EMPOWR Revision Symmetric Knee Cones

K240324 · Encore Medical, L.P. (Dba Enovis) · JWH · Apr 26, 2024 · Orthopedic

Device Facts

Record IDK240324
Device NameEMPOWR Revision Symmetric Knee Cones
ApplicantEncore Medical, L.P. (Dba Enovis)
Product CodeJWH · Orthopedic
Decision DateApr 26, 2024
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3560
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The EMPOWR Revision Knee™ Symmetric Cones are intended for use in skeletally mature patients with bone defect or poor bone quality (osteoporotic bone) or in case of sclerotic bone that requires supplemental fixation in the clinical judgment of the surgeon. The EMPOWR Revision Knee™ Symmetric Cones are intended for uncemented fixation to the bone and are fixed to the femoral and tibial implants using bone cement.

Device Story

EMPOWR Revision Knee Symmetric Cones are modular, single-use orthopedic accessories for primary or revision total knee arthroplasty. Designed for patients with bone defects or poor bone quality, these cones provide supplemental fixation in the distal femur or proximal tibia. The device is implanted uncemented to the bone while being fixed to femoral or tibial implants using bone cement. Used by orthopedic surgeons in clinical/OR settings, the cones serve as structural support to address bone loss. They are compatible with existing EMPOWR knee components. The device benefits patients by restoring joint stability and configuration in complex revision or primary cases where standard fixation is insufficient.

Clinical Evidence

No clinical data submitted. Substantial equivalence is supported by bench testing, including dynamic fatigue testing (ASTM F1800-19e1) and MR conditional labeling evaluation (ASTM F2052-21, F2213-17, F2119-07, F2182-10e2, F2503-20).

Technological Characteristics

Materials: Ti6Al4V alloy (ASTM F1472) with porous surface of commercially pure titanium (ASTM F67). Principle: Modular orthopedic accessory for supplemental fixation in knee arthroplasty. Fixation: Uncemented to bone, cemented to femoral/tibial implants. Sterilization: Sterile, single-use. Connectivity: None. Software: None.

Indications for Use

Indicated for skeletally mature patients requiring total joint replacement due to degenerative, post-traumatic, or rheumatoid arthritis; avascular necrosis of the femoral condyle; post-traumatic loss of joint configuration (patellofemoral erosion, dysfunction, or prior patellectomy); or moderate valgus, varus, or flexion deformities. Also indicated for salvage of failed surgical attempts where bone loss requires augments, sleeves, cones, or extensions, specifically for patients with bone defects, poor bone quality (osteoporotic), or sclerotic bone requiring supplemental fixation.

