← Product Code [JWH](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JWH) · K253144

# Freedom® Total Knee System  Titan PCK Components (K253144)

_Maxx Orthopedics, Inc. · JWH · Oct 22, 2025 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JWH/K253144

## Device Facts

- **Applicant:** Maxx Orthopedics, Inc.
- **Product Code:** [JWH](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JWH.md)
- **Decision Date:** Oct 22, 2025
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3560
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Freedom® Total Knee System is indicated for patients with severe knee pain and the disability due to: • Severe knee joint pain, loss of mobility, and disability due to: rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis. • Correction of functional deformities. • Post-traumatic loss of knee joint contour, particularly when there is patellofemoral erosion, dysfunction, or prior patellectomy. • Moderate valgus, varus, or flexion trauma. • Knee fractures untreatable by other methods. • Revision surgery where sufficient bone stock and soft tissue integrity are present. (For PCK Components and Primary PCK Components only) The Freedom® Total Knee System – Titan PCK Components are intended for cemented and single use only.

## Device Story

Freedom® Total Knee System – Titan PCK Components are orthopedic implants for total knee replacement. System includes femoral augments, tibial augments, offset junctions, and stem extensions. Devices are fabricated from Ti-6Al-4V (ASTM F136) with Titanium Niobium Nitride (TiNbN) coating. Components are intended for cemented, single-use in revision or primary knee arthroplasty to address bone loss or structural deformity. Implants are surgically attached to previously cleared stemmed femoral and tibial components using screws, pins, and taper junctions. Coating provides surface modification to previously cleared uncoated designs. Surgeons use these components to restore joint function, stability, and mobility in patients with severe knee pathology. Benefits include improved implant fixation and structural support in cases of significant bone loss.

## Clinical Evidence

Bench testing only. No clinical data presented. Performance testing included coating characterization (thickness, hardness, adhesion). Mechanical and functional testing (static/dynamic properties, modular disassembly, range of motion, stability) was leveraged from previously cleared uncoated predicate devices, as the TiNbN coating does not affect these mechanical properties. Biocompatibility was leveraged from the previously cleared TiNbN-coated Freedom® Knee (K200912).

## Technological Characteristics

Materials: Ti-6Al-4V (ASTM F136) with TiNbN coating. Components: Femoral/tibial augments, offset junctions, stem extensions. Modular assembly via screws, pins, and taper junctions. Intended for cemented use. No software or energy source.

## Regulatory Identification

A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).

## Predicate Devices

- Freedom® PCK Femoral Components ([K131481](/device/K131481.md))
- Freedom® Stemmed Tibial Components ([K111785](/device/K111785.md))

## Reference Devices

- Freedom® TiNbN Coated Knee ([K200912](/device/K200912.md))
- Freedom® Total Knee System ([K240863](/device/K240863.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

October 22, 2025

Maxx Orthopedics Inc.
% Bittu Jha
Senior Manager - Regulatory Affairs
Meril Healthcare Pvt. Ltd.
H1 - H3, Meril Park, Survey No 135/2/B &amp; 174/2
Muktanand Marg, Chala
Vapi, Gujarat 396191
India

Re: K253144
Trade/Device Name: Freedom® Total Knee System - Titan PCK Components
Regulation Number: 21 CFR 888.3560
Regulation Name: Knee Joint Patellofemorotibial Polymer/Metal/Polymer Semi-Constrained Cemented Prosthesis
Regulatory Class: Class II
Product Code: JWH
Dated: September 23, 2025
Received: September 25, 2025

Dear Bittu Jha:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

U.S. Food &amp; Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K253144 - Bittu Jha
Page 2

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-

{2}

K253144 - Bittu Jha
Page 3

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Lixin Liu -S

Lixin Liu, Ph.D
Assistant Director
DHT6A: Division of Joint Arthroplasty Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF

|  DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.  |
| --- | --- |
|  510(k) Number (if known) K253144  |   |
|  Device Name Freedom® Total Knee System – Titan PCK Components  |   |
|  Indications for Use (Describe) The Freedom® Total Knee System is indicated for patients with severe knee pain and the disability due to: • Severe knee joint pain, loss of mobility, and disability due to: rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis. • Correction of functional deformities. • Post-traumatic loss of knee joint contour, particularly when there is patellofemoral erosion, dysfunction, or prior patellectomy. • Moderate valgus, varus, or flexion trauma. • Knee fractures untreatable by other methods. • Revision surgery where sufficient bone stock and soft tissue integrity are present. (For PCK Components and Primary PCK Components only) The Freedom® Total Knee System – Titan PCK Components are intended for cemented and single use only.  |   |
|  Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)  |   |
|  CONTINUE ON A SEPARATE PAGE IF NEEDED.  |   |
|  This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*  |   |
|  The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov  |   |
|  "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."  |   |

{4}

maxx orthopedics
510(k) SUMMARY
K253144 Page 1 of 7

# I. SUBMITTER

Maxx Orthopedics Inc.
2460 General Armistead Ave Suite 100
Norristown PA 19403 USA
Telephone No: +1.484.342.0092
Web site: www.maxxortho.com

