← Product Code [KRO](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KRO) · K023667

# RT-PLUS SOLUTION & RT-PLUS MODULAR SOLUTION KNEE (K023667)

_Plus Orthopedics · KRO · Dec 24, 2002 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KRO/K023667

## Device Facts

- **Applicant:** Plus Orthopedics
- **Product Code:** [KRO](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KRO.md)
- **Decision Date:** Dec 24, 2002
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3510
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The RT-PLUS Knee System is a tricompartmental rotating hinged prosthesis o the total condylar type. The system consists of femoral, tibial and patellar components. It is indicated for use for replacement of the total knee joint in which significant bone loss and/or ligamentous deficiencies have occurred due to tumors, trauma, infections, revision or connective tissue disorders. RT-PLUS Modular Knee provides joint stability when any or all of the following structures are non-functional: MCL, LCL, PCL, ACL or the illiotibial band.

## Device Story

Hinged tri-compartmental knee prosthesis; includes femoral, tibial, and patellar components. Used for total knee joint replacement in cases of significant bone loss or ligamentous deficiency (MCL, LCL, PCL, ACL, or iliotibial band). Modifications include additional sizes for femoral/tibial baseplates, new CoCrMo augmentation blocks, and additional titanium alloy stem sizes. Cemented application. Provides joint stability in complex revision or trauma scenarios.

## Clinical Evidence

Bench testing only. Compliance with ISO 14789-1 (endurance properties of knee tibial trays) demonstrated.

## Technological Characteristics

Tricompartmental rotating hinged prosthesis. Materials: Ultra-high molecular weight polyethylene (ASTM F648), titanium alloy (ASTM F136), CoCrMo (ASTM F75). Cemented fixation. Modular femoral and tibial stems available.

## Regulatory Identification

A knee joint femorotibial metal/polymer constrained cemented prosthesis is a device intended to be implanted to replace part of a knee joint. The device limits translation or rotation in one or more planes and has components that are linked together or affined. This generic type of device includes prostheses composed of a ball-and-socket joint located between a stemmed femoral and a stemmed tibial component and a runner and track joint between each pair of femoral and tibial condyles. The ball-and-socket joint is composed of a ball at the head of a column rising from the stemmed tibial component. The ball, the column, the tibial plateau, and the stem for fixation of the tibial component are made of an alloy, such as cobalt-chromium-molybdenum. The ball of the tibial component is held within the socket of the femoral component by the femoral component's flat outer surface. The flat outer surface of the tibial component abuts both a reciprocal flat surface within the cavity of the femoral component and flanges on the femoral component designed to prevent distal displacement. The stem of the femoral component is made of an alloy, such as cobalt-chromium-molybdenum, but the socket of the component is made of ultra-high molecular weight polyethylene. The femoral component has metallic runners which align with the ultra-high molecular weight polyethylene tracks that press-fit into the metallic tibial component. The generic class also includes devices whose upper and lower components are linked with a solid bolt passing through a journal bearing of greater radius, permitting some rotation in the transverse plane, a minimal arc of abduction/adduction. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).

## Predicate Devices

- RT-PLUS Knee ([K013340](/device/K013340.md))
- RT-PLUS Modular Knee ([K003504](/device/K003504.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K023(667

## Amended 510(k) Summary

Date November 26, 2002

Submitter

PLUS ORTHOPEDICS 6055 Lusk Blvd San Diego, CA 92121

#### Contact person

J.D. Webb 1001 Oakwood Blvd Round Rock, TX 78681 512-388-4694

#### Common name

Hinged knee

#### Classification name

Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer (per 21 CFR section 888.3510)

#### Equivalent Device

The RT-PLUS and RT-PLUS Knee systems are equivalent to the RT-PLUS Knee (K013340) and RT-PLUS Modular Knee (K003504).

#### Device Description

The RT-PLUS Knee and RT-PLUS Modular Knee are hinged tri-compartmental prostheses with femoral, tibial, and patella components and are of the cemented total condylar type. They are intended to be used for replacement of the total knee ioint in which significant bone loss and/or ligamentous deficiencies have occurred due to tumors, trauma, infections, revision or connective tissue disorders. Both of these knee systems provide joint stability when any or all of the following structures are non-functional: MCL, LCL, PCL, ACL or the illiotibial band. Both systems include polyethylene tibial augmentation blocks. The Modular knee includes modular femoral and tibial stems.

The modifications described in this submission include:

- Changes to the outside surfaces of the ultra high molecular weight polyethylene (ASTM F648) 1. augmentation blocks and addition of blocks for Size 2 and 10 tibial baseplates. (Original cleared on K003504)
- Additional titanium alloy (ASTM F136) stem sizes (Original cleared on K003504) 2.
- Addition of CoCrMo (ASTM F75) augmentation blocks 3.
- Changes to femoral and tibial baseplate components to accept CoCrMo blocks (Original cleared on 4. K003504)
- 5. Addition of size 2 and 10 femoral and tibial baseplate components (Non-modular cleared on K013340)

#### Intended Use

The RT-PLUS Knee System is indicated for use for replacement of the total knee joint in which significant bone loss and/or ligamentous deficiencies have occurred due to tumors, trauma, infections, revision or connective tissue disorders. RT-PLUS Modular Knee provides joint stability when any or all of the following structures are non-functional: MCL, LCL, PCL, ACL or the illiotibial band.

#### Summary of Technological Characteristics Compared to Predicate Device

The RT-PLUS and RT-PLUS Modular Knee systems with modifications have the same indications, materials and fundamental scientific technology as the RT-PLUS Knee (K013340) and RT-PLUS Modular Knee (K003504).

### Summary Nonclinical Tests

ISO 14789-1 "Implants for surgery -- Total knee-joint prostheses -- Part 1: Determination of endurance properties of knee tibial trays" was complied with.

DEC 2 4 2002

{1}------------------------------------------------

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol consists of three stylized human profiles facing to the right, with flowing lines connecting them. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

DEC 2 4 2002

PLUS Orthopedics c/o Mr. J.D. Webb 1001 Oakwood Road Round Rock, Texas 78681

Re: K023667

Trade/Device Name: RT-PLUS Solution and RT-PLUS Modular Solution Knee Regulation Number: 21 CFR §888.3510 Regulation Name: Knee Joint Femorotibial Metal/Polymer

Constrained Cemented Prosthesis

Regulatory Class: II Product Code: KRO Dated: November 29, 2002 Received: December 2, 2002

Dear Mr. Webb:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{2}------------------------------------------------

## Page 2 - Mr. J.D. Webb

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,

ia Witten, Ph.D., M.D. Ce Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

510(k) number (if known):

K023(6-)

Device Name: RT-PLUS & RT-PLUS Modular Knee Systems

Indications for Use:

# RT-PLUS™ & RT-PLUS™ Modular Knee Systems Indications for Use .

The RT-PLUS Knee System is a tricompartmental rotating hinged prosthesis o the total condylar type. The system consists of femoral, tibial and patellar components. It is indicated for use for replacement of the total knee joint in which significant bone loss and/or ligamentous deficiencies have occurred due to tumors, trauma, infections, revision or connective tissue disorders. RT-PLUS Modular Knee provides joint stability when any or all of the following structures are non-functional: MCL, LCL, PCL, ACL or the illiotibial band.

(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE

Prescription Use _X (per 21 CFR 801.109)OROver-the-Counter Use ____(Optional format 1-2-96)

Gella Welle (Division Sign-off)
Division of General, Neurological and Restorative Devices

sion of General, Neurological and Restorative Devices

510(k) Number KO23667

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KRO/K023667](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/KRO/K023667)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
