K231333 · Globus Medical, Inc. · JDQ · Aug 4, 2023 · Orthopedic
Device Facts
Record ID
K231333
Device Name
STRETTO Cable System
Applicant
Globus Medical, Inc.
Product Code
JDQ · Orthopedic
Decision Date
Aug 4, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3010
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The STRETTO™ Cable System is indicated for use in: general orthopedic trauma surgery (e.g., fractures of the olecranon, patella, femur - including periprosthetic, pelvis, acetabulum, humerus and ankle, and acromioclavicular dislocations); prophylactic banding during total joint procedures; and, temporary reduction techniques for ORIF (Open Reduction Internal Fixation) procedures. STRETTO™ Screw Anchors are indicated for fractures that may not be securely held by either a screw or a cerclage device alone, and where cerclage is used in combination with bone screws and/or plates to provide internal fixation of fractured bone. STRETTO™ Press-In Anchors are indicated for use with a cerclage cable and plate to augment long bone fracture fixation, particularly when the use of screws would be inhibited, as in the presence of intramedullary implants.
Device Story
STRETTO™ Cable System consists of cables, crimps, and anchors used for bone fixation. Anchors include press-in styles for ANTHEM fracture plates and screw-in styles for ANTHEM screw hex recesses. System provides internal fixation for fractures or prophylactic banding. Used in orthopedic trauma surgery by surgeons. Device functions by securing bone fragments via cable tensioning and crimping, often in conjunction with plates or screws. Benefits include stabilization of fractures where traditional screw fixation is insufficient or inhibited by existing implants.
Clinical Evidence
Bench testing only. Mechanical testing (static and dynamic tension) performed using ASTM E8 as a guide. Bacterial endotoxin testing (BET) conducted per ANSI/AAMI ST72:2011. No clinical data provided.
Technological Characteristics
Materials: titanium alloy, commercially pure titanium, cobalt chromium alloy, cobalt chromium molybdenum alloy, or stainless steel. Components: cables, crimps, and anchors. Anchors include press-in and screw-in variants. Mechanical testing guided by ASTM E8. Sterilization/biocompatibility testing per ANSI/AAMI ST72:2011.
Indications for Use
Indicated for patients requiring orthopedic trauma surgery for fractures (olecranon, patella, femur, pelvis, acetabulum, humerus, ankle), acromioclavicular dislocations, prophylactic banding during total joint procedures, or temporary reduction for ORIF. Screw anchors indicated for fractures requiring combined cerclage and screw/plate fixation. Press-in anchors indicated for long bone fracture augmentation when screw use is inhibited by intramedullary implants.
Regulatory Classification
Identification
A bone fixation cerclage is a device intended to be implanted that is made of alloys, such as cobalt-chromium-molybdenum, and that consists of a metallic ribbon or flat sheet or a wire. The device is wrapped around the shaft of a long bone, anchored to the bone with wire or screws, and used in the fixation of fractures.
Predicate Devices
Pioneer Laboratories SDB Cerclage System (K992616)
K031162 — SMITH & NEPHEW ORTHOPAEDIC CABLING SYSTEM · Smith & Nephew, Inc. · May 1, 2003
K980690 — CABLE SCREW WITH INTEGRAL CRIMP · Pioneer Surgical Technology · May 22, 1998
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration".
August 4, 2023
Globus Medical Inc. Jennifer Antonacci Group Manager, Regulatory Affairs 2560 General Armistead Ave. Audubon, Pennsylvania 19403
Re: K231333
Trade/Device Name: STRETTO™ Cable System Regulation Number: 21 CFR 888.3010 Regulation Name: Bone Fixation Cerclage Regulatory Class: Class II Product Code: JDQ, HWC, HRS Dated: May 8, 2023 Received: May 8, 2023
Dear Jennifer Antonacci:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/1/Picture/5 description: The image shows the name "Shumaya Ali -S" in a large, clear font. The text is horizontally oriented and appears to be the primary focus of the image. The background is plain, with a faint design that does not distract from the text.
Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K231333
Device Name STRETTO™ Cable System
#### Indications for Use (Describe)
The STRETTO™ Cable System is indicated for use in: general orthopedic trauma surgery (e.g., fractures of the olecranon, patella, femur - including periprosthetic, pelvis, acetabulum, humerus and ankle, and acromioclavicular dislocations); prophylactic banding during total joint procedures; and, temporary reduction techniques for ORIF (Open Reduction Internal Fixation) procedures.
STRETTO™ Screw Anchors are indicated for fractures that may not be securely held by either a screw or a cerclage device alone, and where cerclage is used in combination with bone screws and/or plates to provide internal fixation of fractured bone.
STRETTO™ Press-In Anchors are indicated for use with a cerclage cable and plate to augment long bone fracture fixation, particularly when the use of screws would be inhibited, as in the presence of intramedullary implants.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary: STRETTO™ Cable System
Company: Globus Medical Inc. 2560 General Armistead Ave. Audubon, PA 19403 610-930-1800
Contact: Jennifer Antonacci, Ph.D. Group Manager, Requlatory Affairs
- August 3, 2023 Date Prepared:
Device Name: STRETTO™ Cable System
Common Name: Bone fixation cerclage cable
Classification: Per 21 CFR as follows: &88.3010 Bone Fixation Cerclage (primary) §888.3030 Single/multiple component metallic bone fixation appliances and accessories §888.3040 Smooth or threaded metallic bone fixation fastener
> Product Code: JDQ (primary), HWC, HRS Regulatory Class: II
Primary Predicate: Pioneer Laboratories SDB Cerclage System (K992616)
# Additional
| Predicates: |
|-------------------------------------------------------------------------------------------------------|
| Pioneer Laboratories Hex Button Device (K992617)<br>Synthes Variable Angle Positioning Pins (K162124) |
# Purpose:
The purpose of this submission is to request clearance for the STRETTO™ Cable System.
### Device Description:
The STRETTO™ Cable System implants are comprised of cables, crimps and anchors. Anchors include a thru hole for the cables to pass through, and are available in multiple styles. Press-in anchors mate with ANTHEM fracture plates and screw anchors mate with the hex recess of ANTHEM screws. The cable is secured with the crimp and may be used with the anchors. STRETTO™ Implants are manufactured from titanium allov. commercially pure titanium. cobalt chromium alloy, cobalt chromium molybdenum alloy, or stainless steel.
{4}------------------------------------------------
# Indications for Use:
The STRETTO™ Cable System is indicated for use in general orthopedic trauma surgery, including olecranon, patella, femur (including periprosthetic fractures), pelvic, acetabular, humeral and ankle fractures, acromioclavicular dislocations, prophylactic banding during total joint procedures, and temporary reduction techniques for ORIF (Open Reduction Internal Fixation) procedures.
STRETTO™ Screw Anchors are indicated for fractures that may not be securely held by either a screw or a cerclage device alone, and where cerclage is used in combination with bone screws and/or plates to provide internal fixation of fractured bone.
STRETTO™ Press-In Anchors are indicated for use with a cerclage cable to augment fracture stabilization with plates used in long bone fixation, and when the use of screws would be inhibited, as in the presence of intramedullary implants.
# Performance Data:
Mechanical testing (static and dynamic tension) was conducted using ASTM E8 as a guide. Performance data demonstrate substantial equivalence to the predicate devices. An engineering analysis was conducted for the anchors to demonstrate substantial equivalence to the predicate devices. Bacterial endotoxin testing (BET) was conducted in accordance with ANSI/AAMI ST72:2011.
### Technological Characteristics:
The STRETTO™ Cable System has similar technological characteristics as the predicate devices including design, intended use, material composition, function, and range of sizes.
### Basis of Substantial Equivalence:
The STRETTO™ Cable System has been found to be substantially equivalent to the predicate devices with respect to technical characteristics, performance, and intended use. The information provided supports substantial equivalence of the subject devices to the predicate devices.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.