← Product Code [JWH](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JWH) · K173331

# JOURNEY II XR Knee Instruments (K173331)

_Smith & Nephew, Inc. · JWH · Nov 16, 2017 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JWH/K173331

## Device Facts

- **Applicant:** Smith & Nephew, Inc.
- **Product Code:** [JWH](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JWH.md)
- **Decision Date:** Nov 16, 2017
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3560
- **Device Class:** Class 2
- **Review Panel:** Orthopedic

## Indications for Use

Smith & Nephew Journey II XR Instruments are accessory devices intended to be used to assist in the implantation of Smith & Nephew Journey II XR Knee System and their cleared Indications for Use. Total knee components are indicated for rheumatoid arthritis, osteoarthritis, osteoarthritis, or degenerative arthritis; failed osteotomies, unicompartmental replacement, or total knee replacement. The Journey II XR Knee system components are indicated for use only with cement and are single use devices.

## Device Story

Journey II XR Knee Instruments are surgical accessory tools used by orthopedic surgeons in clinical settings to assist in the implantation of the Journey II XR Knee System. These instruments facilitate bone preparation and component positioning during total knee arthroplasty. The device set includes modified tibial impactors and keel preparation instruments. Surgeons use these tools manually to ensure proper fit and alignment of the cemented knee prosthesis. The instruments are designed to be compatible with existing Smith & Nephew knee systems, ensuring consistent surgical outcomes and mechanical performance during the implantation procedure.

## Clinical Evidence

Bench testing only. No clinical data was required. Testing included biocompatibility assessments per ISO 10993-1 and comparative mechanical performance testing of modified tibial impactors and keel preparation instruments against predicate designs.

## Technological Characteristics

Manual surgical instruments for orthopedic knee arthroplasty. Materials assessed for biocompatibility per ISO 10993-1. Design modifications include updated tibial impactor and keel preparation geometries. Non-powered, mechanical instrumentation.

## Regulatory Identification

A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).

## Predicate Devices

- Smith & Nephew, Inc. Journey II XR Knee system ([K141471](/device/K141471.md))
- Total Knee Instruments ([K121393](/device/K121393.md))

## Reference Devices

- Smith & Nephew, Inc. Journey II XR Knee system ([K152726](/device/K152726.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 16, 2017

Smith & Nephew, Inc. Shereen Bienz Senior Regulatory Affairs Specialist 1450 E. Brooks Road Memphis, Tennessee 38116

### Re: K173331

Trade/Device Name: JOURNEY II XR Knee Instruments Regulation Number: 21 CFR 888.3560 Regulation Name: Knee Joint Patellofemorotibial Polymer Semi-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: JWH Dated: October 20, 2017 Received: October 23, 2017

Dear Shereen Bienz:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);

{1}------------------------------------------------

and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

# Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K173331

Device Name Journey II XR Knee Instruments

#### Indications for Use (Describe)

Smith & Nephew Journey II XR Instruments are accessory devices intended to be used to assist in the implantation of Smith & Nephew Journey II XR Knee System and their cleared Indications for Use.

Total knee components are indicated for rheumatoid arthritis, osteoarthritis, osteoarthritis, or degenerative arthritis; failed osteotomies, unicompartmental replacement, or total knee replacement.

The Journey II XR Knee system components are indicated for use only with cement and are single use devices.

| Type of Use ( <i>Select one or both, as applicable</i> ) |
|----------------------------------------------------------|
|----------------------------------------------------------|

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

| Submitted by:                                | Smith & Nephew, Inc.<br>Orthopaedic Division<br>1450 East Brooks Road<br>Memphis, Tennessee 38116               |
|----------------------------------------------|-----------------------------------------------------------------------------------------------------------------|
| Date of Summary:                             | November 13, 2017                                                                                               |
| Contact Person and Address:                  | Shereen Bienz<br>Senior Regulatory Affairs Specialist<br>T (901) 800-3161<br>F (901)                            |
| Name of Device:                              | Smith & Nephew, Inc. Journey II XR Knee System Instruments                                                      |
| Common Name:                                 | Knee Prosthesis                                                                                                 |
| Device Classification Name and<br>Reference: | 21 CFR 888.3560 Knee joint patellofemorotibial<br>polymer/metal/polymer semi-constrained cemented<br>prosthesis |
| Device Class:                                | Class II                                                                                                        |
| Panel Code:                                  | Orthopaedics/87                                                                                                 |
| Product Code:                                | JWH                                                                                                             |

