← Product Code [JWH](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JWH) · K230846

# iTotal Identity Knee Replacement System; Identity Imprint Knee Replacement System (K230846)

_Conformis, Inc. · JWH · Jul 17, 2023 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JWH/K230846

## Device Facts

- **Applicant:** Conformis, Inc.
- **Product Code:** [JWH](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JWH.md)
- **Decision Date:** Jul 17, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3560
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The iTotal CR Knee Replacement System (KRS) is intended for use as a total knee replacement in patients with knee joint pain and disability whose conditions cannot be solely addressed by the use of a prosthetic device that treats only one or two of the three knee compartments, such as a unicondylar, patellofemoral or bicompartmental prosthesis. The Indications for Use include: Painful joint disease due to osteoarthritis, traumatoid arthritis, rheumatoid arthritis or osteonecrosis of the knee. Post traumatic loss of joint function. Moderate varus, valgus or flexion deformity in which the ligamentous structures can be returned to adequate function and stability. Failed osteotomies, hemiarthroplasties, and unicondylar, patellofemoral or bicompartmental implants. Revision procedures provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging scans. This implant is intended for cemented use only. The CS (cruciate sacrificing) insert option should be utilized when additional anterior-posterior constraint is desired.

## Device Story

Total knee replacement system consisting of femoral, tibial, and patellar components. iTotal Identity uses patient-specific implants designed from CT scans; Identity Imprint uses standard-sized implants selected via patient imaging. Components include CoCrMo femoral implants, titanium alloy tibial trays, and UHMWPE inserts/patellar components. System includes single-use, patient-specific surgical instruments for intraoperative positioning and bone cutting, plus reusable manual instruments. Used by orthopedic surgeons in clinical settings to treat severe knee pain/disability. Software updates incorporated to improve manufacturing efficiencies. Implants provide structural replacement of knee joint compartments to restore function and reduce pain.

## Clinical Evidence

Bench testing only. Verification and validation testing confirmed software performance met or exceeded predicate performance.

## Technological Characteristics

Materials: CoCrMo alloy (femoral), titanium alloy (tibial tray), UHMWPE (inserts/patellar). Semi-constrained cemented prosthesis. Includes patient-specific single-use instruments and reusable manual instruments. Software updates for manufacturing efficiency.

## Regulatory Identification

A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).

## Predicate Devices

- iTotal Identity Cruciate Retaining (CR) Knee Replacement System (KRS) ([K210809](/device/K210809.md))
- iTotal Identity Posterior Stabilizing (PS) Knee Replacement System (KRS) ([K210252](/device/K210252.md))
- Identity Imprint Cruciate Retaining (CR) Knee Replacement System (KRS) ([K221404](/device/K221404.md))
- Identity Imprint Posterior Stabilizing (PS) Knee Replacement System (KRS) ([K221404](/device/K221404.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a smaller font.

July 17, 2023

Conformis, Inc. Nancv Giezen Manager Regulatory Affairs 600 Technology Park Dr. Billerica, Massachusetts 01830

## Re: K230846

Trade/Device Name: iTotal Identity Cruciate Retaining (CR) Knee Replacement System (KRS); iTotal Identity Posterior Stabilizing (PS) Knee Replacement System (KRS); Identity Imprint Cruciate Retaining (CR) Knee Replacement System (KRS); Identity Imprint Posterior Stabilizing (PS) Knee Replacement System (KRS) Regulation Number: 21 CFR 888.3560 Regulation Name: Knee Joint Patellofemorotibial Polymer Semi-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: JWH, OIY, OOG Dated: March 27, 2023 Received: June 15, 2023

Dear Nancy Giezen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of 

{1}------------------------------------------------

Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Lixin Liu -S

Lixin Liu, Ph.D. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K230846

#### Device Name

iTotal Identity Cruciate Retaining (CR) Knee Replacement System (KRS), iTotal Identity Posterior Stabilizing (PS) Knee Replacement System (KRS), Identity Imprint Cruciate Retaining (CR) Knee Replacement System (KRS), Identity Imprint Posterior Stabilizing (PS) Knee Replacement System (KRS)

