K211877 · Total Joint Orthopedics, Inc. · JWH · Jun 22, 2022 · Orthopedic
Device Facts
Record ID
K211877
Device Name
Klassic Knee System
Applicant
Total Joint Orthopedics, Inc.
Product Code
JWH · Orthopedic
Decision Date
Jun 22, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3560
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Klassic Knee System is intended for prosthetic replacement in treatment of the following: · Patient conditions of non-inflammatory degenerative joint disease (NIDJD): avascular necrosis and osteoarthritis - · Patient conditions of inflammatory joint disease (IJD): rheumatoid arthritis - · Patients with failed previous surgery where pain, deformity, or dysfunction persists - · Correctable varus-valgus deformity and moderate flexion contracture - · Revision of a previously failed knee arthroplasty - · Patients who require a total knee replacement The Klassic Knee System is indicated for cemented use only, except for the Klassic Femur with Cobalt 3D™, and the Klassic Tibial Baseplate with Ti-Coat®, which are indicated for cementless use.
Device Story
Klassic Knee System provides prosthetic knee replacement components; includes CR/Congruent and Ultra-PS tibial inserts fabricated from E^X Poly (crosslinked UHMWPE with alpha-tocopherol). Device components implanted by orthopedic surgeons during total knee arthroplasty to restore joint function and alleviate pain in patients with degenerative or inflammatory joint disease. System utilizes cemented fixation for standard components; cementless fixation options available for specific femoral and tibial baseplate configurations. Components available in various sizes/thicknesses to match patient anatomy. Clinical benefit derived from mechanical replacement of damaged joint surfaces, improving mobility and reducing pain. Device performance validated through bench testing of wear, stability, stress distribution, range of motion, and modular disassembly.
Clinical Evidence
Bench testing only. Evaluations included wear testing, tibial-femoral stability, stress distribution, range of motion, and tibial modular disassembly characteristics. Biocompatibility evaluated per FDA guidance; LAL testing performed for orthopedic implant compliance.
Technological Characteristics
Components fabricated from E^X Poly (crosslinked UHMWPE with alpha-tocopherol). System includes metal/polymer semi-constrained cemented and porous-coated uncemented configurations. Modular design with various sizes and thicknesses. Sterilization and specific ASTM standards not explicitly detailed in provided text.
Indications for Use
Indicated for patients requiring total knee replacement due to non-inflammatory degenerative joint disease (avascular necrosis, osteoarthritis), inflammatory joint disease (rheumatoid arthritis), failed previous surgery with persistent pain/deformity/dysfunction, correctable varus-valgus deformity, moderate flexion contracture, or revision of failed knee arthroplasty.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
Predicate Devices
Klassic Knee System Tibial Inserts with E-link Poly (K180159)
K180159 — Klassic® Knee System · Total Joint Orthopedics, Inc. · Mar 9, 2018
K162422 — Klassic Knee System · Total Joint Orthopedics, Inc. · Oct 28, 2016
K153075 — Klassic Knee System · Total Joint Orthopedics, Inc. · Nov 13, 2015
K232414 — Klassic® Tibial Insert, PS-Post, Klassic All Poly Tibia, PS-Post · Total Joint Othopedics, Inc. · Sep 8, 2023
K151266 — Physica CR Knee System · Lima Corporate S.P.A. · Oct 6, 2015
Submission Summary (Full Text)
{0}------------------------------------------------
June 22, 2022
Image /page/0/Picture/1 description: The image contains the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized human figure. To the right is the FDA logo, with the acronym "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Total Joint Orthopedics, Inc. % Holly Rhodes VP, Orthopedic Regulatory Affairs MCRA, LLC 1050 K Street NW. Suite 1000 Washington, District of Columbia 20001
Re: K211877
Trade/Device Name: Klassic Knee System Regulation Number: 21 CFR 888.3560 Regulation Name: Knee Joint Patellofemorotibial Polymer Semi-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: JWH. OIY. MBH Dated: May 19, 2022 Received: May 19, 2022
