VERSYS EPOCH FULLCOAT HIP PROSTHESIS, MODEL 4088 SERIES
K073499 · Zimmer, Inc. · JDL · Aug 7, 2008 · Orthopedic
Device Facts
Record ID
K073499
Device Name
VERSYS EPOCH FULLCOAT HIP PROSTHESIS, MODEL 4088 SERIES
Applicant
Zimmer, Inc.
Product Code
JDL · Orthopedic
Decision Date
Aug 7, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3320
Device Class
Class 3
Attributes
Therapeutic
Indications for Use
The VerSys Epoch FullCoat Hip Prosthesis is indicated for: - Patients suffering from severe hip pain and disability due to rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis of the femoral head, and nonunion of previous fractures of the femur. - Patients with congenital hip dysplasia, protrusio acetabuli, or slipped capital femoral epiphysis. - Patients suffering from disability due to previous fusion. - Patients with acute femoral neck fractures. This device is intended for cementless use only.
Device Story
Proximal femoral hip implant for total hip arthroplasty; replaces proximal femur. Device features Zimaloy (CoCrMo alloy) core, injection-molded PEEK layer, and titanium fiber metal (FM) porous coating. Includes 12/14 Morse-type taper for modular femoral head attachment. Proximal geometry is trapezoidal; available in Reduced Neck Length (RNL) with 125°/135° angles and Low Head Center (LHC) with 125° angle. Implanted by orthopedic surgeons in clinical/OR settings. Provides structural support for hip joint; enables joint articulation via modular head. Benefits patients by restoring hip function and reducing pain associated with degenerative or traumatic conditions.
Clinical Evidence
No clinical data provided; device equivalence established via mechanical bench testing.
Indicated for patients with severe hip pain/disability due to rheumatoid, osteoarthritis, traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis, femoral nonunion, congenital hip dysplasia, protrusio acetabuli, slipped capital femoral epiphysis, previous fusion, or acute femoral neck fractures. For cementless use only.
Regulatory Classification
Identification
A hip joint metal/metal semi-constrained, with a cemented acetabular component, prosthesis is a two-part device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that consist of a femoral and an acetabular component, both made of alloys, such as cobalt-chromium-molybdenum. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
K030079 — VERSYS BEADED FULLCOAT BOWED REVISION HIP PROSTHESIS · Zimmer, Inc. · Feb 5, 2003
K974294 — ENCORE LINEAR POROUS COATED HIP · Encore Orthopedics, Inc. · Jan 12, 1998
K192660 — Zimmer M/L Taper Hip Prosthesis · Zimmer, Inc. · Feb 28, 2020
K041109 — ZIMMER ANATOMIC II HIP PROSTHESIS · Zimmer, Inc. · Jul 20, 2004
K984154 — APF POROUS COATED LINE EXTENSION · Biomet, Inc. · Feb 11, 1999
Submission Summary (Full Text)
{0}------------------------------------------------
K073499 (pg. 1 of 2)
# Summary of Safety and Effectiveness
| Submitter : | Zimmer, Inc.<br>P.O. Box 708<br>Warsaw, IN 46581-0708 |
|---------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Eric S. Pittman<br>Associate, Corporate Regulatory Affairs<br>Telephone: (574) 371-8369<br>Fax: (574) 372-4605 |
| Date: | March 15, 2008 |
| Trade Name: | VerSys® Epoch® FullCoat Hip System |
| Common Name: | Proximal Femoral Prosthesis |
| Classification Name<br>and Reference: | Hip joint metal/metal semi-constrained, with a<br>cemented acetabular component, prosthesis (21<br>CFR § 888.3320-JDL) |
| Predicate Device: | VerSys® Epoch® FullCoat Hip System,<br>manufactured by Zimmer, Inc., K052321, cleared<br>February 15, 2006. |
| Device Description: | The VerSys Epoch FullCoat Hip stem is an addition<br>to the currently marketed VerSys Epoch FullCoat<br>proximal femoral hip implant family. It is a set of<br>fully porous coated implants that are comprised of<br>Zimaloy™, a cobalt chromium molybdenum<br>(CoCrMo) alloy core surrounded by a layer of<br>injection molded polyetheretherketone (PEEK)<br>encased in a layer of commercially pure titanium<br>fiber metal (FM). The stem is designed to replace<br>the proximal femur in a total hip arthroplasty. It<br>features a 12/14 Morse-type taper to accommodate<br>the attachment of modular femoral heads. The<br>proximal geometry of the device is trapezoidal and<br>comes in two options Reduced Neck Length (RNL) |
{1}------------------------------------------------
K073499 2 of 2 ్రామ్మ
with 125° and 135° angles and Low Head Center
(LHC) with a 125° angle.
:
| Intended Use: | The VerSys Epoch FullCoat Hip Prosthesis is<br>indicated for:<br>- Patients suffering from severe hip pain and<br>disability due to rheumatoid arthritis, osteoarthritis,<br>traumatic arthritis, polyarthritis, collagen disorders,<br>avascular necrosis of the femoral head, and<br>nonunion of previous fractures of the femur.<br>- Patients with congenital hip dysplasia, protrusio<br>acetabuli, or slipped capital femoral epiphysis.<br>- Patients suffering from disability due to previous<br>fusion.<br>- Patients with acute femoral neck fractures.<br>This device is intended for cementless use only. |
|----------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Comparison to Predicate Device: | Compared to the predicate, the subject device has<br>the same intended use, similar physical and<br>performance characteristics and is manufactured<br>using similar processes. |
| Performance Data (Nonclinical<br>and/or Clinical): | Non-Clinical Performance and Conclusions: |
| | Mechanical testing of the device indicated that it is<br>substantially equivalent to the legally market<br>predicate. |
| | Clinical Performance and Conclusions: |
| | Clinical data and conclusions were not needed for<br>this device. |
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image contains the text "Public Health Service". The text is written in a simple, sans-serif font. The words are arranged horizontally, with "Public" and "Health" on the first line and "Service" on the second line. The text is black against a white background.
Image /page/2/Picture/2 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features a stylized eagle with its head turned to the left. The eagle's body is formed by three curved lines. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" is arranged in a circular pattern around the eagle.
NOV 182008
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Zimmer, Inc. % Ms. Natalie S. Heck Sr. Manager, Regulatory Affairs P.O. Box 708 Warsaw, Indiana 46581-0708
Re: K073499 Trade/Device Name: VerSys® Epoch® FullCoat Hip System Regulation Number: 21 CFR 888.3358 Regulation Name: Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis Regulatory Class: II Product Code: LPH, JDL Dated: July 28, 2008 Received: July 30, 2008
Dear Ms. Heck:
This letter corrects our substantially equivalent letter of August 7, 2008.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{3}------------------------------------------------
Page 2 - Ms. Natalie S. Heck
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance. please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
## 510(k) Number (if known): K073499
### Device Name:
VerSys® Epoch® FullCoat Hip System
### Indications for Use:
The VerSys Epoch FullCoat Hip Prosthesis is indicated for:
- Patients suffering from severe hip pain and disability due to rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis of the femoral head, and nonunion of previous fractures of the femur.
- Patients with congenital hip dysplasia, protrusio acetabuli, or slipped capital femoral epiphysis.
- Patients suffering from disability due to previous fusion.
- Patients with acute femoral neck fractures.
This device is intended for cementless use only.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(Please do not write below this line - Continue on another page if needed)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Or Division of General, Restorative, and Neurological Devices
073499
510(k) Number
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.