Indicated for use only with bone cement for patients suffering from: Severe knee joint pain, loss of mobility and disability due to: rheumatoid arthritis, osteoarthritis, traumatic arthritis, Correction of functional deformities. Posttraumatic loss of knee joint contour, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy; Moderate valgus, Varus or flexion trauma. Knee fractures untreatable by other methods.
Device Story
Genius / Genuin Total Knee System is a posterior stabilized (PS) knee prosthesis for cemented application. Components include a CoCrMo alloy femoral component, a CoCrMo alloy tibial tray, a UHMWPE tibial insert, and a UHMWPE patellar component. The system replaces articulating surfaces of the distal femur, tibia, and patella to restore joint function. It is used by orthopedic surgeons in a clinical/surgical setting. The device facilitates flexion-extension (0° to 155°) and internal-external rotation (±15° in extension, ±25° in flexion). The tibial insert locks into the tibial tray via a dovetail mechanism. By replacing damaged joint surfaces, the device aims to reduce pain, correct deformities, and improve patient mobility.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by non-clinical bench testing, including cyclic fatigue (ASTM F1800-12), tibial insert endurance (ASTM F277-16), contact pressure/area (ASTM F1672-14, F2083-12), constraint testing (ASTM F1223-14), accelerated aging (ASTM F2003-02), locking mechanism strength (ASTM F1814-15), and biocompatibility (ISO 10993-1:2018).
Technological Characteristics
Materials: CoCrMo alloy (ASTM F75-18, ISO 5832-4:2014) for femoral/tibial components; UHMWPE (ASTM F648-14, ISO 5834-2:2019) for inserts/patella. Design: Posterior stabilized, cemented, semi-constrained. Femoral component: 8 sizes. Tibial tray: 6 sizes, 0.8mm cement pocket. Tibial insert: 8 sizes, 8 thicknesses (7-17mm). Patellar component: 6 sizes. No surface coatings.
Indications for Use
Indicated for patients with severe knee pain, mobility loss, or disability due to rheumatoid, osteoarthritis, or traumatic arthritis; functional deformities; post-traumatic knee contour loss (patellofemoral erosion/dysfunction/prior patellectomy); moderate valgus, varus, or flexion trauma; or knee fractures untreatable by other methods. For use only with bone cement.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
K191192 — SurgTech GENOLL Total Knee System · Surgtech, Inc. · Aug 1, 2019
K142388 — Neo Total Knee System · New Era Orthopedics, LLC · Dec 5, 2014
K180446 — b-ONE Total Knee System · B-One Ortho, Corp. · Oct 12, 2018
K111433 — DEPUY STTUNE (TM) PS KNEE SYSTEM · DePuy Orthopaedics, Inc. · Aug 30, 2011
K021657 — UKNEE TOTAL KNEE SYSTEM · United Orthopedic Corp. · Aug 12, 2002
Submission Summary (Full Text)
{0}------------------------------------------------
May 5, 2021
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Biorad Medisys PVT LTD Sudhakar Saxena General Manager Regulatory Affairs Survey No 48/3 & 48/7, Pashan Sus Road, Sus Village, Taluka Mulshi Pune, Maharashtra 411021 INDIA
Re: K200765
Trade/Device Name: Genius / Genuin Total Knee System Regulation Number: 21 CFR 888.3560 Regulation Name: Knee Joint Patellofemorotibial Polymer Semi-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: JWH Dated: March 26, 2021 Received: April 5, 2021
Dear Sudhakar Saxena:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Ting Song, Ph.D., R.A.C. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K200765
Device Name Genius / Genuin Total Knee System
#### Indications for Use (Describe)
Indicated for use only with bone cement for patients suffering from: Severe knee joint pain, loss of mobility and disability due to: rheumatoid arthritis, osteoarthritis, traumatic arthritis, Correction of functional deformities. Posttraumatic loss of knee joint contour, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy; Moderate valgus, Varus or flexion trauma. Knee fractures untreatable by other methods.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> | <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| SPONSOR | |
|-------------------------|--------------------------------------------------------------------------------------------------------|
| Name: | Biorad Medisys Pvt Ltd. |
| Address: | Survey No 48/3 & 48/7, Pashan Sus Road Sus Village, Taluka<br>Mulshi Pune, India. Maharashtra - 411021 |
| Phone: | +91 020-30912000 |
| Fax: | +91 020-30912050 |
| Name of Contact Person: | Sudhakar Saxena |
| Designation: | General Manager Regulatory |
| Date Prepared: | May 04, 2021 |
## DEVICE
| Name of Device: | Genius / Genuin Total Knee System |
|----------------------|---------------------------------------------------------------------------------------------------------------|
| Common Name: | Total Knee System |
| Classification Name: | Prosthesis, Knee, Patellofemorotibial, Semi-Constrained,<br>Cemented, Polymer/Metal/Polymer (21 CFR 888.3560) |
| Product code: | JWH |
### PREDICATE DEVICE
PFC Sigma Knee System (K182301), manufactured by DePuy Orthopaedics, Inc
{4}------------------------------------------------
## DEVICE DESCRIPTION
The Genius / Genuin Total Knee System has 3 major components:
- Femoral component
- . Tibial component - Includes Tibial Tray and Tibial Insert
- Patellar component
The brief description of each major component is provided below:
Femoral Component: Femoral component is a posterior stabilized (PS) design for femoral prosthesis fabricated from CoCrMo Alloy (Cobalt-Chromium-Molybdenum) conforming to ASTM F75 – 18 & ISO 5832-4:2014. It is intended for cemented application to replace the articulating surface of the distal femur. The articulating surface of femoral component interfaces with the Tibial Insert to facilitate flexion-extension and internal-external movement. Femoral component of Genius / Genuin Total Knee System design has both Left and Right configurations. Each of the Left and Right configurations are available in eight different sizes namely A, B, C, D, E, F, G & H where A corresponds to the smallest femoral component and H corresponds to the largest femoral component. The sizes are based on the anterior/posterior (A/P) and medial/lateral (M/L) dimensions.