## Applicant Information

Mr. Corey Perine
Chief Operating Officer
Maxx Orthopedics Inc.
E-mail: corey.perine@maxxortho.com

## Primary Correspondent Information

Bittu Jha
Senior Manager – Regulatory Affairs
Meril Healthcare Pvt. Ltd.
E-mail: bittu.jha@merillife.com

## Alternate Correspondent Information

Umesh Sharma
Sr. General Manager – Quality Assurance/ Regulatory Affairs
Meril Healthcare Pvt. Ltd.
E-mail: umesh.sharma@merillife.com

Date Prepared: October 09, 2025

{5}

maxx orthopedics

510(k) SUMMARY

K253144 Page 2 of 7

## II. SUBJECT DEVICE

|  Trade / Proprietary Name | Freedom® Total Knee System – Titan PCK Components  |
| --- | --- |
|  Common Name | Total Knee Replacement  |
|  Classification | Knee Joint Patellofemorotibial Polymer/Metal/Polymer Semi-Constrained Cemented Prosthesis  |
|  Regulatory Class | II  |
|  Product Code | JWH  |
|  Regulation Number | 21CFR 880.3560  |
|  Review Panel | Orthopedic  |

## III. PREDICATE DEVICES

|  Device Category | Product Code | Trade Name | Manufacturer | 510(k)  |
| --- | --- | --- | --- | --- |
|  Primary Predicate | JWH | Freedom® PCK Femoral Components | Maxx Orthopedics Inc. | K131481  |
|  Secondary Predicate | JWH | Freedom® Stemmed Tibial Components |   | K111785  |

## IV. REFERENCE DEVICES

|  Device Category | Product Code | Trade Name | Manufacturer | 510(k)  |
| --- | --- | --- | --- | --- |
|  Reference Device | JWH | Freedom® TiNbN Coated Knee | Maxx Orthopedics Inc. | K200912  |
|  Reference Device | JWH | Freedom® Total Knee System |   | K240863  |

{6}

maxx orthopedics
510(k) SUMMARY
K253144 Page 3 of 7

# V. DEVICE DESCRIPTION

The Freedom® Total Knee System is comprised of a femoral component, all-poly tibial component, patellar component, tibial base plate and tibial articular surface. The Freedom® Total Knee System's Femoral Component is offered as different versions such as stemmed PCK design, primary PCK, cruciate retaining, posterior stabilizing. The Freedom® Total Knee System's Tibial Base Plate is offered with stemmed design and without stemmed design. The Freedom® Total Knee System was originally cleared under the 510(k) number K082019. Later on, several modifications were made and were cleared under 510(k)s K091280, K192148, K090411, K182574, K131481, K111785, K200912, and K240863 respectively.

This submission seeks the clearance of Titanium Niobium Nitride (TiNbN) coated version of previously cleared Femoral Augment (non-coated version cleared in K131481), Tibial Augments, Stem Extension &amp; Offset Junction (non-coated versions cleared in K111785). The coated versions are now branded as Freedom® Total Knee System – Titan PCK Components.

Below is the description of the coated components.

# Freedom® Titan Femoral Augments

The Freedom® Titan Femoral Augments are fabricated from Titanium-6Aluminum-4Vanadium (Ti-6A1-4V), complaint with ASTM F136, and are coated with Titanium Niobium Nitride (TiNbN). These augments are intended to be screwed to the internal distal and posterior surfaces of the previously cleared PCK Stemmed Femoral Component (K131481) and Titan PCK Stemmed Femoral Component (K240863) when required in cases of significant bone loss. The augments are designed to be stackable and for use on either the medial or lateral side.

Augment Screws, fabricated from Titanium-6Aluminum-4Vanadium (Ti-6A1-4V), complaint with ASTM F136, and are coated with Titanium Niobium Nitride (TiNbN), are available as an accessory to the system to fix the augments to the Stemmed Femoral Component and other augments.

{7}

maxx orthopedics
510(k) SUMMARY
K253144 Page 4 of 7

The Freedom® Titan Femoral Augments is available in distal and posterior design configuration. Each configuration is further available in 6 different sizes.

## Freedom® Titan Tibial Augments

The Freedom® Titan Tibial Augments are fabricated from Titanium-Aluminum-Vanadium (Ti-6A1-4V), complaint with ASTM F136, and are coated with Titanium Niobium Nitride (TiNbN). These augments are intended to be screwed to the distal surface of the previously cleared Stemmed Tibial Base Plate (K111785) &amp; Titan Stemmed Tibial Base Plate (K240863) when required in cases of significant bone loss. The augments are designed to be stackable, using a system of pins, bosses and screws, and reversible for use on either the medial or lateral side.

Augment Screws, fabricated from Titanium-6Aluminum-4Vanadium (Ti-6A1-4V), complaint with ASTM F136, and are coated with Titanium Niobium Nitride (TiNbN), are available as an accessory to the system to fix the augments to the Stemmed Tibial Base Plate and other augments.