#### Indications for Use (Describe)

iTotal Identity Cruciate Retaining (CR) Knee Replacement System (KRS)

The iTotal CR Knee Replacement System (KRS) is intended for use as a total knee replacement in patients with knee jount pain and disability whose conditions cannot be solely addressed by the use of a prosthetic device that treats only one or two of the three knee compartments, such as a unicondylar, patellofemoral or bicompartmental prosthesis. The Indications for Use include:

Painful joint disease due to osteoarthritis, traumatoid arthritis or osteonecrosis of the knee. Post traumatic loss of joint function.

Moderate varus, valgus or flexion deformity in which the ligamentous structures can be returned to adequate function and stability.

Failed osteotomies, hemiarthroplasties, and unicondylar, patellofemoral or bicompartmental implants.

Revision procedures provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging scans

This implant is intended for cemented use only.

The CS (cruciate sacrificing) insert option should be utilized when additional anterior-posterior constraint is desired.

iTotal Identity Posterior Stabilizing (PS) Knee Replacement System (KRS)

The iTotal Identity Posterior Stabilized Knee Replacement System is intended for use as a total knee replacement in patients with knee joint pain and disability whose conditions cannot be solely addressed by the use of a prosthetic device that treats only one or two of the three compartments, such as a unicondylar, patellofemoral or bicompartmental prosthesis.

The Indications for Use include:

Pamful joint disease due to osteoarthritis, traumatoid arthritis, polyarthritis, osteonerosis of the knee Post-traumatic loss of joint function

Moderate varus, valgus, or flexion deformity

Failed osteotomies, hemiarthroplasties, and unicondylar, patellofemoral or bicompartmental implants

Revision procedures provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging scans

This implant is intended for cemented use only.

Identity Imprint Cruciate Retaining (CR) Knee Replacement System (KRS)

The Identity™ Imprint™ CR Total Knee Replacement System is intended for use as a total knee replacement in patients with knee joint pain and disability whose conditions cannot be solely addressed by the use of a prosthetic device that treats only one or two of the three knee compartments, such as a unicondylar, patellofemoral or bicompartmental prosthesis. The Indications for Use include:

Painful joint disease due to osteoarthritis, traumatic arthritis or osteonecrosis of the knee.

Post traumatic loss of joint function.

Moderate varus, valgus or flexion deformity in which the ligamentous structures can be returned to adequate function and stability.

Failed osteotomies, hemiarthroplasties, and unicondylar, patellofemoral or bicompartmental implants.

Revision procedures provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging scans.

This implant is intended for cemented use only.

{3}------------------------------------------------

Identity Imprint Posterior Stabilizing (PS) Knee Replacement System (KRS)

The Identity™ Imprint™ PS Total Knee Replacement System is intended for use as a total knee replacement in patients with knee joint pain and disability whose conditions cannot be solely addressed by the use of a prosthetic device that treats only one or two of the three knee compartments, such as a unicondylar, patellofemoral or bicompartmental prosthesis.

The Indications for Use include:

Painful joint disease due to osteoarthritis, traumatic arthritis, rheumatoid arthritis, polyarthritis, osteonecrosis of the knee.

Post-traumatic loss of joint function.

Moderate varus, valgus, or flexion deformity.

Failed osteotomies, hemiarthroplasties, and unicondylar, patellofemoral or bicompartmental implants.

Revision procedures provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging scans.

This implant is intended for cemented use only.

