← Product Code [JDR](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JDR) · K190166

# TriMed Nitinol Staple System (K190166)

_TriMed, Inc. · JDR · Jul 1, 2019 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JDR/K190166

## Device Facts

- **Applicant:** TriMed, Inc.
- **Product Code:** [JDR](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JDR.md)
- **Decision Date:** Jul 1, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The TriMed Nitinol Staple System is intended for use in fractures, osteotomy, and arthrodesis of the following areas of the body: small bones in the hand and foot.

## Device Story

TriMed Nitinol Staple System is a sterile, non-bioabsorbable, implantable bone fixation device. Manufactured from medical-grade Nitinol, staples are available in symmetrical and asymmetrical leg configurations. Used in orthopedic procedures for fractures, osteotomies, and arthrodesis of small bones in the hand and foot. Implanted using reusable instruments for site preparation and insertion. Provides mechanical fixation to stabilize bone fragments during healing. Intended for prescription use by surgeons in clinical settings.

## Clinical Evidence

Bench testing only. Performance data includes pull-out fixation strength, elastic static bending, constant amplitude bending fatigue, corrosion susceptibility testing, quantification of titanium oxide layer thickness, and transformation temperature (At) analysis.

## Technological Characteristics

Medical-grade Nitinol construction. Symmetrical and asymmetrical leg configurations. Implantable bone fixation appliance. Sterile, non-bioabsorbable. Utilizes reusable instrumentation for site preparation and insertion.

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Eleos Staple ([K112794](/device/K112794.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized depiction of a human figure. To the right is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

July 1, 2019

TriMed, Inc. % Jean Asquith Regulatory Affairs Consultant Jean Asquith 4221 Lost Oasis Hollow Austin, Texas 78739

Re: K190166

Trade/Device Name: TriMed Nitinol Staple System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances and Accessories Regulatory Class: Class II Product Code: JDR Dated: May 24, 2019 Received: May 30, 2019

Dear Jean Asquith:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For Vesa Vuniqi Acting Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K190166

Device Name TriMed Nitinol Staple System

#### Indications for Use (Describe)

The TriMed Nitinol Staple System is intended for use in fractures, osteotomy, and arthrodesis of the following areas of the body: small bones in the hand and foot.

| Type of Use (Select one or both, as applicable)                                                                     |  |
|---------------------------------------------------------------------------------------------------------------------|--|
| <div> <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)    </span> </div> |  |
| <div> <span> <span style="font-size: 16px;">□</span> Over-The-Counter Use (21 CFR 801 Subpart C)    </span> </div>  |  |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

#### 510(k) Summary

#### 1. SUBMITTER INFORMATION

TriMed, Inc. 27533 Avenue Hopkins Santa Clarita, CA 91355

| Company Contact:           | David Medoff<br>Tel: (661) 255-7406<br>Fax: (661) 254-8485 |
|----------------------------|------------------------------------------------------------|
| Email:                     | davidmedoff@trimedortho.com                                |
| 510(k) Consultant Contact: | Jean Asquith<br>Tel: (512) 680-5802                        |
| Email:                     | jtasquith.ra@gmail.com                                     |
| Date Prepared:             | June 21, 2019                                              |

### 2. DEVICE IDENTIFICATION

| Device Trade Name:      | TriMed Nitinol Staple                                                         |
|-------------------------|-------------------------------------------------------------------------------|
| Device Common Name:     | Staple, Fixation, Bone                                                        |
| Regulation Number:      | 21 CFR 888.3030                                                               |
| Regulation Description: | Single/multiple component metallic bone fixation appliance<br>and accessories |
| Product Code:           | JDR                                                                           |
| Device Class:           | Class II                                                                      |
| Classification Panel:   | Orthopedic                                                                    |

#### 3. DEVICE DESCRIPTION

The TriMed Staple is a sterile, non-bioabsorbable, implantable device designed to address common types of deformity corrections, fractures and osteotomies in extremities amenable to staple fixation, all depending on the size, shape, and location of the fractured bone and bone fracture fragments. TriMed Staples are manufactured from medical grade Nitinol. Two types of staple leg configuration are available; symmetrical and asymmetrical. The staple is implanted with a set of re-usable instruments designed for: preparation of the implant site and insertion of the implant into bone.

### 4. INDICATIONS FOR USE

The TriMed Nitinol Staple System is intended for use in fractures, osteotomy, and arthrodesis of the following areas of the body: small bones in the hand and foot.

{4}------------------------------------------------

# 5. COMPARISON TO PREDICATE DEVICES

The TriMed Nitinol Staple System is substantially equivalent to the following device:

| 510(k) Number | Device                                                                             | Applicant |
|---------------|------------------------------------------------------------------------------------|-----------|
| K112794       | Eleos Staple (included as part of<br>the TriMed EOS Small Bone<br>Fixation System) | TriMed    |

The TriMed Nitinol Staple is substantially equivalent to the Eleos Staple when comparing the following characteristics:

- Intended Use ●
- Principles of Use
- Fundamental Technology
- Basic Design ●
- Materials Used ●
- Where used ●
- Sterilization Method o

# 6. TESTING AND PERFORMANCE DATA

To demonstrate product performance, TriMed conducted the following tests and compared the results to the predicate device.

- Pull-Out Fixation Strength; ●
- Elastic Static Bending;
- Constant Amplitude Bending Fatigue; ●
- Corrosion Susceptibility Testing; and ●
- Quantification of the titanium oxide layer thickness ●

In addition, the transformation temperature (At) was quantified.

# 7. CONCLUSIONS

The TriMed Nitinol Staple is substantially equivalent to the predicate device as the basic design features and intended uses are the same. Any differences between the subject and predicate devices are considered minor and do not raise questions concerning safety or effectiveness. Based on the indications for use, technological characteristics, and the summary of data submitted, TriMed, Inc. has determined that the proposed device is substantially equivalent to the currently marketed predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JDR/K190166](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JDR/K190166)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
