Osteonics Corp. Dba Stryker Orthopaedics
325 Corporate Drive
Mahwah, NJ 07430
Contact Person
Richa Sharma
Principal Regulatory Affairs Specialist
Howmedica Osteonics Corp
325 Corporate Drive
Mahwah, NJ 07430
+44 7814 546161
Richa.sharma@stryker.com
Alternate Contact
Ka Zoua Xiong
Regulatory Affairs Consultant
Howmedica Osteonics Corp
325 Corporate Drive
Mahwah, NJ 07430
kazoua.xiong@stryker.com
Date Prepared:
December 14, 2025
Proprietary Name:
Triathlon® Tritanium® Asymmetric Patella
Triathlon® Tritanium® Symmetric Patella
Triathlon® Symmetric Patella
Triathlon® Asymmetric Patella
Scorpio® Universal Dome Patella
Triathlon® Hinge Bumper
Common Name:
Artificial Knee Replacement Components – Patellar Components and Hinge Knee Bumper
Classification Name:
Knee joint femorotibial metal/polymer constrained cemented Prosthesis (21 CFR §888.3510)
Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis (21 CFR §888.3560)
Knee joint patellofemorotibial metal/polymer porous-coated uncemented prosthesis (21 CFR §888.3565)
Product Codes:
KRO, JWH, MBH
{6}
K252898 - Page 2 of 4
Legally Marketed Predicate Device to Which Substantial Equivalence is Claimed:
The primary predicate for this submission is K242831.
Legally Marketed Additional Predicate Devices
| Device | Previous Premarket Notifications |
| --- | --- |
| Triathlon® Tritanium® Asymmetric Patella | K132624, K141056 |
| Triathlon® Tritanium® Symmetric Patella | |
| Triathlon® Symmetric Patella | K040267, K141056 |
| Triathlon® Asymmetric Patella | |
| Scorpio® Universal Dome Patella | K972967, K243817 |
| Triathlon® Hinge Bumper | K223528, K230416, K251665 |
Rationale for Bundling:
The change proposed is only related to updating the packaging configuration of the Ultra High Molecular Weight Polyethylene (UHMWPE) devices that are packaged in a Nitrogen Vacuum (N₂Vac) environment. The same proposed packaging configurations may be used among multiple generic product types, and the same supporting data can be used across all product types. In all testing performed to support the proposed packaging change, worst-case devices, representative of all products, were used. Therefore, the testing data represents all products currently bundled in the subject 510(k) submission.
Device Description:
The devices covered by this Traditional 510(k) Premarket Notification are Stryker UHMWPE devices (including knee patella devices and hinged knee bumper) that are packaged in a N₂Vac environment. The packaging for these devices consists of an impermeable foil pouch in a Polyethylene Terephthalate Glycol (PETG) blister sealed with a Tyvek lid. All subject devices are commercially available and have been found to be substantially equivalent in previous 510(k)s.
Intended Use:
The subject devices have the same intended use as those specified in the 510(k) submissions for the predicate devices listed. In general, these devices are intended for use in primary or revision knee arthroplasty.
{7}
K252898 - Page 3 of 4
## Indications for Use:
### Triathlon® Symmetric and Asymmetric Patella
- Painful, disabling joint disease of the knee resulting from: noninflammatory degenerative joint disease (including osteoarthritis, traumatic arthritis, or avascular necrosis), rheumatoid arthritis or post-traumatic arthritis.
- Post-traumatic loss of knee joint configuration and function.
- Moderate varus, valgus, or flexion deformity in which the ligamentous structures can be returned to adequate function and stability.
- Revision of previous unsuccessful knee replacement or other procedure.
- Fracture of the distal femur and/or proximal tibia that cannot be stabilized by standard fracture management techniques.
### Triathlon® Tritanium® Asymmetric and Symmetric Patella
- Painful, disabling joint disease of the knee resulting from: noninflammatory degenerative joint disease (including osteoarthritis, traumatic arthritis, or avascular necrosis), rheumatoid arthritis or post-traumatic arthritis.
- Post-traumatic loss of knee joint configuration and function.
- Moderate varus, valgus, or flexion deformity in which the ligamentous structures can be returned to adequate function and stability.
- Revision of previous unsuccessful knee replacement or other procedure.
- Fracture of the distal femur and/or proximal tibia that cannot be stabilized by standard fracture management techniques.
The Triathlon Tritanium Tibial Baseplate and Tritanium Metal-Backed Patella components are indicated for both uncemented and cemented use.
### Triathlon® Hinge Bumper
Triathlon® Hinge Knee System is intended to be implanted with bone cement for the following condition(s):
- There is destruction of the joint surfaces, with or without significant bone deformity.
- The cruciate and/or collateral ligaments do not stabilize the knee joint.
- The ligaments are inadequate and/or the musculature is weak. And/or
- Revision is required of a failed prosthesis where there has been gross instability, with or without bone loss or inadequate soft tissue.
When used with MRH femur and/or MRH tibial baseplate replacement indicated in revision of an existing prosthesis:
- Revision is required of a failed prosthesis where there has been gross instability, with or without bone loss or inadequate soft tissue.
When used with compatible GMRS components:
- Where segmental resection and/or replacement of femur and/or proximal tibia is required.
### Scorpio® Universal Dome Patella
- Painful, disabling joint disease of the knee resulting from: degenerative arthritis, rheumatoid arthritis or post-traumatic arthritis.
- Post-traumatic loss of knee joint configuration and function.
{8}
K252898 - Page 4 of 4
- Moderate varus, valgus, or flexion deformity in which the ligamentous structures can be returned to adequate function and stability.
- Revision of previous unsuccessful knee replacement or other procedure.
## Summary of Technological Characteristics:
The subject devices are equivalent in intended use, indications for use, product design, product materials, and operational principles to their corresponding predicates. The only difference between the subject devices and corresponding predicate devices are their packaging configurations. The packaging components of the predicate devices contain an inner and outer blister of Barex® Modified Acrylonitrile copolymer sealed with inner and outer foil lids purged with alternating vacuum and nitrogen (N₂Vac). The packaging of the subject devices consist of an impermeable foil pouch in a Polyethylene Terephthalate Glycol (PETG) blister sealed with Tyvek lid followed by vacuum and nitrogen (N₂VAC) purging.
## Non-Clinical Testing:
The following non-clinical laboratory testing was performed to determine substantial equivalence:
A study was performed on the subject UHMWPE devices to evaluate the effect of the proposed packaging modifications. The study compared the following material properties of the subject UHMWPE devices for the proposed and current packaging configurations:
- Small Punch per ASTM F2977
- Oxygen Content
- Oxidation Index per ASTM F2102
A ship test study was completed on the subject devices to qualify the proposed packaging configurations. Testing was completed per ISO 11607-1, ASTM F88 / F88M, and ASTM F2096.
An Accelerated Aging study was completed to validate a shelf-life of five (5) years (inclusive of final month) for the subject devices marketed as STERILE per ISO 11607-1, ISO 11607-2, and ASTM F1980.
Product endotoxin and cytotoxicity testing were executed as the proposed packaging configuration constitutes a change in packaging materials that contact the product after final cleaning. Endotoxin testing was completed per AAMI ST72, and cytotoxicity testing was completed per ISO 10993-5.
## Clinical Testing:
Clinical testing was not required as a basis for substantial equivalence.
## Conclusion:
Based upon a comparison of the intended use, indications for use, design, material, sterilization method, technical and performance characteristics, and operational principles, the subject devices are substantially equivalent to the predicate devices identified in this Premarket Notification.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.