Triathlon® Hinge Knee System

K223528 · Howmedica Osteonics Corp., Dba Stryker Orthopaedics · KRO · Jan 20, 2023 · Orthopedic

Device Facts

Record IDK223528
Device NameTriathlon® Hinge Knee System
ApplicantHowmedica Osteonics Corp., Dba Stryker Orthopaedics
Product CodeKRO · Orthopedic
Decision DateJan 20, 2023
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3510
Device ClassClass 2
AttributesTherapeutic

Intended Use

The THK System components are sterile, single-use devices intended for cemented use in primary and revision total knee arthroplasty. The THK System is intended to be used for the treatment of a severely unstable knee, particularly with the loss of collateral support in both primary and revision cases. The THK System can also be incorporated into implants for distal femoral and total femoral replacements performed for the treatment of bone tumors. It is a partially constrained prosthesis with limited rotational capability, which allows more natural motion and reduces torque on the fixation.

Device Story

Triathlon Hinge Knee (THK) System is a tricompartmental knee prosthesis; line extension to Triathlon Total Knee System. Device components: hinge femoral component, tibial bearing component, hinge tibial insert, revision tibial baseplate, and optional distal femoral/tibial augments. Operates via hinge axle/bushing assembly for flexion/extension; bearing surface articulation provides internal/external rotation. Used in orthopedic surgery for primary/revision total knee arthroplasty or distal/total femoral replacement. Implanted by orthopedic surgeons using bone cement. Provides varus/valgus stability and rotational constraint; reduces torque on fixation; restores joint function in severely unstable knees.

Clinical Evidence

No clinical data. Substantial equivalence supported by non-clinical laboratory testing and engineering analysis, including UHMWPE material characterization, fatigue testing (ASTM F1800-19e1), wear analysis (ISO 14243-1/3), range of motion/rotational freedom analysis, and biocompatibility (ISO 10993-1:2020).

Technological Characteristics

Tricompartmental knee prosthesis; materials include UHMWPE. Features hinge femoral component, revision tibial baseplate, hinge inserts, and tibial bearing components. Mechanical constraint via hinge axle/bushing and bearing surface radius. Sterile, single-use, cemented fixation. Bench testing per ASTM F1800-19e1, ISO 14243, ISO 10993-1, ISO 11607, and ANSI/AAMI ST72.

Indications for Use

Indicated for patients with destruction of joint surfaces (with/without bone deformity), knee joint instability due to inadequate cruciate/collateral ligaments, weak musculature, revision of failed prosthesis with gross instability (with/without bone loss or inadequate soft tissue), or cases requiring segmental resection and replacement of the distal femur.