#### Device Description

Subject of this Special Premarket Notification are modified Journey II XR Instruments. The subject devices are modifications of existing instrumentation cleared for use with the Journey II XR Knee system.

#### Technological Characteristics

Biocompatibility assessments for the instruments were conducted per the recommendations of ISO 10993-1 and the FDA guidance document, Use of International Standard ISO 10993-1, "Biological evaluation of medical device – Part 1: Evaluation and testing within a risk management process, " issued June 16, 2016. A review of the results indicates that the modified knee instruments are equivalent to existing, legally marketed predicate instrumentation with regards to mechanical performance and that there are no new issues related to the safety and effectiveness of the subject devices. Clinical data was not needed to support the safety and effectiveness of the subject devices.

#### Intended Use

Journey II XR Instruments are accessory devices and intended to assist in the implantation of Smith & Nephew Journey II XR Knee systems. Table 1 includes the Smith & Nephew Total Knee systems to be used in conjunction with the subject devices.

| Manufacturer         | Description                                       | Submission<br>Number | Clearance Date |
|----------------------|---------------------------------------------------|----------------------|----------------|
| Smith & Nephew, Inc. | Smith & Nephew, Inc. Journey II<br>XR Knee system | K141471              | 11/14/2014     |
| Smith & Nephew, Inc. | Smith & Nephew, Inc. Journey II<br>XR Knee system | K152726              | 10/21/2015     |

| Table 2: Smith & Nephew Inc. Compatible Total Knee Systems |
|------------------------------------------------------------|
|------------------------------------------------------------|

{4}------------------------------------------------

#### Indications for Use

Smith & Nephew Journey II XR Instruments are accessory devices intended to be used to assist in the implantation of Smith & Nephew Journey II XR Knee System and their cleared Indications for Use.

Total knee components are indicated for rheumatoid arthritis; post-traumatic arthritis, or degenerative arthritis; failed osteotomies, unicompartmental replacement, or total knee replacement.

The Journey II XR Knee system components are indicated for use only with cement and are single use devices.

#### Substantial Equivalence Information

The substantial equivalence of the Journey II XR Instruments is based on its similarities in indications for use, design features, and operational principles to the predicate systems listed in the following table.

#### Table 1: Predicate Devices

| Manufacturer         | Description                                       | Submission<br>Number | Clearance Date |
|----------------------|---------------------------------------------------|----------------------|----------------|
| Smith & Nephew, Inc. | Smith & Nephew, Inc. Journey II<br>XR Knee system | K141471              | 11/14/2014     |
| Smith & Nephew, Inc. | Total Knee Instruments                            | K121393              | 08/08/2012     |

#### Performance and Biocompatibility Testing

The following testing was conducted to evaluate substantial equivalence:

- Biocompatibility risk assessment on the materials was performed in accordance with FDA . guidance, Use of International Standard ISO 10993-1, "Biological evaluation of medical device - Part 1: Evaluation and testing within a risk management process," issued June 16, 2016
- Testing was performed to evaluate impact resistance of the proposed tibial impactor design to ● the previously tested current design
- . Testing was performed to evaluate the performance of the modified keel prep instruments as compared to the predicate keel prep instruments

#### Conclusion

As previously noted, this Special 510|k) Premarket Notification is being submitted to request clearance for modified Journey II XR Instruments. Based on the similarities to the predicate components and a review of the testing performed, the device is substantially equivalent to above predicate systems.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JWH/K173331](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JWH/K173331)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