## Freedom® Titan Offset Junction

The Freedom® Titan Offset Junctions are fabricated from Titanium-Aluminum-Vanadium (Ti-6A1-4V), complaint with ASTM F136, and are coated with Titanium Niobium Nitride (TiNbN). Offset Junctions are intended to be used with the previously cleared Stemmed Tibial Base Plate (K111785) &amp; Titan Stemmed Tibial Base Plate (K240863) to provide an additional 4mm or 6mm offset between the tibial keel and stem. The Offset Junction is attached to the tibial component and stem through taper junctions. A set-screw is supplied with the Offset Junction to provide additional locking during extraction.

## Freedom® Titan Stem Extension

The Freedom® Titan Stem Extensions are fabricated from Titanium-Aluminum-Vanadium (Ti-6A1-4V), complaint with ASTM F136, and are coated with Titanium Niobium Nitride (TiNbN). The Titan Stem Extensions are intended for use with the previously cleared Stemmed Tibial Base Plate (K111785) &amp; Titan Stemmed Tibial Base Plate (K240863). The Stem Extensions are

{8}

maxx orthopedics
510(k) SUMMARY
K253144 Page 5 of 7

available in a range of diameters and lengths and are fluted distally with a distal slot in the larger lengths and diameters.

## VI. INDICATIONS FOR USE

The Freedom® Total Knee System is indicated for patients with severe knee pain and the disability due to:

- Severe knee joint pain, loss of mobility, and disability due to: rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis.
- Correction of functional deformities.
- Post-traumatic loss of knee joint contour, particularly when there is patellofemoral erosion, dysfunction, or prior patellectomy.
- Moderate valgus, varus, or flexion trauma.
- Knee fractures untreatable by other methods.
- Revision surgery where sufficient bone stock and soft tissue integrity are present. (For PCK Components and Primary PCK Components only)

The Freedom® Total Knee System – Titan PCK Components are intended for cemented and single use only.

## VII. PERFORMANCE TESTING

The presence of TiNbN coating is the only change in this submission. To evaluate the device function and performance of the coating for its intended use, the coated components as well as representative samples with the TiNbN Coating was subjected to the following coating tests:

- Coating Thickness
- Coating Hardness
- Coating Adhesion Strength

Additionally, the below listed mechanical and functional tests are leveraged from the testing performed on Freedom® PCK Components (K131481) and Freedom® Stemmed Tibial

{9}

maxx
orthopedics
510(k) SUMMARY
K253144 Page 6 of 7

Components (K111785). The cleared uncoated devices are the part of Freedom® Total Knee System and are identical to the Titan Femoral Augment, Titan Tibial Augment, Titan Stem Extension and Titan Offset Junction except for the TiNbN coating on the surface. However, the TiNbN coating does not have any effect on these mechanical and functional tests. Therefore, the testing performed on the cleared uncoated Freedom® PCK Components and Freedom Stemmed Tibial Components can be leveraged for the subject device.

- Static and Dynamic Properties of the Freedom PCK Tibial Post
- Modular Disassembly Characteristics of the Freedom PCK Tibial Insert
- Determination of the Range of Motion of the Freedom PCK
- Stability Characteristics of the Freedom PCK Components
- Stability Characteristics of the Freedom PCK Components at High Flexion
- Axial Disassembly Properties of the Freedom Modular Stemmed Tibial Component
- Disassembly Properties of the Freedom Modular Tibial Augment
- Modular Augment Damage Analysis of the Freedom Knee Stemmed Tibial Tray Assembly
- Modular Taper Analysis of the Freedom Knee System Stemmed Tibial Tray

## VIII. BIOCOMPATIBILITY

The biocompatibility tests are leveraged from the testing performed on Freedom® TiNbN Coated Knee (K200912). The cleared K200912 device has the same coating as the subject devices. Therefore, the biocompatibility testing performed on the cleared Freedom® TiNbN Coated Knee can be leveraged for the subject device.

## IX. SUBSTANTIAL EQUIVALENCE DISCUSSION

There are no significant technological differences between the subject and predicate device. The subject devices Freedom® Total Knee System – Titan PCK are same as the predicate devices in terms of fundamental scientific technology; design and dimensions, geometry and sizing,

{10}

maxx
orthopedics
510(k) SUMMARY
K253144 Page 7 of 7

device's operating mechanism, base material composition, and intended use and indications for use.

The presence of TiNbN coating is the only change. The coating is equivalent to the previously cleared Freedom® TiNbN Coated Knee (K200912) and Freedom® Total Knee System – Titan Stemmed PCK Femoral Component &amp; Titan Stemmed Tibial Base Plate (K240863). Based on the results from design hazard analysis and performance testing, it is concluded that the device modification does not raise different questions on the device's safety and effectiveness. The subject device is substantially equivalent to the proposed predicate devices.

## X. CONCLUSION

Based on the intended use, the indications for use and fundamental scientific technology supported by performance and biocompatible testing, it is concluded that the subject device is substantially equivalent to the legally marketed predicate devices.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JWH/K253144](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JWH/K253144)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
