Treace Medical Concepts (TMC) Implant Fixation System
K222645 · Treace Medical Concepts · JDR · Nov 21, 2022 · Orthopedic
Device Facts
Record ID
K222645
Device Name
Treace Medical Concepts (TMC) Implant Fixation System
Applicant
Treace Medical Concepts
Product Code
JDR · Orthopedic
Decision Date
Nov 21, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The system is intended to be used for fracture fixation, osteotomy fixation, and joint arthrodesis of the foot and ankle.
Device Story
TMC Compression Implant System consists of metallic bone staples and associated instrumentation; used for fracture fixation, osteotomy fixation, and joint arthrodesis of foot and ankle. Implants provided in various bridge lengths, leg lengths, and cross-sections to accommodate patient anatomy. System utilized by surgeons in clinical/OR settings to provide mechanical stabilization of bone segments. Implants facilitate bone healing by providing compression across fracture or arthrodesis sites. Device is implanted via surgical procedure; instrumentation assists in proper placement. Benefits include stabilization of bone segments to promote fusion or healing.
Clinical Evidence
Bench testing only. Performance testing conducted per ASTM F564 for metallic bone staples comparing subject device to primary predicate (Z-Staple). Testing demonstrated subject device met all acceptance criteria.
Technological Characteristics
Implant grade titanium alloy (Ti-6Al-4V-ELI) per ASTM F136. Bone fixation staples with varying bridge/leg lengths and cross-sections. Sterile via gamma irradiation. Mechanical fixation principle.
Indications for Use
Indicated for fracture fixation, osteotomy fixation, and joint arthrodesis of the foot and ankle in patients requiring bone fixation.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Wright Medical FuseForce™ Flex Dynamic Compression System (K203832)
Reference Devices
Treace Medical Concepts (TMC) Plating System (K220136)
Related Devices
K232387 — Treace Medical Concepts (TMC) Compression Implant System · Treace Medical Concepts · Aug 28, 2023
K242415 — TMC Compression Implant System · Treace Medical Concepts · Sep 6, 2024
K243658 — TMC Compression Implant System · Treace Medical Concepts · Dec 26, 2024
K172973 — Life Spine Foot and Ankle Plating System · Life Spine, Inc. · Dec 21, 2017
K161037 — Tarsa-Link Stand-Alone Wedge Fixation System · Life Spine, Inc. · Aug 15, 2016
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration".
February 8, 2023
Treace Medical Concepts Brittany Grochala Sr. Regulatory Affairs Specialist 203 Fort Wade Road, Suite 150 Ponte Vedra, Florida 32081
Re: K222645
Trade/Device Name: Treace Medical Concepts (TMC) Compression Implant System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: JDR
Dear Brittany Grochala:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated November 21, 2022. Specifically, FDA is updating this SE Letter as an administrative correction to the Trade/Device Name from "Treace Medical Concepts (TMC) Implant Fixation System" to "Treace Medical Concepts (TMC) Compression Implant System".
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Limin Sun, Ph.D., OHT6: Office of Orthopedic Devices, 301-796-7056, Limin.Sun(@fda.hhs.gov.
Sincerely,
Image /page/0/Picture/9 description: The image shows the text "Limin Sun -S" in a large, sans-serif font. The text is black and the background is white. The words are arranged horizontally, with "Limin" and "Sun" being the most prominent words. The "-S" is smaller and slightly offset to the right.
Limin Sun, Ph.D. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 21, 2022
Treace Medical Concepts Brittany Grochala Sr. Regulatory Affairs Specialist 203 Fort Wade Road, Suite 150 Ponte Vedra, Florida 32081
Re: K222645
Trade/Device Name: Treace Medical Concepts (TMC) Implant Fixation System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: JDR Dated: August 31, 2022 Received: September 1, 2022
Dear Brittany Grochala:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Limin Sun -S
Limin Sun, Ph.D. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K222645
Device Name
Treace Medical Concepts (TMC) Compression Implant System
Indications for Use (Describe)
The system is intended to be used for fracture fixation, and joint arthrodesis of the foot and ankle.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image is a logo for Treace Medical Concepts, Inc. The logo features the word "TREACE" in gray, with a red swoosh above the "A". Below the word "TREACE" is the text "Medical Concepts, Inc." in gray. Below that is the text "The Leader in Hallux Valgus Surgery" in red.