Dear Holly Rhodes:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Ting Song. PhD Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K211877
Device Name Klassic Knee System
Indications for Use (Describe)
The Klassic Knee System is intended for prosthetic replacement in treatment of the following:
· Patient conditions of non-inflammatory degenerative joint disease (NIDJD): avascular necrosis and osteoarthritis
- · Patient conditions of inflammatory joint disease (IJD): rheumatoid arthritis
- · Patients with failed previous surgery where pain, deformity, or dysfunction persists
- · Correctable varus-valgus deformity and moderate flexion contracture
- · Revision of a previously failed knee arthroplasty
- · Patients who require a total knee replacement
The Klassic Knee System is indicated for cemented use only, except for the Klassic Femur with Cobalt 3D™, and the Klassic Tibial Baseplate with Ti-Coat®, which are indicated for cementless use.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
| Manufacturer: | Total Joint Orthopedics, Inc.<br>1567 E. Stratford Avenue<br>Salt Lake City, UT 84106<br>Phone: 801.486.6070<br>Fax: 801.486.6117 |
|---------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Mr. Chris Weaber<br>Director of Research and Development |
| Prepared By: | MCRA, LLC<br>803 7th Street, NW, Third Floor<br>Washington, DC 20001<br>Phone: 202.552.5800 |
| Date Prepared: | June 21, 2022 |
| Device Trade Name: | Klassic Knee System |
| Device Common Name: | Tibial Inserts |
| Regulations: | 21 CFR 888.3560 – Knee joint patellofemorotibial<br>polymer/metal/polymer semi-constrained cemented<br>21 CFR 888.3565 Knee joint patellofemorotibial metal/polymer<br>porous-coated uncemented prosthesis |
| Classification: | Class II |
| Product Codes: | JWH, OIY, MBH |
## Indications for Use:
The Klassic Knee® System is intended for prosthetic replacement in treatment of the following:
- Patient conditions of non-inflammatory degenerative joint disease (NIDJD): avascular necrosis and osteoarthritis
- . Patient conditions of inflammatory joint disease (IJD): rheumatoid arthritis
- . Patients with failed previous surgery where pain, deformity, or dysfunction persists
- Correctable varus-valgus deformity and moderate flexion contracture
- Revision of a previously failed knee arthroplasty •
- . Patients who require a total knee replacement
The Klassic® Knee System is indicated for cemented use only, except for the Klassic® Femur with Cobalt 3D™, and the Klassic® Tibial Baseplate with Ti-Coat®, which are indicated for cementless use.
{4}------------------------------------------------
#### Device Description:
The purpose of this 510(k) is to add Klassic® Knee CR/Congruent Tibial Inserts and Ultra-PS® Tibial Inserts, which are fabricated with E^X Poly, to the Klassic® Knee System. E^X Poly is a crosslinked UHMWPE containing alpha-tocopherol. The subject Klassic® Knee CR/Congruent and Ultra-PS® Tibial Inserts are available in various sizes and thicknesses to match patient anatomy.
### Predicate Devices:
The subject Klassic® Knee CR/Congruent and Ultra-PS® Tibial Inserts with E^X Poly are substantially equivalent to the primary predicate Klassic® Knee System Tibial Inserts with E-link Poly (K180159) with respect to indications, material, design, and function. Additionally, the submission includes reference devices, E-link (K141972) and Biomet E1 (K100048).
### Substantial Equivalence:
Bench testing and engineering analyses were performed on subject Klassic® Knee CR/Congruent and Ultra-PS® Tibial Inserts Knee Components with E^X Poly to evaluate wear, tibial-femoral stability characteristics, stress distributions and range of motion and tibial modular disassembly characteristics. The results of these analyses and testing indicate that the subject CR/Congruent and Ultra-PS® Tibial Inserts with E^X Poly are substantially equivalent to the predicate components. The biocompatibility of the subject tibial inserts was evaluated in accordance with FDA's Biocompatibility guidance document. Lastly, the subject tibial inserts are in compliance with LAL testing requirements for orthopedic implants.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.