← Product Code [JWH](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JWH) · K200653

# AMF Revision TT Cones (K200653)

_Lima Corporate S.P.A. · JWH · May 7, 2020 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JWH/K200653

## Device Facts

- **Applicant:** Lima Corporate S.P.A.
- **Product Code:** [JWH](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JWH.md)
- **Decision Date:** May 7, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3560
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

Physica total knee system is indicated for use in knee arthroplasty in skeletally mature patients with the following conditions: - · Non-inflammatory degenerative joint disease: including osteoarthritis, or avascular necrosis; - · Inflammatory degenerative joint disease including rheumatoid arthritis; - · Correction of functional deformity; - · Revision procedures where other treatments or devices have failed; and - · Treatment of fractures that are unmanageable using other techniques. In patients with preserved and well functioning collateral ligaments. Physica PS components are also for: - · Absent or not-functioning posterior cruciate ligament; - · Severe antero-posterior instability of the knee joint AMF Revision TT Cones are intended for use in skeletally mature patients with bone defect or poor bone quality (osteoporotic bone) or in case of sclerotic bone that requires supplemental metaphysed fixation in the clinical judgment of the surgeon. Tibial and femoral cones are intended fixation to the bone and are fixed to the femoral and tibial implants using bone cement.

## Device Story

AMF Revision TT Cones are accessory components for Total Knee Arthroplasty; used with Physica tibial plates and PS femoral components. Devices are one-piece, conically shaped, cannulated implants. Intended for patients with bone defects or poor bone quality requiring supplemental metaphyseal fixation. Surgeon implants cones into bone (uncemented fixation to bone) and secures them to femoral/tibial implants using bone cement. Provides structural support in revision or complex primary knee arthroplasty; aids in stabilizing implants where bone stock is compromised.

## Clinical Evidence

No clinical data. Substantial equivalence demonstrated via comparison of intended use, materials, technological characteristics, and non-clinical bench testing (fatigue resistance per ASTM F1800).

## Technological Characteristics

One-piece, conically shaped, cannulated metal implants. Designed for uncemented fixation to bone and cemented fixation to femoral/tibial implants. Fatigue resistance tested per ASTM F1800.

## Regulatory Identification

A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).

## Predicate Devices

- Trabecular Metal Knee System Augments ([K053340](/device/K053340.md), [K102896](/device/K102896.md), [K103517](/device/K103517.md))
- Triathlon Tritanium Cone Augments ([K143393](/device/K143393.md))

## Reference Devices

- Physica tibial plate ([K141934](/device/K141934.md), [K151266](/device/K151266.md), [K152008](/device/K152008.md))
- Physica PS femoral component ([K152008](/device/K152008.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

May 7, 2020

Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.

LimaCorporate S.p.A % Lacey Harbour Regulatory Manager Lima USA Inc 2001 NE Green Oaks Blvd. Ste.100 Arlington, Texas 76006

Re: K200653

Trade/Device Name: AMF Revision TT Cones Regulation Number: 21 CFR 888.3560 Regulation Name: Knee Joint Patellofemorotibial Polymer Semi-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: JWH, MBH Dated: January 31, 2020 Received: March 12, 2020

Dear Lacey Harbour:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Ting Song, Ph.D., R.A.C. Acting Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K200653

Device Name AMF Revision TT Cones

Indications for Use (Describe)

Physica total knee system is indicated for use in knee arthroplasty in skeletally mature patients with the following conditions:

- · Non-inflammatory degenerative joint disease: including osteoarthritis, or avascular necrosis;
- · Inflammatory degenerative joint disease including rheumatoid arthritis;
- · Correction of functional deformity;
- · Revision procedures where other treatments or devices have failed; and
- · Treatment of fractures that are unmanageable using other techniques.