# 510(k) Summary Treace Medical Concepts (TMC) Compression Implant System August 31st, 2022
| Company: | Treace Medical Concepts, Inc.<br>203 Fort Wade Road, Suite 150<br>Ponte Vedra, FL 32081 |
|-----------------------------|-----------------------------------------------------------------------------------------|
| Establishment Registration: | 3011623994 |
| Primary Contact: | Brittany Grochala, Sr. Regulatory Affairs Specialist |
| | Phone: 515-865-0494 |
| | Fax: 904-834-7169 |
| | Email: bgrochala@treace.net |
| Secondary Contact: | Kristina Hall, Director, Regulatory Affairs |
| | Phone: 904-373-5940 ext. 1321 |
| | Fax: 904-834-7169 |
| | Email: khall@treace.net |
| Trade Name: | Treace Medical Concepts (TMC) Compression Implant |
| | System |
| Common Name: | Staple, Fixation, Bone |
| Classification: | Class II |
| Regulation Number: | 21 CFR 888.3030 Single/multiple component metallic bone |
| | fixation appliances and accessories. |
| Panel: | 87- Orthopedic |
| Product Code(s): | JDR Staple, Fixation, Bone |
# Predicate Device(s):
Primary Predicate:
- Z-Medical Z-Staple (K121277, S.E. 11/07/2012) .
Additional Predicate:
- Wright Medical FuseForce™ Flex Dynamic Compression System (K203832, S.E. . 05/07/2021)
Reference Device:
- Treace Medical Concepts (TMC) Plating System (K220136, S.E. 02/16/2022) .
# Device Description:
The Treace Medical Concepts (TMC) Compression Implant System consists of implants and related instrumentation for implantation. The implant is offered in multiple combinations of bridge lengths, leg lengths, cross sections and cannulated versions to accommodate various anatomies. The system is intended to be used for fracture fixation, osteotomy fixation, and joint arthrodesis of the foot and ankle.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for TREACE Medical Concepts, Inc. The logo is in gray, except for a red swoosh above the letter A. Below the company name is the tagline "The Leader in Hallux Valgus Surgery" in red.
The purpose of this traditional 510(k) submission is for the introduction of the Treace Medical Concepts (TMC) Compression Implant System.
All implantable components are manufactured from implant grade titanium alloy (Ti-6AI-4V-ELI) per ASTM F136 and are provided sterile by gamma irradiation.
# Indications for Use:
The system is intended to be used for fracture fixation, osteotomy fixation, and joint arthrodesis of the foot and ankle.
## Substantial Equivalence:
The subject TMC Compression Implant System is substantially equivalent to the following predicate and reference devices:
Primary Predicate:
- Z-Medical Z-Staple (K121277, S.E. 11/07/2012) .
Additional Predicate:
- Wright Medical FuseForce™ Flex Dynamic Compression System (K203832, S.E. . 05/07/2021)
Reference Device:
- Treace Medical Concepts (TMC) Plating System (K220136, S.E. 02/16/2022) .
The subject compression implants are intended to be used for fracture fixation, osteotomy fixation, and joint arthrodesis of the foot and ankle, equivalent to the predicate devices. The subject compression implants are manufactured from implant grade titanium alloy (Ti-6AI-4V-ELI) per ASTM F136 and also share similar materials, geometry, construction, packaging, and overall design with the predicate and reference devices (Z-Staple and TMC Plating System). Thus, it can be concluded that the subject devices do not raise new questions about safety and effectiveness and are substantially equivalent to the predicate and reference devices.
# Performance Testing:
The subject device was compared to the primary predicate device. Z-Staples, per ASTM F564 for Bone Metallic Staples. The testing demonstrated that the subject device met all acceptance criteria. Therefore, the subject device is substantially equivalent to the predicate device.
### Conclusion:
The TMC Compression Implant System has similar intended use, overall design, materials, and mechanical properties to that of the predicate and reference devices. Therefore, it can be concluded that the subject device is at least as safe and substantially equivalent to the predicate and reference devices.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.