Regulatory Classification

Identification

A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. April 26, 2024 Encore Medical, L.P. (dba Enovis) Trey Thorsen Regulatory Consultant 9800 Metric Boulevard Austin, Texas 78758 Re: K240324 Trade/Device Name: EMPOWR Revision Symmetric Knee Cones Regulation Number: 21 CFR 888.3560 Regulation Name: Knee Joint Patellofemorotibial Polymer Semi-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: JWH. MBH Dated: February 1, 2024 Received: February 2, 2024 Dear Trey Thorsen: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" {1}------------------------------------------------ (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). > Sincerely. Peter G. Allen Digitally signed by Peter G. > Allen -s Date: 2024.04.26 11:12:48 နာ -04'00 For Lixin Liu, Ph.D. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health {2}------------------------------------------------ # Indications for Use Submission Number (if known) K240324 Device Name EMPOWR Revision Symmetric Knee Cones Indications for Use (Describe) Total joint replacement is indicated for patients suffering from disability due to: -degenerative, post-traumatic or rheumatoid arthritis; -avascular necrosis of the femoral condyle; -post-traumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy; -moderate valgus, varus or flexion deformities; This device may also be indicated in the salvage of previously failed surgical attempts where bone loss may require the use of augments, sleeves cones or extensions. The EMPOWR Revision Knee™ Symmetric Cones are indicated for the following conditions: EMPOWR™ Revision Knee™ Symmetric Cones are indicated for use in skeletally mature patients with bone defect or poor bone quality (osteoporotic bone) or in case of sclerotic bone that requires supplemental fixation in the clinical judgment of the surgeon. EMPOWR™ Revision Knee™ Symmetric Cones are indicated for uncemented fixation to the bone and are fixed to the femoral and tibial implants using bone cement. Type of Use (Select one or both, as applicable) < | Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) # CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. # *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for "enovis". The first two letters, "en", are in red, followed by a red and black circular design. The last three letters, "ovis", are in black. #### 510(k) Summary #### I. SUBMITTER Encore Medical, L.P. 9800 Metric Blvd. Austin, TX 78758 Contact Person: Trey Thorsen Email: trey.thorsen@enovis.com Phone: 850-450-3932 # II. DEVICE Name of Device: EMPOWR Revision Knee™ Symmetric Cones Common or Usual Name: Total Knee Implant Classification Name: Knee joint patellofemorotibial polymer semiconstrained cemented prosthesis > Knee joint patellofemorotibial metal/polymer porous-coated uncemented prosthesis. Regulation: 21 CFR 888.3560 / 21 CFR 888.3565 Regulatory Class: II Primary Product Code: JWH Secondary Product Code: MBH # III. PREDICATE / REFERENCE DEVICES Predicate: AMF Revision TT Cones, (K200653) Reference: LINK TrabecuLink® Femoral Cones, (K201364) 3D Metal Diaphyseal Femoral Cones, (K200075) # IV. DEVICE DESCRIPTION Envois is adding a non-mating modular component accessory to the EMPOWR™ Knee product line and is compatible with the EMPOWR Revision Knee™ Femur (K213793), the EMPOWR PS Knee® Femur (K160342), the EMPOWR ™ Universal Tibial Baseplate (K173723), and the EMPOWR™ Knee Cemented Stems and Extender Accessories (K173723, K213793, and K230441). The EMPOWR Revision Knee™ Symmetric Cone implants (also known as Cones) are an optional accessory in primary or revision Total Knee Arthroplasty. The Cones are sterile, single-use device that are compatible for use with the EMPOWR Revision Knee™ components. The EMPOWR™ Cones are composed of Ti6Al4V alloy per ASTM F1472 and contain a modified surface composed of commercially pure titanium per ASTM F67. This modified surface (i.e. porous coating) is exactly the same as applied to existing FMP™ Porous Coated Acetabular Shells (K072888). {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for "enovis". The first two letters, "e" and "n", are in red. The "o" is a stylized black and gray circle. The last three letters, "vis", are in black. The EMPOWR Revision Knee™ Symmetric Cones are intended to the mating femoral and/or tibial component using bone cement. The cones are intended for fixation as an assembled construct in the distal femur and/or proximal tibia, without bone cement. # V. INDICATIONS FOR USE / INTENDED USE # Intended Use The EMPOWR Revision Knee™ Symmetric Cones are intended for use in skeletally mature patients with bone defect or poor bone quality (osteoporotic bone) or in case of sclerotic bone that requires supplemental fixation in the clinical judgment of the surgeon. The EMPOWR Revision Knee™ Symmetric Cones are intended for uncemented fixation to the bone and are fixed to the femoral and tibial implants using bone cement. #### Indications for Use The EMPOWR Revision Knee™ Symmetric Cone implants (also known as Cones) are an optional accessory in primary or revision Total Knee Arthroplasty. The Cones are sterile, single-use device that are compatible for use with the EMPOWR Revision Knee™ components. The EMPOWR Revision Knee™ : Total joint replacement indicated for patients suffering from disability due to: - . degenerative, post-traumatic or rheumatoid arthritis; - avascular necrosis of the femoral condyle; - post-traumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy; - . moderate valgus, varus or flexion deformities; This device may also be indicated in the salvage of previously failed surgical attempts where bone loss may require the use of augments, sleeves cones or extensions. The EMPOWR Revision Knee™ Symmetric Cones are indicated for the following conditions: - . EMPOWR™ Revision Knee™ Symmetric Cones are indicated for use in skeletally mature patients with bone defect or poor bone quality (osteoporotic bone) or in case of sclerotic bone that requires supplemental fixation in the clinical judgment of the surgeon. - EMPOWR™ Revision Knee™ Symmetric Cones are indicated for uncemented fixation to the bone and are fixed to the femoral and tibial implants using bone cement. {5}------------------------------------------------ Image /page/5/Picture/1 description: The image shows the logo for "enovis". The word "eno" is in red, followed by a stylized "o" that is black and red. The letters "vis" are in black. There is a trademark symbol in the lower right corner of the logo. # VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE The EMPOWR Revision Knee™ Symmetric Cones are an optional accessory in primary or revision Total Knee Arthroplasty. Comparative testing demonstrates substantial equivalence between the subject and predicate device. #### Biocompatibility testing The biocompatibility evaluation for the EMPOWR Cones was conducted in accordance with the FDA guidance, Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing, May 1, 1995, and International Standard ISO 10993-1 Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process, as recognized by FDA. The EMPOWR Revision Knee™ Symmetric Cones are intended for permanent implantation, contacting tissue/bone. Biocompatibility was ensured through the use of qualified materials and contact agents. #### Performance Testing The following testing was performed to FDA recognized standards and internal protocols: - . Dynamic Fatigue Testing: ASTM F1800-19e1 (modified) - . MR Conditional Labeling: ASTM F2052 -21, ASTM F2213-17, ASTM F2119-07(2013), ASTM F2182-10e2, ASTM F2503-20 #### Animal Studies No animal data submitted. #### Clinical Studies No clinical data submitted. #### VIII. CONCLUSIONS All testing and evaluations demonstrate that the subject device is substantially equivalent to the predicate.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...