The range of motion for the Genius / Genuin Total Knee System - femoral component is designed to range from 00 to 155° of flexion and provides the l/E rotation up to ±15° in extension and up to ±25° in flexion.
Tibial Component: The Tibial component comprises of two parts:
- . Tibial tray
- Tibial Insert
Tibial Tray: Tibial tray comprises of a tibial platform that securely holds the Tibial Insert with a dovetail locking mechanism, and fabricated from CoCrMo (Cobalt-Chromium-Molybdenum) Alloy conforming to ASTM F75 - 18 & ISO 5832-4:2014. It is intended for cemented application. The bottom of the tibial tray component has a 0.8mm cement pocket depth for optimum fixation of the prosthesis to bone. Tibial tray is available in 6 different sizes namely size 1, 2, 3, 4, 6 & 7 wherein 1 corresponds to the smallest tibial size and 7 corresponds to largest tibial size.
{5}------------------------------------------------
## 510(K) SUMMARY
Tibial Insert: Tibial insert is configured to have articulating surfaces over one side on which the articulating surface of the femoral component rolls/moves to achieve required flexionextension and rotational movements. The other side comprises of a dovetail locking (snap-fit) feature that gets securely interlocked with the tibial tray during the surgery. The tibial insert is fabricated from UHMWPE conforming to ASTM F648 – 14 & ISO 5834-2:2019. Tibial insert component is available in 8 sizes and each of the inserts are available with 8 different thicknesses i.e. Thickness-7, 8, 9, 10, 11, 13, 15 and 17mm.
Patella Component: The patella component is a dome shaped prosthesis made of UHMWPE conforming to ASTM F648 – 14 & ISO 5834-2:2019. It has 3 pegs on the opposite side of dome to facilitate better fixation with the resurfaced patellar bone. It is intended for cemented use only. lt is designed to provide conforming contact during normal and high flexion activities. Patellar components are available in 6 sizes (diameters). Each size is available with different thickness.
There is no surface coating on any of the components of the Genius / Genuin Total Knee System.
## INDICATIONS FOR USE
Indicated for use only with bone cement for patients suffering from: Severe knee joint pain, loss of mobility and disability due to: rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis, Correction of functional deformities. Post-traumatic loss of knee joint contour, particularly when there is patellofemoral erosion, dysfunction or prior patellectomy; Moderate valgus, Varus or flexion trauma. Knee fractures untreatable by other methods.
## COMPARISON TO TECHNOLOGY CHARACTERISTICS:
The Genius / Genuin Total Knee System is equivalent to the predicate device with respect to intended use, device design, material and method of sterilization.
All the necessary performance test to determine the mechanical characteristics have been performed on the device i.e. Genius / Genuin Total Knee System.
{6}------------------------------------------------
## PERFORMANCE DATA
## Non-Clinical Performance
The following tests have been conducted on the Genius / Genuin Total Knee Joint Replacement system:
- . Cyclic Fatigue Testing of Metal Tibial Tray Component of Total Knee Joint Replacement (ASTM F1800-12)
- Evaluating Tibial Insert Endurance under High Flexion (220,000 cycles) (ASTM F277-16)
- . Determination of Total Knee Implant Femoral-Patellar Contact Pressure/Area (ASTM F1672- 14)
- . Determination of Total Knee Implant Femoral-Tibial Contact Pressure/Area (ASTM F2083-12)
- Determination of Total Knee Replacement Constraint (ASTM F1223- 14)
- Accelerated Aging of UHMWPE after Gamma Irradiation in AIR (ASTM F2003-02)
- Locking mechanism strength test data; static anterior and posterior shear, static medial and lateral shear, static tensile pull off (ASTM F1814-15)
- 1m cycles testing (F/N curve) for shear resistance of tibial post in posterior stabilized tibial bearing (ASTM F1814-15)
- Range of Motion (ASTM F1223-14)
- . UHMWPE Material Property Characterization
- Biocompatibility Testing as per ISO 10993-1:2018
## Clinical Performance Data/Information
Not provided as the device is proven to be substantially equivalent to predicate device.
## CONCLUSION
The Genius / Genuin Total Knee System components are identical to the predicate PFC Sigma Knee System components. Performance data and analyses demonstrate the device is as safe and effective and is substantially equivalent to the legally marketed predicate device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.