MONOTUBE TUBE TO RING CLAMP
Device Facts
| Record ID | K961618 |
|---|---|
| Device Name | MONOTUBE TUBE TO RING CLAMP |
| Applicant | Howmedica Corp. |
| Product Code | JEC · Orthopedic |
| Decision Date | May 28, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 888.3040 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Indications for Use
This device is intended to provide stabilization of open and/or unstable fractures and where soft tissue injury precludes the use of other fracture treatments such as IM rodding or casting.
Device Story
External fixation clamp; connects Monticelli/Spinelli circular frame rings to Monotube external fixation system tubes. Enables angulation between circular frame and tube components. Used in orthopedic surgery for fracture stabilization. Physician-operated in clinical/OR settings. Provides hybrid construct combining pin fixation with ring/wire fixation.
Clinical Evidence
Bench testing only. Performance comparison of Monotube Hybrid Frame versus Monticelli/Spinelli Hybrid Frame using Hoffmann External Fixation System components.
Technological Characteristics
Materials: Stainless steel and aluminum. Function: Mechanical fixation clamp for external frame systems. Design: Articulating coupling for angulation between circular rings and fixation tubes.
Indications for Use
Indicated for patients with open and/or unstable fractures where soft tissue injury precludes alternative treatments like IM rodding or casting.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
- Howmedica Monticelli/Spinelli Sliding Clamp (with Articulating Coupling)
- Synthes Ring-to-Rod Clamp
- Zimmer Torus Ring-to-Rod Clamp
Reference Devices
- Monotube External Fixation System
- Monticelli/Spinelli Circular External Fixation System
- Hoffmann External Fixation System
Related Devices
- K960782 — TITANIUM TUBE TO TUBE CLAMP · Synthes (Usa) · May 8, 1996
- K011948 — HOFFMAN II HYBRID RING CLAMP · Howmedica Osteonics Corp. · Jul 13, 2001
- K960269 — TITANIUM UNIVERSAL CLAMP · Synthes (Usa) · Apr 4, 1996
- K111376 — CONSTRUX EXTERNAL FIXATION DEVICE · Ebi, LLC · Aug 3, 2011
- K960014 — PRACTIFIX EXTERNAL FIXATION SYSTEM · Orthopaedic Innovations, Inc. · Mar 18, 1996