← Product Code [JWH](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JWH) · K200395

# LOSPA II Knee System (EXULT Knee Replacement System) (K200395)

_Corentec Co., Ltd. · JWH · May 7, 2020 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JWH/K200395

## Device Facts

- **Applicant:** Corentec Co., Ltd.
- **Product Code:** [JWH](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JWH.md)
- **Decision Date:** May 7, 2020
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3560
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

LOSPA II Knee System (EXULT Knee Replacement System) is indicated for the treatment of diseases as follows: · Painful, disabling joint disease of the knee resulting from non-inflammatory degenerative joint disease (including osteoarthritis, traumatic arthritis or avascular necrosis) or rheumatoid arthritis; · Post-traumatic loss of knee joint configuration and function; · Moderate varus, valgus, or flexion deformity in which the ligamentous structures can be returned to adequate function and stability; · Correction of unsuccessful osteotomy, arthrodesis, or failure of previous arthroplasty procedure. LOSPA II Knee System (EXULT Knee Replacement System) is intended for cemented application only.

## Device Story

The LOSPA II Knee System is a total knee joint replacement prosthesis. This submission represents a line extension adding new sizes (#13) and a modified size (#12) to the existing femoral component portfolio (PS and CR types). The device is a mechanical implant used by orthopedic surgeons in a clinical/hospital setting to replace damaged knee joint surfaces. It functions by restoring joint configuration and stability in patients with degenerative or post-traumatic joint disease. The system is intended for cemented application. The device does not involve software or electronic processing; it is a passive orthopedic implant.

## Clinical Evidence

Bench testing only. No clinical data was provided. The system's safety and effectiveness were supported by mechanical testing of worst-case predicate components, including tibial plate fatigue, constraint testing, contact analysis, insert disassembly, tibial insert post shear, wear testing, dislocation/jump distance, and range of motion, conducted in accordance with ASTM F1800, ASTM F1223, ASTM F2083, ASTM F1814, ISO 14243, and ISO 21536.

## Technological Characteristics

Total knee joint replacement prosthesis. Femoral components manufactured from cobalt-chromium-molybdenum alloy (ASTM F75, UNS R30075). System includes PS (Posterior Stabilized) and CR (Cruciate Retaining) types. Supplied sterile. Mechanical, passive implant. No software or electronic components.

## Regulatory Identification

A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).

## Predicate Devices

- LOSPA II Knee System ([K192507](/device/K192507.md))
- LOSPA Knee System ([K130673](/device/K130673.md))
- LOSPA Knee System ([K110404](/device/K110404.md))
- LOSPA Knee System ([K121037](/device/K121037.md))
- LOSPA Knee System ([K160157](/device/K160157.md))
- LOSPA Knee System ([K190402](/device/K190402.md))
- Triathlon Total Knee System ([K173849](/device/K173849.md))
- Triathlon Total Knee System ([K172326](/device/K172326.md))
- Scorpio NRG Knee System ([K042343](/device/K042343.md))
- Scorpio NRG Knee System ([K071991](/device/K071991.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/2 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath in a smaller font.

May 7, 2020

Corentec Co., Ltd % Yoorim Bae Asst. Manager-RA Corentec Co., Ltd. Banpo-daero 20-gil, 33-2, Seocho Gu, Seoul, Seoul 06649 Republic of Korea

Re: K200395

Trade/Device Name: LOSPA II Knee System (EXULT Knee Replacement System) Regulation Number: 21 CFR 888.3560 Regulation Name: Knee Joint Patellofemorotibial Polymer Semi-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: JWH Dated: April 7, 2020 Received: April 7, 2020

Dear Yoorim Bae:

We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Ting Song, Ph.D., R.A.C. Acting Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K200395

Device Name

LOSPA II Knee System (EXULT Knee Replacement System)

Indications for Use (Describe)

LOSPA II Knee System (EXULT Knee Replacement System) is indicated for the treatment of diseases as follows:

· Painful, disabling joint disease of the knee resulting from non-inflammatory degenerative joint disease (including osteoarthritis, traumatic arthritis or avascular necrosis) or rheumatoid arthritis;

· Post-traumatic loss of knee joint configuration and function;

· Moderate varus, valgus, or flexion deformity in which the ligamentous structures can be returned to adequate function and stability;

· Correction of unsuccessful osteotomy, arthrodesis, or failure of previous arthroplasty procedure.

LOSPA II Knee System (EXULT Knee Replacement System) is intended for cemented application only.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(K) SUMMARY Corentec Co.,Ltd. LOSPA II Knee System (EXULT Knee Replacement System) - Specification Inclusion

February 14, 2020

## ADMINISTRATIVE INFORMATION

| Manufacturer:     | Corentec Co., Ltd.                                  |
|-------------------|-----------------------------------------------------|
|                   | 12, Yeongsanhong 1-gil, Ipjang-Myeon, Seobuk-Gu     |
|                   | Cheonan-si, Chungchongnam-do, Rep. of Korea -31056  |
|                   | Telephone: +82-41-585-7114; Fax: +82-41-585-7113    |
| Official Contact: | Sungwon Yang                                        |
|                   | Director – QA&RA                                    |
|                   | 12, Yeongsanhong 1-gil, Ipjang-Myeon, Seobuk-Gu     |
|                   | Cheonan-si, Chungchongnam-do, Rep. of Korea -31056  |
|                   | Ph: +82-41-410-7176 (Direct) ; Fax: +82-41-585-7113 |
|                   | Email: sungwon.yang@corentec.com                    |

#### DEVICE NAME AND CLASSIFICATION

| Trade/Proprietary Name: LOSPA II Knee System (EXULT Knee Replacement System) |
|------------------------------------------------------------------------------|
| Common Name: Total Knee Joint Replacement Prosthesis                         |
| Classification Regulations: 21 CFR 888.3560                                  |
| Class: II                                                                    |
| Product Codes: JWH                                                           |
| Classification Panel: Orthopedic Products Panel                              |
| Reviewing Branch: Orthopedic Devices Branch                                  |

## INDICATIONS FOR USE

The intended use of the added specification has not changed as a result of the modification of the predicate device cleared under LOSPA II Knee System, K192507.