Regulatory Classification

Identification

A knee joint femorotibial metal/polymer constrained cemented prosthesis is a device intended to be implanted to replace part of a knee joint. The device limits translation or rotation in one or more planes and has components that are linked together or affined. This generic type of device includes prostheses composed of a ball-and-socket joint located between a stemmed femoral and a stemmed tibial component and a runner and track joint between each pair of femoral and tibial condyles. The ball-and-socket joint is composed of a ball at the head of a column rising from the stemmed tibial component. The ball, the column, the tibial plateau, and the stem for fixation of the tibial component are made of an alloy, such as cobalt-chromium-molybdenum. The ball of the tibial component is held within the socket of the femoral component by the femoral component's flat outer surface. The flat outer surface of the tibial component abuts both a reciprocal flat surface within the cavity of the femoral component and flanges on the femoral component designed to prevent distal displacement. The stem of the femoral component is made of an alloy, such as cobalt-chromium-molybdenum, but the socket of the component is made of ultra-high molecular weight polyethylene. The femoral component has metallic runners which align with the ultra-high molecular weight polyethylene tracks that press-fit into the metallic tibial component. The generic class also includes devices whose upper and lower components are linked with a solid bolt passing through a journal bearing of greater radius, permitting some rotation in the transverse plane, a minimal arc of abduction/adduction. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. January 20, 2023 Howmedica Osteonics Corp., dba Stryker Orthopaedics Allison Byrne Senior Specialist, Regulatory Affairs 325 Corporate Drive Mahwah, New Jersey 07430 Re: K223528 Trade/Device Name: Triathlon® Hinge Knee System Regulation Number: 21 CFR 888.3510 Regulation Name: Knee Joint Femorotibial Metal/Polymer Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: KRO Dated: November 22, 2022 Received: November 23, 2022 Dear Allison Byrne: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. # Ting Song -S Ting Song, Ph.D. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) #### K223528 Device Name Triathlon® Hinge Knee System #### Indications for Use (Describe) This Rotating Hinge Knee System is intended to be implanted with bone cement for the following condition(s): - · There is destruction of the joint surfaces, with or without significant bone deformity. - · The cruciate and/or collateral ligaments do not stabilize the knee joint. - · The ligaments are inadequate and/or the musculature is weak. And/or - · Revision is required of a failed prosthesis where there has been gross instability, with or without bone loss or madequate soft tissue. - · And/or where segmental resection and replacement of the distal femur is required. | Type of Use (Select one or both, as applicable) | | |--------------------------------------------------------------------------------------------------------|---------------------------------------------| | <span style="text-decoration: underline;"><b>X</b></span> Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ #### 510(k) Summary | Sponsor | Stryker Orthopaedics<br>325 Corporate Drive<br>Mahwah, NJ 07430 | |----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Person | Allison Byrne<br>Senior Specialist, Regulatory Affairs<br>Howmedica Osteonics Corp<br>325 Corporate Drive<br>Mahwah, NJ 07430<br>Phone: (201) 831-5732<br>Email: allison.byrne@stryker.com | | Alternate Contact | Margaret Klippel<br>Chief Specialist, Regulatory Affairs<br>Howmedica Osteonics Corp<br>325 Corporate Drive<br>Mahwah, NJ 07430<br>Phone: (201) 831-5559<br>Email: magaret.klippel@stryker.com | | Date Prepared: | November 22, 2022 | | Proprietary Name: | Triathlon® Hinge Knee System | | Common Name: | Rotating Hinge Knee System<br>Artificial Knee System<br>Total Knee Joint Prosthesis | | Classification Name: | Knee joint femorotibial metal/polymer constrained cemented<br>prosthesis. (21 CFR Part 888.3510) | | Product Codes: | KRO | #### Legally Marketed Primary Predicate Devices to Which Substantial Equivalence is Claimed: - o The components of the Modular Rotating Hinge Knee System: - Modular Rotating Hinge (MRH) Knee Femoral Components – K994207, K002552 - MRH Hinge Tibial Inserts K994207, K002552 ■ - MRH Sleeve - K994207, K002552 - MRH Tibial Baseplate - K994207, K002552 - MRH Tibial Bearing Component K994207, K001548, K001957, K002552 ■ - MRH Bumper components K994207, K002552 ■ - MRH Knee Femoral Distal Block – K994207 {4}------------------------------------------------ ## Legally Marketed Additional Predicate Devices Used to Support Substantial Equivalence: - o The components of the Triathlon® Total Knee System: - Triathlon® Total Stabilized (TS) Femoral Components – K070095, K141056, K172326 - Triathlon® TS Plus Inserts K070095, K072221, K141056, K172326, K172634 ■ - Triathlon® Universal Baseplate K053514, K141056, K172326 ■ - Triathlon® Femoral Distal Augments - K070095, K141056, K172326 - 트 Triathlon® Tibial Augments - K053514, K141056, K172326 ## Legally Marketed Reference Devices Used to Support Substantial Equivalence in terms of biocompatibility: - Triathlon® Tritanium Metal Back Patella K132624, K141056, K172326 . - Triathlon® Posterior Stabilized (PS) Tibial Insert K031729, K141056, K172326 . - Osteonics® Omnifit® X Plus Cemented Hip Stem K983226. K153345 ■ ## Reason for 510(k) Submission: The purpose of this submission is to introduce a line extension to the Triathlon® Total Knee System, specifically a new Triathlon® Hinge Knee (THK) System. ## Device Description: The subject THK System is a line extension to the existing Triathlon® Knee System and will be a modified version of the predicate Modular