The MedicineLodge, Inc. (MLI) - Cross Pin is used to secure soft tissue graft fixations during cruciate ligament reconstruction surgeries.
Device Story
Partially threaded metal fixation fastener; attaches soft tissue allografts or autografts to host bone during cruciate ligament reconstruction. Device features proximal spiral threads for bone fixation and tapered, unthreaded distal section for graft passage. Manually inserted by surgeon using cannulated 3.5mm hex driver over 0.042" guide wire. Cannulation facilitates accurate placement and reduces tunnel screw divergence. Device provides mechanical stability for graft-to-bone attachment; benefits patient by securing graft during healing process.
Clinical Evidence
Bench testing only. Mechanical 3-point bending tests performed on subject device and DePuy Cross Pin predicate to compare load-to-failure properties. Statistical analysis (Student's t-test) showed no significant difference (p > 0.05) between the two designs.
Technological Characteristics
Material: Ti-6Al-4V ELI (ASTM F-136). Surface: Anodized (AMS 2488C). Dimensions: 6.5mm diameter; 40-70mm lengths. Features: Proximal threads (ISO 5835), central cannulation for 0.042" guide wire, 3.5mm hex head (ASTM F-116-85). Energy: Manual mechanical insertion. Sterilization: Not specified.
Indications for Use
Indicated for patients undergoing cruciate ligament reconstruction surgery requiring soft tissue graft fixation.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
DePuy Cross Pin Screw System
Arthrex Cross Pin
Related Devices
K961905 — MEDICINELODGE, INC. SET SCREW · Medicinelodge, Inc. · Aug 6, 1996
{0}
SEP 17 1996
K961920
# 13. 510 (K) SUMMARY OF SAFETY AND EFFECTIVENESS INFORMATION
## 13.1 Date of Summary Preparation
May 15, 1996
## 13.2 Manufactures Contact Person
Wade Fallin
Vice President of Research and Development/General Manager
152 South 600 West
Logan, UT 84321
Telephone: 801-753-7675
Fax: 801-753-7698
## 13.3 Proprietary Device Name
MLI - Cross Pin
## 13.4 Common Name
Threaded Fixation Pin
## 13.5 Classification Name
Smooth or Threaded Metallic Bone Fixation Fastener
## 13.6 Classification Reference
21CFR 888.3040
## 13.7 Device Product Code
87 JDW
## 13.8 Regulatory Class
Class II
## 13.9 Special Controls
At this time, the Food and Drug Administration generated performance standards applicable to the MLI - Cross Pin are not in force.
## 13.10 Materials
The following material is used in the manufacturing of the MLI - Cross Pin (All Sizes). Ti-6Al-4V ELI per ASTM F-136.
## 13.11 Indications for Use
The MedicineLodge, Inc. (MLI) - Cross Pin is used to secure soft tissue graft fixations during cruciate ligament reconstruction surgeries.
{1}
35
# 13.12 Device Description
The MedicineLodge, Inc. (MLI) Cross Pin is a partially threaded metal fixation fastener used to attach soft tissue allografts or autografts to host bone.
To begin, the device is similar in design to currently marketed threaded fixation pins. The spiral threads (per ISO 5835) on the proximal one-third of the screw are designed to firmly grasp the host bone during insertion. The distal two-thirds of the MLI Cross Pin possesses two design characteristics common to cross pin devices.
a. **Unthreaded Distal Two-thirds**; The absence of threads along this portion of the cross pin allows the passage of the soft tissue graft around the screw.
b. **Tapered Distal Two-thirds**; The taper of the cross pin along the distal two-thirds allow for easy passage and positioning of this portion of the device through the tibio-femoral tunnel.
The MLI Cross Pin head conforms to ISO 5835 and the device is designed to be manually inserted using a standard cannulated 3.5mm (cannulation of 0.046") hex driver, per ASTM F-116-85.
The MLI Cross Pin will be provided with a 6.5mm outer diameter (Per ISO 5835) offered in lengths of 40mm, 50mm, 60mm and 70mm lengths. Table below lists the various MLI Cross Pin combinations along with corresponding part numbers.
All MLI Cross Pins will be manufactured from Ti-6Al-4V ELI material (ASTM F-136) and are anodized per AMS 2488C for a smoother and more uniform surface finish. The anodized layer additionally provides an inert coating which helps to prevent galling and corrosion by providing a passive surface.
As mentioned previously, the MLI Cross Pin possesses a central cannulation to allow the use of guide wires for accurate placement of the device in the host bone. In application, the use of guide wires reduces the chances of tunnel screw divergence. All MLI Cross Pins are designed for use with a 0.042 inch diameter guide wire, which is also provided by MedicineLodge, Inc. All instrumentation designed for use with the MLI Cross Pin is defined in Section 6, with engineering drawings presented in Appendix B.