The LOSPA II Knee System (EXULT Knee Replacement System) is indicated for the treatment of diseases as follows:

- o Painful, disabling joint disease of the knee resulting from non-inflammatory degenerative joint disease (including osteoarthritis, traumatic arthritis or avascular necrosis) or rheumatoid arthritis;
- . Post-traumatic loss of knee joint configuration and function;

{4}------------------------------------------------

- o Moderate varus, valgus, or flexion deformity in which the ligamentous structures can be returned to adequate function and stability;
- . Correction or revision of unsuccessful osteotomy, arthrodesis, or failure of previous arthroplasty procedure.

The LOSPA II Knee System (EXULT Knee Replacement System) is intended for cemented application only.

# Legally Marketed Device to Which Substantial Equivalence is Claimed:

The LOSPA II Knee System (EXULT Knee Replacement System) (CR, PS) Femoral Component - K192507 The LOSPA II Knee System (EXULT Knee Replacement System) Instrumentation -K192507, K110404, K130673, K160157

# DEVICE DESCRIPTION

The Additional components being added to the LOSPA II Knee Replacement System (EXULT Knee Replacement System) are:

- Revised Size 12 PS, CR Femoral Component -
- Additional Size 13 PS, CR Femoral Component -
- -Instrumentation

The subject LOSPA II Knee System (EXULT Knee Replacement System) components specification inclusions are a line extension of Femoral Components and Instrumentation system. The following are the additional components,

# A) LOSPA II Femoral Component (PS-type and CR-type)

Femoral Components of LOSPA II Knee System (EXULT Knee Replacement System) are designed based on Femoral Components of LOSPA II Knee System. Femoral component has two different types which match to Tibial Insert types: PS-type and CR-type. There is one additional size, #13 of each design (CR and PS), each for left and right sides, and one modified size #12 (CR and PS, each left and right). For both CR and PS designs, the ranges of dimensions are the same. All femoral components are manufactured from cobalt-chromium-molybdenum alloy conforming to ASTM F75. Standard Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Castings and Casting Alloy for Surgical Implants (UNS R30075)

## B) Instrumentation

There are additional instruments for use with the system implant components.

{5}------------------------------------------------

#### EXULT Knee Replacement System

# SUBSTANTIAL EQUIVALENCE

LOSPA II Knee System is substantially equivalent in indications and design principles to the following predicate devices, each of which has been determined by FDA to be substantially equivalent to pre-amendment devices:

| Device Type              | Manufacturer                                               | Trade or Proprietary<br>or Model Name | 510(k)                                              |
|--------------------------|------------------------------------------------------------|---------------------------------------|-----------------------------------------------------|
| Primary Predicate        | Corentec Co. Ltd.                                          | LOSPA II Knee System                  | K192507                                             |
| Additional<br>Predicates | Corentec Co. Ltd.                                          | LOSPA Knee System                     | K130673<br>K110404<br>K121037<br>K160157<br>K190402 |
|                          | Stryker Orthopaedics<br>(Aka Howmedica<br>Osteonics Corp.) | Triathlon Total Knee System           | K173849<br>K172326                                  |
|                          | Stryker Orthopaedics<br>(Aka Howmedica<br>Osteonics Corp.) | Scorpio NRG Knee System               | K042343<br>K071991                                  |

Substantially equivalent products for LOSPA II Knee System are as follows:

# PERFORMANCE TESTING - BENCH

The LOSPA II Knee System (EXULT Knee Replacement System) additional components were justified by a series of testing requirements conducted on the worst-case predicate components of the system to demonstrate substantial equivalence and included methods described in the following standards: ASTM F1800: ASTM F1223; ASTM F2083: ASTM F1814, ISO 14243 and ISO 21536, which consisted of tibial plate fatigue testing, constraint testing, contact analysis, insert disassembly testing, tibial insert post shear testing, wear testing, dislocation/jump distance and range of motion. Mechanical testing of the additional components was not necessary as the minor size changes are consistent with the previously evaluated predicate device components.

## CONCLUSIONS

The LOSPA II Knee System (EXULT Knee Replacement System) additional components are comparable to predicate devices and should be as safe and effective as predicate devices. Any differences in technological characteristic between the subject and predicate devices do not raise new issues of safety or efficacy. At a high level, the LOSPA II Knee System (EXULT Knee Replacement System) has the following similarities to the predicate devices:

- has the same intended use, O
- has the same indications for use, O
- uses the same operating principles, O
- incorporates the same basic designs, O
- incorporates the same/similar materials, and O
- is supplied sterile O

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JWH/K200395](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JWH/K200395)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
