← Product Code [JWH](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JWH) · K161360

# U2 Total Knee System, PSA Tibial Insert (K161360)

_United Orthopedic Corporation · JWH · Feb 1, 2017 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JWH/K161360

## Device Facts

- **Applicant:** United Orthopedic Corporation
- **Product Code:** [JWH](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JWH.md)
- **Decision Date:** Feb 1, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3560
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

This device is indicated in knee arthroplasty in skeletally mature patients with severe knee pain and disability due to rheumatoid arthritis, osteoarthritis, primatic arthritis, polyarthritis, collagen disorders, avascular necrosis of the femoral condyle or pseudogout, posttraumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectorny, moderate valgus, varus, or flexion contraction. This device is intended for use in patients who require augmentations due to inadequate bone stock and/or require increased stabilization for tibiofemoral joint due to soft tissue imbalance. The femoral and tibial augments are to be attached to their respective components with a fixation screw or screws. Note: In the US, this device is for cemented use only.

## Device Story

U2 Total Knee System - PSA Tibial Insert is a semi-constrained, cemented knee prosthesis component; part of a patellofemorotibial polymer/metal/polymer system. Device provides stabilization for tibiofemoral joint in patients with soft tissue imbalance or bone stock deficiency. Available in PSA and PSA-LC (low constrained) configurations. Used by orthopedic surgeons in clinical settings during total knee arthroplasty. Device functions as a mechanical implant to restore joint configuration and function. Benefits include pain relief and improved mobility for patients with severe knee pathology.

## Clinical Evidence

No clinical data provided. Substantial equivalence is based on non-clinical bench testing, including range of motion and spine fatigue testing. Wear performance is supported by data from the predicate U2 XPE Tibial Insert (K103733). Bacterial endotoxin testing met USP <161> limits.

## Technological Characteristics

Semi-constrained, cemented knee prosthesis. Materials and locking mechanism are identical to predicate K082424. Available in PSA and PSA-LC configurations. Sterilization method is consistent with predicate devices. No software or electronic components.

## Regulatory Identification

A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).

## Predicate Devices

- UNITED U2 Total Knee System – PSA Type ([K082424](/device/K082424.md))
- UNITED U2 XPE Total Knee System ([K103733](/device/K103733.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 1, 2017

United Orthopedic Corporation Karen Ho Regulatory Affairs Manager No 57, Park Ave 2, Science Park Hsinchu 300 Taiwan

Re: K161360 Trade/Device Name: U2 Total Knee System, PSA Tibial Insert Regulation Number: 21 CFR 888.3560 Regulation Name: Knee Joint Patellofemorotibial Polymer/Metal/Polymer Semi-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: JWH Dated: December 13, 2016 Received: December 14, 2016

Dear Karen Ho:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

### Indications for Use

510(k) Number (if known) K161360

#### Device Name

U2 Total Knee System - PSA Tibial Insert

This device is indicated in knee arthroplasty in skeletally mature patients with severe knee pain and disability due to rheumatoid arthritis, osteoarthritis, primatic arthritis, polyarthritis, collagen disorders, avascular necrosis of the femoral condyle or pseudogout, posttraumatic loss of joint configuration, particularly when there is patellofemoral erosion, dysfunction or prior patellectorny, moderate valgus, varus, or flexion contraction. This device is intended for use in patients who require augmentations due to inadequate bone stock and/or require increased stabilization for tibiofemoral joint due to soft tissue imbalance. The femoral and tibial augments are to be attached to their respective components with a fixation screw or screws.

Note: In the US, this device is for cemented use only.

Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(k) Summary of Safety and Effectiveness

