K062301 · DePuy Orthopaedics, Inc. · KRO · Oct 13, 2006 · Orthopedic
Device Facts
Record ID
K062301
Device Name
DEPUY LPS PROXIMAL TIBIAL COMPONENT
Applicant
DePuy Orthopaedics, Inc.
Product Code
KRO · Orthopedic
Decision Date
Oct 13, 2006
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3510
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The DePuy LPS is intended for use in replacement of the mid-shaft portion of the femur, proximal, distal and/or total femur, and proximal tibia, especially in cases that require extensive resection and replacement. Specific diagnostic indications for use include: - malignant tumors (e.g., osteosarcomas, chondrosarcomas, giant cell tumors, bone tumors) . requiring extensive resection and replacement; - . patient conditions of noninflammatory degenerative joint disease (NIDJD), e.g. avascular necrosis, osteoarthritis, and inflammatory joint disease (IJD), e.g., theumatoid arthritis, requiring extensive resection and replacement: - . revision cases for a failed previous prosthesis requiring extensive resection and replacement: - . severe trauma requiring extensive resection and replacement. The LPS is also intended for use in bone loss post-infection, where the surgeon has elected to excise the bone and replacement is required. The distal femoral component, the S-ROM Tibial Tray and the non-porous coated straight and bowed stems are intended for cemented use only.
Device Story
Modular proximal tibial component; replaces proximal tibia in oncology or end-stage revision cases requiring extensive bone resection. Features female Morse-type taper for connection to segment components or tibial stems; central stem hole for tibial plateau/hinge assembly; polished mirror finish for articulation; anterior surface with Porocoat beaded Co-Cr-Mo porous coating. Used in surgical settings by orthopedic surgeons. Provides structural replacement for resected bone; facilitates joint reconstruction. Benefits patient by restoring limb function following severe bone loss or tumor resection.
Clinical Evidence
Bench testing only.
Technological Characteristics
Modular proximal tibial component; materials include Co-Cr-Mo alloy with Porocoat beaded porous coating. Features female Morse-type taper and central stem hole. Designed for cemented use in specific configurations.
Indications for Use
Indicated for patients requiring extensive bone resection and replacement of the proximal tibia or femur due to malignant tumors, noninflammatory degenerative joint disease (e.g., avascular necrosis, osteoarthritis), inflammatory joint disease (e.g., rheumatoid arthritis), failed previous prosthesis, severe trauma, or bone loss post-infection.
Regulatory Classification
Identification
A knee joint femorotibial metal/polymer constrained cemented prosthesis is a device intended to be implanted to replace part of a knee joint. The device limits translation or rotation in one or more planes and has components that are linked together or affined. This generic type of device includes prostheses composed of a ball-and-socket joint located between a stemmed femoral and a stemmed tibial component and a runner and track joint between each pair of femoral and tibial condyles. The ball-and-socket joint is composed of a ball at the head of a column rising from the stemmed tibial component. The ball, the column, the tibial plateau, and the stem for fixation of the tibial component are made of an alloy, such as cobalt-chromium-molybdenum. The ball of the tibial component is held within the socket of the femoral component by the femoral component's flat outer surface. The flat outer surface of the tibial component abuts both a reciprocal flat surface within the cavity of the femoral component and flanges on the femoral component designed to prevent distal displacement. The stem of the femoral component is made of an alloy, such as cobalt-chromium-molybdenum, but the socket of the component is made of ultra-high molecular weight polyethylene. The femoral component has metallic runners which align with the ultra-high molecular weight polyethylene tracks that press-fit into the metallic tibial component. The generic class also includes devices whose upper and lower components are linked with a solid bolt passing through a journal bearing of greater radius, permitting some rotation in the transverse plane, a minimal arc of abduction/adduction. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
Predicate Devices
DePuy Orthogenesis LPS Proximal Tibial Replacement System (K011810)
K110940 — ZIMMER SEGMENT SYSTEM TRABECULAR METAL PROXIMAL TIBIAL COMPONENT, TRABECULAR METAL PROXIMAL COMPONENT, ADDITIONAL SEGMEN · Zimmer, Inc. · Jul 1, 2011
K063686 — DEPUY LPS DISTAL FEMORAL COMPONENT · DePuy Orthopaedics, Inc. · Dec 20, 2007
Submission Summary (Full Text)
{0}------------------------------------------------
## 510(k) SUMMARY
OC7 1 3 2006
| NAME OF FIRM: | DePuy Orthopaedics Inc. 700 Orthopaedic Drive Warsaw, Indiana 46581-0988 Establishment Registration No.: 1818910 |
|-----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(K) CONTACT: | Nancy Friddle Team Leader, Regulatory Affairs Tel: (574) 371-4923 Fax: (574) 371-4987 |
| TRADE NAME: | DePuy LPS Proximal Tibial Component |
| COMMON NAME: | Proximal Tibial Replacement |
| CLASSIFICATION: | Knee joint femorotibial metal/polymer constrained cemented prosthesis: Class II per 21 CFR §888.3510 |
| DEVICE PRODUCT CODE: | KRO |
| SUBSTANTIALLY EQUIVALENT DEVICES: | DePuy Orthogenesis LPS Proximal Tibial Replacement System (K011810)<br>DePuy LPS (K033959) |
### DEVICE DESCRIPTION:
The DePuy LPS Proximal Tibial Component is a modular component that is designed to replace the proximal portion of the tibia. Unlike primary knee systems, the LPS System is used when the amount of bone resection and replacement is extreme (e.g. in oncology cases or endstage revision).