Type of Use (*Select one or both, as applicable*)

| <span style="text-decoration: underline;"></span> <span style="font-size: 14px;"> <svg fill="none" height="16" viewbox="0 0 16 16" width="16" xmlns="http://www.w3.org/2000/svg"> <path d="M13.2222 2.77783H2.77778V13.2223H13.2222V2.77783Z" stroke="black" stroke-width="1.5"></path> <path d="M4.88892 4.88892L11.1111 11.1111" stroke="black" stroke-width="1.5"></path> <path d="M4.88892 11.1111L11.1111 4.88892" stroke="black" stroke-width="1.5"></path> </svg>          Prescription Use (Part 21 CFR 801 Subpart D)</span> | <span style="text-decoration: underline;"></span> <span style="font-size: 14px;"> <svg fill="none" height="16" viewbox="0 0 16 16" width="16" xmlns="http://www.w3.org/2000/svg"> <path d="M13.2222 2.77783H2.77778V13.2223H13.2222V2.77783Z" stroke="black" stroke-width="1.5"></path> </svg>         Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

K230846 Page 1 of 3

Image /page/4/Picture/1 description: The image shows the logo for Conformis. The logo consists of a stylized letter "C" in orange and black, with the word "CONFORMIS" written in black and orange letters below it. The letters "CON" are in black, while the letters "FORMIS" are in orange.

## 510(k) Summary

Prepared on: 2023-07-17

#### Contact Details

| Applicant Name:              | Conformis, Inc.                                          |
|------------------------------|----------------------------------------------------------|
| Applicant Address:           | 600 Technology Park Dr. Billerica MA 01830 United States |
| Applicant Contact Telephone: | (781) 345-9058                                           |
| Applicant Contact:           | Nancy Giezen                                             |
| Applicant Contact Email:     | Nancy.Giezen@conformis.com                               |

#### Device Name

Device Trade Name:

iTotal Identity Cruciate Retaining (CR) Knee Replacement System (KRS) iTotal Identity Posterior Stabilizing (PS) Knee Replacement System (KRS) Identity Imprint Cruciate Retaining (CR) Knee Replacement System (KRS) Identity Imprint Posterior Stabilizing (PS) Knee Replacement System (KRS)

## Common Name:

Knee replacement system

#### Classification Name:

Knee joint patellofemorotibial polymer semi-constrained cemented prosthesis

Regulation Number: 888.3560

Product Code:

JWH, OIY, OOG

#### Legally Marketed Predicate Devices

Predicate #: Predicate Trade Name: Product Codes: iTotal Identity Cruciate Retaining (CR) Knee Replacement System (KRS) JWH. OIY. OOG K210809 iTotal Identity Posterior Stabilizing (PS) Knee Replacement System (KRS) JWH, OIY, OOG K210252 K221404 Identity Imprint Cruciate Retaining (CR) Knee Replacement System (KRS) JWH, OIY, OOG Identity Imprint Posterior Stabilizing (PS) Knee Replacement System (KRS) JWH. OIY, OOG

#### Device Description Summary

The iTotal Identity and Identity Imprint Knee Replacement Systems (KRS) are semi-constrained total knee prosthetic devices consisting of femoral, tibial, and patellar components. These devices are intended for treatment of severe pain and/or disability of a knee damaged by osteoarthritis or trauma.

Patient imaging (CT scans) can be used to design patient-specific implants (iTotal Identity) or to select standard sized implants (Identity Imprint), to meet the geometric and anatomic requirements of the specific patient. The femoral components of the subject devices are manufactured from cobalt chromium molybdenum (CoCrMo) alloy. The tibial component includes a metal tray manufactured from titanium alloy and polyethylene inserts. The patellar components are manufactured from UHMWPE.

For user convenience, single-use, patient-specific surgical instruments are provided to assist in the positioning of the total knee replacement components intraoperatively and in guiding the cutting of bone. In addition, reusable orthopedic manual surgical instruments are provided separately.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for Conformis. The logo consists of a stylized letter "C" in orange and black, with the word "CONFORMIS" written in black and orange letters below it. The first four letters of the word are in black, and the last four letters are in orange.

## Intended Use/Indications For Use

## iTotal Identity Cruciate Retaining (CR) Knee Replacement System (KRS)

The iTotal CR Knee Replacement System (KRS) is intended for use as a total knee replacement in patients with knee joint pain and disability whose conditions cannot be solely addressed by the use of a prosthetic device that treats only one or two of the three knee compartments, such as a unicondylar, patellofemoral or bicompartmental prosthesis.

