APEX PINS

K061493 · Stryker Howmedica Osteonics · JDW · Jul 7, 2006 · Orthopedic

Device Facts

Record IDK061493
Device NameAPEX PINS
ApplicantStryker Howmedica Osteonics
Product CodeJDW · Orthopedic
Decision DateJul 7, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3040
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

Apex® Pins are intended for use in conjunction with external fixators in children and adults. Specific indications include, but not limited to the following conditions: bone fracture fixation for open, closed and/or unstable fractures, osteotomies, arthrodeses, deformity correction including bone transport, revision procedures where other treatments or devices have been unsuccessful or bone reconstruction procedures. The pins will be available with and without hydroxylapatite (HA) coating.

Device Story

Apex® Pins are metallic external fixation pins used with external fixator systems for orthopedic procedures. The device line extension introduces hydroxylapatite (HA) coating to improve purchase in osteoporotic bone and reduce pin loosening by enhancing the pin-bone interface. Pins are used by surgeons in clinical settings to stabilize bone fractures, correct deformities, or perform reconstructions. The device functions as a mechanical fastener; it does not involve electronic processing or software. Clinical benefit includes improved long-term fixation stability in compromised bone tissue.

Clinical Evidence

Bench testing only. Mechanical strength testing was conducted to compare HA-coated pins to uncoated predicate pins to ensure no degradation in performance due to the coating process.

Technological Characteristics

Metallic external fixation pins; available with or without hydroxylapatite (HA) coating. Designed for use with external fixator systems. Mechanical fixation principle.

Indications for Use

Indicated for children and adults requiring external fixation for bone fracture fixation (open, closed, unstable), osteotomies, arthrodeses, deformity correction (including bone transport), revision procedures, or bone reconstruction.

Regulatory Classification

Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K061493 JUL - 7 2006 ## 510(k) Summary of Safety and Effectiveness Apex® Pins with HA Coating | Proprietary Name: | Apex® Pins | |--------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Common Name: | External Fixation Pins | | Classification Name/Reference: | Smooth or threaded metallic bone fixation<br>fastener - 21 CFR §888.3040 | | Device Product Code: | 87 JDW | | Proposed Regulatory Class: | Class II | | For Information contact: | Vivian Kelly, Regulatory Affairs Specialist<br>Howmedica Osteonics Corp.<br>325 Corporate Drive<br>Mahwah, NJ 07430<br>Phone: (201) 831-5581 Fax: (201) 831-6038 | | Date Summary Prepared: | May 30, 2006 | ## Description This submission is a line extension to the Apex® Pin product line to include fixation pins coated with hydroxylapatite (HA) to improve purchase in osteoporotic bone and for use in long-term fixation to reduce the incidence of pin loosening by enhancing the fixation at the pin-bone interface. The Apex® Pins are external fixation pins used with external fixators for fracture fixation. They are intended to be used with the other components in Howmedica Osteonics' external fixation systems. ## Indications: Apex® Pins are intended for use in conjunction with external fixators in children and adults. Specific indications include, but not limited to the following conditions: bone fracture fixation for open, closed and/or unstable fractures, osteotomies, arthrodeses, deformity correction including bone transport, revision procedures where other treatments or devices have been unsuccessful or bone reconstruction procedures. The pins will be available with and without hydroxylapatite (HA) coating. ## Substantial Equivalence: Apex® Pins with hydroxylapatite (HA) coating are substantially equivalent to other commercially available uncoated and HA coated pins in regards to intended use, design, materials, and operational principles as external fixation pins. Testing was conducted to compare the subject HA coated pins to its uncoated predicate for mechanical strength and to characterize the HA coating. The results demonstrate that the subject pins are substantially equivalent in strength to the predicate pins with and without HA coating. {1}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" around the perimeter. Inside the circle is a stylized symbol that resembles three overlapping human profiles or faces. Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JUL - 7 2006 Stryker Howmedica Osteonics Corp. % Ms. Vivian Kelly Regulatory Affairs Specialist 325 Corporate Drive Mahwah, New Jersey 07430 Re: K061493 Trade/Device Name: Apex® Pins Regulation Number: 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: II Product Code: JDW Dated: May 30, 2006 Received: May 31, 2006 Dear Ms. Vivian Kelly: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality {2}------------------------------------------------ Page 2 - Ms. Vivian Kelly systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120 . Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours. Barbara Prieur Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ Indications for Use KO61493 510(k) Number (if known): Device Name: Apex® Pins Indications for Use: Apex® Pins are intended for use in conjunction with external fixators in children and adults. Specific indications include, but not limited to the following conditions: bone fracture fixation for open, closed and/or unstable fractures, osteotomies, arthrodeses, deformity correction including bone transport, revision procedures where other treatments or devices have been unsuccessful or bone reconstruction procedures. The pins will be available with and without hydroxylapatite (HA) coating. Prescription Use Over-The-Counter Use X AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Page 1 of 1 Barbara Buckley Division of General, Restorative, and Neurological Devices 510(k) Number K061493
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...