In patients with preserved and well functioning collateral ligaments. Physica PS components are also for:

- · Absent or not-functioning posterior cruciate ligament;
- · Severe antero-posterior instability of the knee joint

AMF Revision TT Cones are intended for use in skeletally mature patients with bone defect or poor bone quality (osteoporotic bone) or in case of sclerotic bone that requires supplemental metaphysed fixation in the clinical judgment of the surgeon. Tibial and femoral cones are intended fixation to the bone and are fixed to the femoral and tibial implants using bone cement.

| Type of Use (Select one or both, as applicable)                                                                                                                           |                                                                                                  |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|
| <div><span style="text-decoration: underline;"> </span></div> <div><span style="text-decoration: line-through;"></span>Prescription Use (Part 21 CFR 801 Subpart D)</div> | <span style="text-decoration: line-through;"></span> Over-The-Counter Use (21 CFR 801 Subpart C) |

---

CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# Summary of Safety and Effectiveness

Date: April 30, 2020

Manufacturer: Limacorporate S.p.A. Via Nazionale, 52 33038 - Villanova di San Daniele Udine - Italy

U.S. Contact Person: Lacey Harbour, MB(ASCP)CM US Regulatory Manager Lacey.harbour@limacorporate.com Limacorporate S.p.A. 2001 NE Green Oaks Blvd. Ste.100 Arlington, Texas 76006 USA www.limacorporate.com

Office Phone: 817.385.0777 ext.200 Cell Phone: 432.638.6615 FAX: 817.385.0377

| Product                  | Common<br>Name       | Product<br>Code | Regulation and Classification Name                                                                                      |
|--------------------------|----------------------|-----------------|-------------------------------------------------------------------------------------------------------------------------|
| AMF Revision<br>TT Cones | Total Knee<br>System | JWH             | Knee joint Patellofemorotibial<br>Polymer/Metal/Polymer Semi-<br>Constrained Cemented Prosthesis per 21<br>CFR 888.3560 |
|                          |                      | MBH             | Knee joint Patellofemorotibial<br>metal/polymer porous-coated<br>uncemented prosthesis per 21 CFR<br>888.3565           |

## Description

The AMF Revision TT Cones are intended to be used as an optional accessory component in Total Knee Arthroplasty in combinations with the Physica tibial plate (K141934, K151266 and K152008) and Physica PS femoral component (K152008). The AMF Revision TT Cones are one-piece devices, conically shaped with cannulation all the way through the cone.

#### Indications for use

Physica total knee system is indicated for use in knee arthroplasty in skeletally mature patients with the following conditions:

- Non-inflammatory degenerative joint disease: including osteoarthritis, traumatic arthritis, or avascular necrosis;
- Inflammatory degenerative joint disease including rheumatoid arthritis; ●
- Correction of functional deformity; ●
- Revision procedures where other treatments or devices have failed; and ●
- Treatment of fractures that are unmanageable using other techniques.

Traditional 510(k) – AMF Revision TT Cones April 09, 2020

{4}------------------------------------------------

In patients with preserved and well functioning collateral ligaments, Physica PS components are also for:

- Absent or not-functioning posterior cruciate ligament: .
- Severe antero-posterior instability of the knee joint ●

AMF Revision TT Cones are intended for use in skeletally mature patients with bone defect or poor bone quality (osteoporotic bone) or in case of sclerotic bone that requires supplemental metaphyseal fixation in the clinical judgment of the surgeon. Tibial and femoral cones are intended for uncemented fixation to the bone and are fixed to the femoral and tibial implants using bone cement.

## Predicate Devices

| Company | Device name                              | Cleared via                     |
|---------|------------------------------------------|---------------------------------|
| ZIMMER  | Trabecular Metal Knee System<br>Augments | K053340, K102896 and<br>K103517 |
| STRYKER | Triathlon Tritanium Cone Augments        | K143393.                        |

#### Summary of technology comparison

The intended use, principles of operation, design, materials, sterility and methods of fixation of AMF Revision TT Cones are substantially equivalent to the ones of the predicate devices. Design Control Activities have been successfully completed.

#### Non-clinical testing

The following test were performed on AMF Revision TT Cones:

- Fatigue resistance of the Physica System with AMF Revision TT Tibial Cones ● (Internal protocol derived from ASTM F1800).
## Clinical testing

Clinical testing was not necessary to demonstrate substantial equivalence of the new AMF Revision TT Cones to the predicate devices.

#### Conclusion

Based upon a comparison of intended use, materials, summary of technological characteristics, and preclinical testing, the AMF Revision TT Cones are substantially equivalent to the predicate devices identified in this premarket notification.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JWH/K200653](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JWH/K200653)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