### Submitter Information

| Name                          | United Orthopedic Corporation                                                                                                                                                                                                                                                                         |
|-------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address                       | No 57, Park Ave 2, Science Park, Hsinchu 300, Taiwan                                                                                                                                                                                                                                                  |
| Phone Number                  | +886-3-5773351 ext. 2212                                                                                                                                                                                                                                                                              |
| Fax Number                    | +886-3-577156                                                                                                                                                                                                                                                                                         |
| Name of Contact Person        | Karen Ho<br>Regulation and Document Management                                                                                                                                                                                                                                                        |
| Date prepared                 | May 12, 2016                                                                                                                                                                                                                                                                                          |
| Name of Device                |                                                                                                                                                                                                                                                                                                       |
| Trade Name                    | U2 Total Knee System – PSA Tibial Insert                                                                                                                                                                                                                                                              |
| Common Name                   | Tibial Insert                                                                                                                                                                                                                                                                                         |
| Regulation Name and<br>Number | The device classification for U2 Total Knee System - PSA<br>Tibial Insert is “Knee joint patellofemorotibial<br>polymer/metal/polymer semi-constrained cemented<br>prosthesis” and is contained in the Code of Federal<br>Regulation, under 21CFR 888.3560. This falls under the<br>Orthopedic Panel. |
| Device Class                  | Class II                                                                                                                                                                                                                                                                                              |
| Classification Panel          | Orthopaedics                                                                                                                                                                                                                                                                                          |
| Product Code                  | JWH                                                                                                                                                                                                                                                                                                   |
| Predicate Device              | 1. “UNITED” U2 Total Knee System – PSA Type (K082424)<br>2. “UNITED” U2 XPE Total Knee System (K103733)                                                                                                                                                                                               |

## Device Description:

U2 Total Knee System - PSA Tibial Insert is an extended design of U2 Total Knee System - PSA Type (K082424) including XPE Tibal Insert, PSA and XPE Tibial Insert,

{4}------------------------------------------------

## O U2 Total Knee System - PSA Tibial Insert

PSA, LC. This System is a patellofemorotibial polymer / metal / polymer, semi-constrained, cemented knee prosthesis, which is intended for use in patients who require augmentation and/or stem extensions due to inadequate bone stock. U2 Total Knee System - PSA Tibial Inserts are available in a range of thicknesses and in two design configurations: PSA and PSA-LC type. For the PSA tibial insert, it is intended for use in patients who require constrained stabilization for tibiofemoral joint due to soft tissue imbalance. PSA-LC (low constrained) tibial insert provide less constrained stabilization than PSA tibial insert.

#### Indications for Use:

This device is indicated in knee arthroplasty in skeletally mature patients with severe knee pain and disability due to rheumatoid arthritis, osteoarthritis, primary and secondary traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis of the femoral condyle or pseudogout, posttraumatic loss of joint configuration, particularly when there is patellofemoral joint surface erosion, dysfunction or prior patellectomy, moderate valgus, varus, or flexion contraction. This device is intended for use in patients who require augmentation and/or stem extensions due to inadequate bone stock and/or require increased stabilization for tibiofemoral joint due to soft tissue imbalance. The femoral and tibial augments are to be attached to their respective components with a fixation screw or screws.

Note: In the US, this device is for cemented use only.

#### Comparison to Predicate Device:

U2 Total Knee System - PSA Tibial Insert has the same basic design, intended use, materials, locking mechanism and sterilization as the predicate devices, U2 Total Knee System - PSA Type (K082424) and U2 XPE Total Knee System (K103733). The only difference between the subject and the predicate devices is the profile design of the spine. However, the performance evaluation of the subject device was conducted and would not post issues about safety and effectiveness. Thus, we believe that the subjected U2 Total Knee System - PSA Tibial Insert is substantially equivalent to the predicate device.

{5}------------------------------------------------

#### Performance Data:

#### ● Non-clinical Performance

Range of Motion and Spine Fatigue Test were conducted to evaluate the safety and effectiveness of the subjected device.

The subject U2 Total Knee System – PSA Tibial Insert has the same curvature design at the articular surface as U2 Total Knee system - PSA Type which was previously cleared by FDA (K082424). Thus, the constraint, contact area and contact pressure of the subject PSA Tibial Insert should be identical with U2 Total Knee system – PSA Type (K082424). In addition, the material, size distribution and the locking mechanism of the subject PSA-LC Tibial Insert is identical to the on-marketed PSA insert (K082424). The wear performance could be represented by the on-marketed U2 XPE Tibial Insert (K103733) due to the same contact area and size distribution.

Bacterial endotoxin testing was conducted and met the endotoxin limit as specified in USP <161>.

### ● Clinical Performance Data/Information

None provided as a basis for substantial equivalence.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JWH/K161360](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JWH/K161360)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
