← Product Code [JDO](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JDO) · K152686

# CONQUEST FN (K152686)

_Smith & Nephew · JDO · Mar 17, 2016 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JDO/K152686

## Device Facts

- **Applicant:** Smith & Nephew
- **Product Code:** [JDO](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JDO.md)
- **Decision Date:** Mar 17, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

CONQUEST FN is intended for use in internal fixation of femoral neck fractures

## Device Story

CONQUEST FN is an internal fixation plating system for femoral neck fractures; consists of contoured locking bone plates and compatible locking/non-locking bone screws. Device provides mechanical stabilization of fractures; used by orthopedic surgeons in clinical/surgical settings. System utilizes POGO screw technology to facilitate fixation. Implants are single-use; provided sterile. Healthcare providers use the system to achieve bone fixation, promoting healing and stability in patients with intracapsular femoral neck fractures.

## Clinical Evidence

Bench testing only. Testing included axial pullout strength, insertion torque performance, removal torque performance, torsional strength, and construct fatigue testing. Results demonstrated that the CONQUEST FN system met all acceptance criteria and exhibited similar or superior performance compared to the predicate devices in simulated femoral neck fracture models.

## Technological Characteristics

Internal fixation plating system; materials: 316L implant-grade stainless steel; components: contoured locking bone plates, locking and non-locking bone screws; sterilization: Gamma irradiation; single-use; mechanical fixation principle.

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Targon FN System ([K102057](/device/K102057.md))
- POGO Screw ([K080649](/device/K080649.md))
- Smith & Nephew Cannulated Screws and Washers ([K111994](/device/K111994.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 17, 2016

Smith & Nephew, Incorporated Bradley Heil Regulatory Affairs Manager 1450 East Brooks Road Memphis, Tennessee 38116

Re: K152686 Trade/Device Name: CONQUEST FN Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: JDO, KTT Dated: February 16, 2016 Received: February 18, 2016

Dear Mr. Heil:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing

{1}------------------------------------------------

(21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

# Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Form Approved: OMB No. 0910-0120

Expiration Date: January 31, 2017

See PRA Statement below.

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

510(k) Number (if known)

#### K152686

Device Name CONQUEST FN

Indications for Use (Describe)

The Smith & Nephew CONQUEST FN is indicated for displaced intracapsular femoral neck fractures.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the Smith & Nephew logo. The logo consists of an orange flower-like symbol on the left, followed by the text "We are smith&nephew". The words "We are" are in gray, while "smith&nephew" is in orange.

| Submitted by:                                | Smith & Nephew, Inc.<br>Advance Surgical Devices Division<br>1450 East Brooks Road<br>Memphis, Tennessee 38116 |
|----------------------------------------------|----------------------------------------------------------------------------------------------------------------|
| Date of Summary:                             | March 15, 2016                                                                                                 |
| Contact Person and Address:                  | Bradley Heil<br>Regulatory Affairs Manager<br>T (901) 399-6339<br>F (901) 566-7831                             |
| Name of Device:                              | CONQUEST FN                                                                                                    |
| Common Name:                                 | Internal Fracture Fixation Device                                                                              |
| Device Classification Name and<br>Reference: | 21 CFR 888.3030 Single/multiple component metallic bone<br>fixation appliances and accessories<br>Class II     |
| Panel Code:                                  | Orthopedics/87                                                                                                 |
| Product Code:                                | JDO, KTT                                                                                                       |

#### Device Description

The subject premarket notification describes the Smith & Nephew CONQUEST FN. The subject internal fixation plating system is comprised of contoured locking bone plates and compatible locking and non-locking bone screws. All described implant devices are manufactured from implant-grade stainless steel (316L) and designed for single-use. They will be provided in a sterile packaged option and will be sterilized via Gamma irradiation.

#### Intended Use

CONQUEST FN is intended for use in internal fixation of femoral neck fractures

#### Indications for Use

The Smith & Nephew CONQUEST FN is indicated for displaced and undisplaced intracapsular femoral neck fractures.

#### Technological Characteristics

Device comparisons described in this premarket notification demonstrated that the proposed CONQUEST FN implants are substantially equivalent to the legally marketed predicate devices listed below with regard to intended use, indications for use, and performance characteristics. The subject CONQUEST FN features characteristics very similar to the Targon FN cleared via K102057, with the primary differences being material changes and the utilization of POGO Screw technology (K080649). To accommodate for differences between the subject device and the predicate cleared via K102057, mechanical testing was conducted to compare the two devices.

#### Substantial Equivalence Information

{4}------------------------------------------------

When compared to the predicate devices listed below, substantial equivalence is based upon similar function, intended use, indications for use and overall design to the devices listed in the table below.

| Manufacturer                     | Description                                     | Submission<br>Number | Clearance<br>Date |
|----------------------------------|-------------------------------------------------|----------------------|-------------------|
| Aesculap Implant<br>Systems, LLC | Targon FN System (Primary Predicate)            | K102057              | 12/21/10          |
| FxDevices                        | POGO Screw                                      | K080649              | 11/25/08          |
| Smith & Nephew                   | Smith & Nephew Cannulated Screws and<br>Washers | K111994              | 10/11/2011        |

Table 1: Substantially Equivalent Predicates to CONQUEST FN

### Summary of Pre-Clinical Testing

- Axial Pullout Strength Evaluation was conducted on the CONQUEST FN Locking screws. Results of the test concluded that the CONQUEST FN Locking Screws met the acceptance criteria and exhibited similar or superior axial pullout strength compared to the predicate.
- . Insertion Torque Performance was conducted on the proposed locking screws. Results of the testing concluded that the CONQUEST FN screws met the acceptance criteria and exhibited similar or superior insertion torque performance compared to the predicate.
- . Removal Torque Performance was conducted on the CONQUEST FN Locking screws. Results of the test concluded that the subject device met the acceptance criterion and is expected to have acceptable torsional performance with regard to removal torque.
- Torsional Strength Evaluation was conducted on the CONQUEST FN Locking Screws. ● Results of the test concluded that the CONQUEST FN Locking Screws met the acceptance criteria and exhibited similar or superior torsional strength compared to the predicate.
- . Construct Fatigue Testing was conducted on the CONQUEST FN System. Results of the test concluded that the CONQUEST FN system met the acceptance criteria and is expected to have similar construct stability during fatigue as compared to the predicate.
- Construct Fatigue Evaluation of the CONQUEST FN System as compared to Three Cannulated Screws in a Simulated Femoral Neck Fracture Model. Results of the test concluded that the CONQUEST FN System met the acceptance criteria and exhibited similar or superior construct fatigue as compared to the predicate.
- Construct Fatigue Evaluation was conducted on the one hole CONQUEST FN locking plate as compared to Three Cannulated Screw. Results of the test concluded that the CONQUEST FN one hole locking plate met the acceptance criteria by having at least similar or superior construct fatigue performance as compared to the predicate.

{5}------------------------------------------------

## Conclusion

This 510(k) premarket notification is being submitted to request clearance for CONQUEST FN. Based on the similarities to the predicate components and a review of the mechanical testing performed, the devices are substantially equivalent to the predicate systems listed in Table 1.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JDO/K152686](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JDO/K152686)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