The Indications for Use include:

Painful joint disease due to osteoarthritis, traumatoid arthritis, rheumatoid arthritis or osteonecrosis of the knee.

Post traumatic loss of joint function.

Moderate varus, valgus or flexion deformity in which the ligamentous structures can be returned to adequate function and stability.

Failed osteotomies, hemiarthroplasties, and unicondylar, patellofemoral or bicompartmental implants. Revision procedures provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging scans

This implant is intended for cemented use only.

The CS (cruciate sacrificing) insert option should be utilized when additional anterior-posterior constraint is desired.

#### iTotal Identity Posterior Stabilizing (PS) Knee Replacement System (KRS)

The iTotal Identity Posterior Stabilized Knee Replacement System is intended for use as a total knee replacement in patients with knee joint pain and disability whose conditions cannot be solely addressed by the use of a prosthetic device that treats only one or two of the three knee compartments, such as a unicondylar, patellofemoral or bicompartmental prosthesis.

The Indications for Use include:

Painful joint disease due to osteoarthritis, traumatic arthritis, rheumatoid arthritis,

osteonecrosis of the knee

Post-traumatic loss of joint function

Moderate varus, valgus, or flexion deformity

Failed osteotomies, hemiarthroplasties, and unicondylar, patellofemoral or bicompartmental implants Revision procedures provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging scans

This implant is intended for cemented use only.

## Identity Imprint Cruciate Retaining (CR) Knee Replacement System (KRS)

The Identity™ Imprint™ CR Total Knee Replacement System is intended for use as a total knee replacement in patients with knee joint pain and disability whose conditions cannot be solely addressed by the use of a prosthetic device that treats only one or two of the three knee compartments, such as a unicondylar, patellofemoral or bicompartmental prosthesis.

The Indications for Use include:

Painful joint disease due to osteoarthritis, traumatoid arthritis, rheumatoid arthritis or osteonecrosis of the

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the ConforMIS logo. The logo features the word "CONFORMIS" in a sans-serif font, with the "FORMIS" portion of the word in orange. Above the word is a stylized "C" shape in orange with a black outline.

knee.

Post traumatic loss of joint function.

Moderate varus, valgus or flexion deformity in which the ligamentous structures can be returned to adequate function and stability.

Failed osteotomies, hemiarthroplasties, and unicondylar, patellofemoral or bicompartmental implants. Revision procedures provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging scans.

This implant is intended for cemented use only.

## Identity Imprint Posterior Stabilizing (PS) Knee Replacement System (KRS)

The Identity™ Imprint™ PS Total Knee Replacement System is intended for use as a total knee replacement in patients with knee joint pain and disability whose conditions cannot be solely addressed by the use of a prosthetic device that treats only one or two of the three knee compartments, such as a unicondylar, patellofemoral or bicompartmental prosthesis.

The Indications for Use include:

Painful joint disease due to osteoarthritis, traumatic arthritis, rheumatoid arthritis, osteonecrosis of the knee.

Post-traumatic loss of joint function.

Moderate varus, valgus, or flexion deformity.

Failed osteotomies, hemiarthroplasties, and unicondylar, patellofemoral or bicompartmental implants. Revision procedures provided that anatomic landmarks necessary for alignment and positioning of the implant are identifiable on patient imaging scans.

This implant is intended for cemented use only.

## Indications For Use Comparison

The indications for use are the same as compared with the predicate devices.

## Technological Comparison

The subject devices which include the implants, single-use patient-specific instruments and reusable instruments remain identical to the predicate devices in operating principle, fundamental technology, design and materials. The purpose of this submission is to incorporate software updates to improve manufacturing efficiencies.

## Non-Clinical and/or Clinical Test Summary and Conclusions

The results of verification and validation testing confirmed that the software performed as well as or better than the predicates. No different questions of safety or effectiveness were raised. Thus the subject devices are considered substantially equivalent to the predicate devices.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JWH/K230846](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JWH/K230846)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
