← Product Code [JEC](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JEC) · K082679

# LAKI HAND FIXATION SYSTEM, MODEL TFX.3X.06.X (K082679)

_Sintea Biotech, Inc. · JEC · Apr 7, 2009 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JEC/K082679

## Device Facts

- **Applicant:** Sintea Biotech, Inc.
- **Product Code:** [JEC](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JEC.md)
- **Decision Date:** Apr 7, 2009
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3040
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The LAKI device is intended to be used for fixation of bone fractures of long bones in fingers. The LAKI device is a semi-invasive device intended to be attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Device Story

LAKI Hand Fixation System is an external fixation device for finger long bone fractures, pseudo-arthrosis, and limb lengthening. System comprises modular elements, rods, nuts, and Kirschner wires for stabilization. Invasive components constructed from 316L stainless steel; external locking connectors from Ti6AL4V titanium alloy. Device attaches through skin to apply traction to skeletal system. Operated by clinicians in clinical settings to facilitate bone re-growth. Output is mechanical stabilization of bone segments.

## Clinical Evidence

No clinical data provided; performance data refers to predicate device K880282.

## Technological Characteristics

Materials: 316L stainless steel (invasive components), Ti6AL4V titanium alloy (external locking connectors). Principle: Mechanical external fixation via modular rods, nuts, and Kirschner wires. Form factor: Modular external stabilizer.

## Regulatory Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- LIMA SEM Modular External Stabilizer ([K880282](/device/K880282.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## 11 K082679

## LAKI Hand Fixation System 510(k) Summary June 2008

I. Sintea Biotech, Inc. Company: 407 Lincoln Rd. Suite 10L Miami Beach, FL 33139 (305) 673-6226

APR - 7 2009

. II. Proprietary Trade Name: LAKI Hand Fixation System

> Regulation Number: 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener. Product Code: . JEC

#### III. Product Description

This 510(k) consists on the submission of an evolution of an external fixation device. The LAKI Hand Fixation System provides opportunity to patients to develop solid bone re-growth in cases of fractures, pseudo-arthrosis, and limb lengthening by use of modular elements, rods, and nuts. The LAKI utilizes a threaded rod bound to two semi modular elements with nuts which allows the former to be locked in the most suitable position for each specific case. Kirschner wires are also used for stabilization. All invasive components of the LAKI Hand Fixation System are made of 316L stainless steel. External locking connector components are all made of Ti6AL4V titanium alloy.

#### IV. Indications

-

The LAKI device is intended to be used for fixation of bone fractures of long bones in fingers. The LAKI device is a semi-invasive device intended to be attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

#### V. Performance Data

Please see 510(k) submission for the SEM Modular External Stabilizer, K880282.

### VI. Substantial Equivalence

Sintea Biotech, Inc. believes that the LAKI Hand Fixation System is substantially equivalent to the LIMA SEM Modular External Stabilizer (K880282) with respect to functional design, indications for use, and principles of operation, performance, and materials.

{1}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/1/Picture/1 description: The image shows the logo for the Department of Health & Human Services, USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" arranged around the perimeter. Inside the circle is a stylized eagle emblem with outstretched wings.

#### Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

# APR - 7 2009

Sintea Biotech, Inc. % Ms. Danielle Wernikowski 407 Lincoln Road, Suite 10L Miami Beach, Florida 33139

Re: K082679

Trade/Device Name: LAKI Hand Fixation System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: II Product Code: JEC Dated: March 12, 2009 Received: March 12, 2009

Dear Ms. Wernikowski:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{2}------------------------------------------------

## Page 2 - Ms. Danielle Wernikowski

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Verlaug Buell MD

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# Indications for Use

510(k) Number (if known): K082679

Device Name: LAKI Hand Fixation System

Indications for Use:

The LAKI device is intended to be used for fixation of bone fractures of long bones in fingers. The LAKI device is a semi-invasive device intended to be attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

Prescription Use _____________________________________________________________________________________________________________________________________________________________

AND/OR

Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________

(Part 21 CFR 801 Subpart D)

(21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Oarbaue Buent
(Division Sign-Off)

Division of General, Restorative. and Neurological Devices

**510(k) Number** K082679

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JEC/K082679](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JEC/K082679)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