Rotating Hinge (MRH) Knee System and Triathlon® Total Knee System. The THK System is a tricompartmental knee system consisting of a new hinge femoral component and a new tibial bearing component connected by a set of previously cleared MRH bushings and an MRH axle (K994207, K002552). A new bumper locks this assembly. This assembly provides motion through the MRH axle/bushings combination in the flexion/extension plane. The articulation between the bearing surfaces on the underside of the new tibial bearing component and a new hinge tibial insert provides motion in the rotating plane. The hinge tibial insert is assembled to a new revision tibial baseplate which incorporates a longitudinal bore to accept a new tibial sleeve or previously cleared MRH tibial sleeve (K994207, K002552). Optional new distal femoral and tibial augments are available to fill bone defects. The THK System is designed to provide varus/valgus stability throughout the range of motion, internal/external rotation about the tibial axis, constraint by the bearing surface radius on the tibial bearing component, and an extensive range of size, modularity, and resection options. The subject THK System consists of: - Hinge Femoral Components in six sizes and in left and right configurations o - Revision Tibial Baseplates in seven sizes O - Hinge Inserts in seven sizes and five thicknesses each hinge insert is packaged with a O standard sleeve subcomponent {5}------------------------------------------------ - Tibial Bearing Components in three sizes O - Bumper inserts in neutral and three degree flexion options O - Revision Tibial Augments in eight sizes, two thicknesses and in right medial/left lateral and O right lateral/left medial options - o Femoral Distal Augments in six sizes and two thicknesses. The components of the subject THK System are sterile, single-use devices intended for cemented use only. They can be used with previously cleared Modular Rotating Hinge (MRH) Knee components (K994207, K002552), the Global Modular Replacement System (GMRS) (K023087), and Triathlon® Knee System components (K172634, K172326, K190991, K143396, K141056. K132624. K070095. K061521. K053514. K052917. K051948. K051146. K040267). #### Intended Use: The component of the subject THK System have the same intended use as those specified in the 510(k) submissions for the components of the predicate devices listed. The THK System components are sterile, single-use devices intended for cemented use in primary and revision total knee arthroplasty. The THK System is intended to be used for the treatment of a severely unstable knee, particularly with the loss of collateral support in both primary and revision cases. The THK System can also be incorporated into implants for distal femoral and total femoral replacements performed for the treatment of bone tumors. It is a partially constrained prosthesis with limited rotational capability, which allows more natural motion and reduces torque on the fixation. #### Indications: This Rotating Hinge Knee System is intended to be implanted with bone cement for the following condition(s): - There is destruction of the joint surfaces, with or without significant bone deformity. ● - The cruciate and/or collateral ligaments do not stabilize the knee joint. ● - The ligaments are inadequate and/or the musculature is weak. And/or - Revision is required of a failed prosthesis where there has been gross instability, with or ● without bone loss or inadequate soft tissue. - And/or where segmental resection and replacement of the distal femur is required. ● #### Summary of Technological Characteristics: The device comparisons and performance testing show that the components of the subject THK System are substantially equivalent to those of the predicate MRH Knee and Triathlon® Total Knee System based on intended use, indications for use, design, materials and sterilization method, performance characteristics, and operational principles. {6}------------------------------------------------ ## Non-Clinical Testing: The following non-clinical laboratory testing and/or engineering analysis were performed to determine substantial equivalence: - Materials Characterization - UHMWPE - Triathlon® Hinge Femoral Component Fatigue Testing ● - Fatigue testing of the Revision Tibial Baseplate Components per ASTM F1800-19e1 ● - o Fatigue testing of the Tibial Bearing components - o Tibial Bearing Component Chair Rise Testing - Shear fatigue analysis of the baseplate/insert locking mechanism - Full construct fatigue worst-case analysis - Wear Analysis per ISO 14243-1 and ISO 14243-3 ● - Analysis of contact area/contact stress - Tibial Bearing/Hinge Insert Components - Patellofemoral Contact Area/Contact Stress and Constraint Analysis - Range of motion and rotational freedom analysis - Revision Tibial Augment and Locking Strength Analysis - Femoral Distal Augment and Locking Strength Analysis - Strength analysis for Hinge Femur Component's boss/stem and Revision Tibial ● Baseplate Component's boss/stem junctions - Biocompatibility evaluated per ISO 10993-1:2020 - Shelf-life validated per the following standards: ● - ISO 11607-1:2019 o - o ISO 11607-2:2019 - o ASTM F1980-21 - Testing performed per the following methods: O - 트 ASTM F1886/F1886M-16 - ASTM F88/88M-21 - . ASTM F2096-11(2019) - o Bacterial endotoxin testing (BET) as specified in ANSI/AAMI ST72:2019 was used for pyrogenicity testing to achieve an endotoxin limit of < 20EU/Device. # Clinical Testing: Clinical testing was not required as a basis for substantial equivalence. # Conclusion: Based upon a comparison of the intended use, indications for use, design, materials and sterilization method, performance characteristics, and operational principles, the components of the subject THK System are substantially equivalent to those of the predicate devices identified in this premarket notification.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...