The distal end of the LPS Proximal Tibial Component has a female Morse-type taper that is designed to accept either a DePuy LPS Segment Component or a tibial stem. The proximal/superior surface is designed with a central stem hole to allow it to accept the stem of the LPS tibial plateau/hinge assembly. The top surface has a highly polished mirror pr nish for smooth articulation with the polyethylene component of the tibial plateau/hinge assembly. The anterior surface has a Porocoat beaded Co-Cr-Mo porous coating.
{1}------------------------------------------------
# INDICATIONS FOR USE:
The DePuy LPS is intended for use in replacement of the mid-shaft portion of the femur, proximal, distal and/or total femur, and proximal tibia, especially in cases that require extensive resection and replacement. Specific diagnostic indications for use include:
- malignant tumors (e.g., osteosarcomas, chondrosarcomas, giant cell tumors, bone tumors) . requiring extensive resection and replacement;
- patient conditions of noninflammatory degenerative joint disease (NIDJD), e.g. avascular . necrosis, osteoarthritis, and inflammatory joint disease (IJD), e.g., theumatoid arthritis. requiring extensive resection and replacement;
- revision cases for a failed previous prosthesis requiring extensive resection and . replacement:
- severe trauma requiring extensive resection and replacement. .
The LPS is also intended for use in bone loss post-infection, where the surgeon has elected to excise the bone and replacement is required.
The distal femoral component, the S-ROM Tibial Tray and the non-porous coated straight and bowed stems are intended for cemented use only.
## BASIS FOR SUBSTANTIAL EQUIVALENCE:
DePuy believes that this device is substantially equivalent to the Orthogenesis LPS Proximal Tibial Replacement System based on similarities in design and indications and the + revits of physical testing and analysis compared to the predicate devices.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its body and wing. The eagle faces to the right. Encircling the eagle is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES . USA" in a circular arrangement.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
### OCT 1 3 2006
DePuy Orthopaedics, Inc. % Ms. Nancy Friddle Team Leader, Regulatory Affairs 700 Orthopaedic Drive Warsaw, Indiana 46581
Re: K062301
Trade/Device Name: LPS Proximal Tibial Component Regulation Number: 21 CFR 888.3510 Regulation Name: Knee joint femortibial metal/polymer constrained cemented prosthesis Regulatory Class: Class II Product Code: KRO Dated: October 5, 2006 Received: October 6, 2006
Dear Mr. Friddle:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Ms. Nancy Friddle
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0210 Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
### Indications for Use Statement
510(k) Number (if known): K062301
### Device Name: DePuy LPS Proximal Tibial Component
The DePuy LPS is intended for use in replacement of the mid-shaft portion of the femur, proximal, distal and/or total femur, and proximal tibia, especially in cases that require extensive resection and replacement. Specific diagnostic indications for use include:
- malignant tumors (e.g., osteosarcomas, chondrosarcomas, giant cell tumors, bone tumors) . requiring extensive resection and replacement;
- . patient conditions of noninflammatory degenerative joint disease (NIDJD), e.g. avascular necrosis, osteoarthritis, and inflammatory joint disease (IJD), e.g., theumatoid arthritis, requiring extensive resection and replacement:
- . revision cases for a failed previous prosthesis requiring extensive resection and replacement:
- . severe trauma requiring extensive resection and replacement.
The LPS is also intended for use in bone loss post-infection, where the surgeon has elected to excise the bone and replacement is required.
The distal femoral component, the S-ROM Tibial Tray and the non-porous coated straight and bowed stems are intended for cemented use only.
| Prescription Use<br>(Part 21 CFR 801 Subpart D) | X | AND/OR | Over-The-Counter Use<br>(21 CFR 801 Subpart C) |
|-------------------------------------------------|---|--------|------------------------------------------------|
|-------------------------------------------------|---|--------|------------------------------------------------|
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of General, Restorative, and Neurological Devices
| 510(k) Number | K062301 |
|---------------|---------|
|---------------|---------|
| Page 1 of 1 |
|-------------|
|-------------|
Page 25
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.