MLI- Cross Pin Sizes
| PART # STERILE | PART # NON-STERILE | DIAMETER (MM) | LENGTH (MM) | DESCRIPTION |
| --- | --- | --- | --- | --- |
| 706540 | 716540 | 6.5mm | 40mm | 6.5mm x 40mm
Cross Pin |
| 706550 | 716550 | 6.5mm | 50mm | 6.5mm x 50mm
Cross Pin |
| 706560 | 716560 | 6.5mm | 60mm | 6.5mm x 60mm
Cross Pin |
| 706570 | 710570 | 6.5mm | 70mm | 6.5mm x 70mm
Cross Pin |
{2}
36
# 13.13 Substantially Equivalent Predicate Devices
| No. | MANUFACTURE'S NAME | DEVICE NAME | 510 (K) APPROVED |
| --- | --- | --- | --- |
| 1. | DePuy, Inc.
Warsaw, IN | DePuy Cross Pin Screw System | Yes |
| 2. | Arthrex, Inc.
Naples, FL | Arthrex Cross Pin | Yes |
# 13.14 Substantial Equivalence Comparison
The following table displays the similarities and differences of the new device to the legally marketed devices to which equivalency is claimed. The materials, method of manufacture, method of sterilization and device packaging are identical to the predicate devices.
| No. | FEATURE | MLI - CROSS PIN | DEPUY CROSS PIN |
| --- | --- | --- | --- |
| 1. | Intended Use | Cruciate Ligament Reconstruction Surgery | Cruciate Ligament Reconstruction Surgery |
| 2. | Diameter | 6.5mm | 4.5mm |
| 3. | Lengths | 40mm, 50mm, 60mm and 70mm | 40mm 45mm, 50mm, 55mm, 60mm, 65mm and 70mm |
| 4. | Material | Ti-6Al-4V ELI per ASTM F-136 | 316 LVM SS |
| 5. | Method of insertion | 3.5mm Hex | 4.0mm Hex |
| 6. | Method of fixation to bone | Screw Threads | Screw Threads |
| 7. | Self tapping feature | Yes | Yes |
| 8. | Method of graft attachment | Loop soft tissue around implant | Loop soft tissue around implant or drill hole through bone end of graft |
| 9. | Cannulation for guide wire | .046 cannulation for use of .042 guide wire | None |
| 10. | Surface finish | Ti anodized | High polish |
| 11. | Thread crest | Soft | Soft |
| 12. | Packaging | Sterile and Non-Sterile | Sterile and Non-Sterile |
{3}
37
# 13.15 Performance Testing (Subject and Predicate Devices)
MedicineLodge, Inc. has conducted mechanical testing on the subject and a predicate device (DePuy Cross Pin; DePuy, Inc., Warsaw, IN) to prove substantial equivalence. The following mechanical tests were performed.
a. Static bending properties of the DePuy Cross Pin in a 3-point bending situation.
b. Static bending properties of the MLI - Cross Pins in a 3-point bending situation.
MedicineLodge, Inc. conducted bending property studies on the subject and predicate device in a non-in-vitro mode to determine the effect of using a load on the cross pin to simulate the load applied on the cross pin via a graft.
Students t-test were done to determine any statistical differences or similarities between the two types of cross pin ligament fixation devices.
In comparing the load to failure between the two cross pin designs, we find no significant differences (p > 0.05).
Based on the mechanical testing conducted by MedicineLodge, Inc. on the subject and predicate device, conclusions can be drawn regarding the design difference between the subject and predicate device. The design differences seen in the subject device offer no inherent weakness and have functioned similar to the predicate device often showing potential for improved performance. MedicineLodge, Inc. would like to claim substantial equivalence to the DePuy Cross Pin which is commercially available in the U.S. market.
# 13.16 Differences from Predicate Devices
The following enumerates the differences in design between the MLI - Cross Pin in comparison with a selected predicate device.
a. Material: The MLI - Cross Pin is manufactured from Ti-6Al-4V ELI per ASTM F-136 where as the DePuy Cross Pin is manufactured from 316 LVM stainless steel per ASTM A-479/A-479M.
b. Cannulated vs. Non-cannulated: The MLI - Cross Pin design allows for the use of a .042" guide wire to help position the cross pin in the tunnel. The DePuy Cross Pin features no such cannulation.
c. 6.5mm vs 4.5mm thread form: The MLI - Cross Pin thread form is that of a 6.5 HB (ISO 5835). The DePuy Cross Pin thread form is that of 4.5 HB (ISO 5835).
d. Screw head vs no head: The MLI - Cross Pin has a screw head (ISO 5835) to accommodate a standard 3.5mm hex to accommodate a hex driver (ASTM F-136). The DePuy Cross Pin has no defined screw head but has a hex nut used in conjunction with a hex driver socket.
# 13.17 Conclusion
Based on the design concept, use of standard material, feature comparisons to selected predicate devices, the device and predicate device testing, MedicineLodge, Inc. believes that sufficient evidence exists to conclude that the MLI - Cross Pin is substantially equivalent to existing legally marketed endosteal cruciate fixation devices.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.