← Product Code [JDI](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JDI) · K191449

# ZCA All-Poly Acetabular Cups (K191449)

_Zimmer, Inc. · JDI · Jul 16, 2020 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JDI/K191449

## Device Facts

- **Applicant:** Zimmer, Inc.
- **Product Code:** [JDI](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JDI.md)
- **Decision Date:** Jul 16, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3350
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The acetabular cup is indicated for cemented use in skeletally mature individuals undergoing primary or revision surgery for rehabilitating hips damaged as a result of noninflammatory degenerative joint disease (NIDID) or its composite diagnoses of osteoarthritis, avascular necrosis, traumatic arthritis, slipped capital epiphysis, fused hip, fracture of the pelvis, and diastrophic variant.

## Device Story

ZCA All-Poly Acetabular Cup is a hip prosthesis component for total hip arthroplasty; manufactured from conventional Ultra-High Molecular-Weight Polyethylene (UHMWPE). Device is intended for cemented fixation in the acetabulum. Used by orthopedic surgeons in clinical settings during primary or revision hip surgery. Functions as a bearing surface for femoral heads to restore hip joint function in patients with degenerative joint disease. Output is a physical implant that replaces damaged acetabular cartilage/bone. Benefits include restoration of joint mobility and reduction of pain. This submission is a retrospective 510(k) for cumulative changes to currently marketed devices.

## Clinical Evidence

No clinical data provided. Bench testing only: includes shelf-life testing (accelerated and real-time aging to 8 years), functional relationship analysis, wear performance, range of motion, and lever-out pull-through strength evaluation.

## Technological Characteristics

Material: Conventional Ultra-High Molecular-Weight Polyethylene (UHMWPE). Form factor: Acetabular cup prosthesis. Fixation: Cemented. Sterilization: Identical to predicates.

## Regulatory Identification

A hip joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and an acetabular resurfacing component made of ultra-high molecular weight polyethylene and is limited to those prostheses intended for use with bone cement (§ 888.3027).

## Predicate Devices

- NON-METAL BACKED ACETABULAR CUP FOR CEMENTED ([K901240](/device/K901240.md))
- ZCA All-Poly Acetabular Cup, Snap-In, Model 8005-946/958-32 ([K030153](/device/K030153.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

July 16, 2020

Zimmer. Inc. Caleb Barylski Specialist, Regulatory Affairs 1800 W. Center Street Warsaw, Indiana 46580

Re: K191449

Trade/Device Name: ZCA All-Poly Acetabular Cups Regulation Number: 21 CFR 888.3350 Regulation Name: Hip Joint Metal/Polymer Semi-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: JDI Dated: July 15, 2020 Received: July 16, 2020

Dear Caleb Barylski:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

FOR Vesa Vuniqi Acting Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K191449

Device Name ZCA All-Poly Acetabular Cups

#### Indications for Use (Describe)

The acetabular cup is indicated for cemented use in skeletally mature individuals undergoing primary or revision surgery for rehabilitating hips damaged as a result of noninflammatory degenerative joint disease (NIDID) or its composite diagnoses of osteoarthritis, avascular necrosis, traumatic arthritis, slipped capital epiphysis, fused hip, fracture of the pelvis, and diastrophic variant.

| Type of Use (Select one or both, as applicable)                                     |                                                                                    |
|-------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
| <span style="font-size:100%;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:100%;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(k) Summary

In accordance with 21 CFR §807.92 and the Safe Medical Devices Act of 1990, the following information is provided for the ZCA All-Poly Acetabular Cups 510(k) premarket notification. The submission was prepared in accordance with the FDA guidance document, 'Format for Traditional and Abbreviated 510(k)s,' issued on September 13, 2019.

| Sponsor:                           | Zimmer, Inc.<br>P.O. Box 708<br>Warsaw, IN 46581-0708<br>Establishment Registration Number: 1822565                                                                                                                                                                                                                                                                                                                                         |                                                                      |                                          |
|------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------|------------------------------------------|
| Contact Person:                    | Caleb Barylski<br>Regulatory Affairs Specialist<br>Telephone: (574-371-0250)<br>Email: Caleb.Barylski@zimmerbiomet.com                                                                                                                                                                                                                                                                                                                      |                                                                      |                                          |
| Date:                              | July 14, 2020                                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                      |                                          |
| Subject Device:                    | Trade Name: ZCA All-Poly Acetabular Cups<br>Common Name: Hip Prosthesis                                                                                                                                                                                                                                                                                                                                                                     |                                                                      |                                          |
|                                    | Classification Name:<br>• JDI– prosthesis, hip, semi-constrained, metal/polymer, cemented (21 CFR 888.3350)                                                                                                                                                                                                                                                                                                                                 |                                                                      |                                          |
| Predicate Device(s):               | K901240                                                                                                                                                                                                                                                                                                                                                                                                                                     | NON-METAL<br>BACKED<br>ACETABULAR<br>CUP FOR<br>CEMENTED             | MEDTEK<br>(purchased by<br>Zimmer, Inc.) |
|                                    | K030153                                                                                                                                                                                                                                                                                                                                                                                                                                     | ZCA All-Poly<br>Acetabular Cup,<br>Snap-In, Model<br>8005-946/958-32 | Zimmer, Inc.                             |
| Purpose and Device<br>Description: | ZCA is an acetabular cup prosthesis intended for use in<br>total hip arthroplasty. It is manufactured from<br>conventional Ultra-High Molecular-Weight Polyethylene<br>(UHMWPE).<br><br>The current submission is a retrospective 510(k) for devices<br>that are currently marketed in the U.S. Through a review of<br>the changes to the device system based on the current FDA<br>Guidance document "Deciding When to Submit a 510(k) for |                                                                      |                                          |

{4}------------------------------------------------

a Change to an Existing Device" (October 25, 2017), Zimmer Inc. has decided to submit a 510(k) for the cumulative changes.

## Intended Use and Indications for Use:

The acetabular cup is indicated for cemented use in skeletally mature individuals undergoing primary or revision surgery for rehabilitating hips damaged as a result of noninflammatory degenerative joint disease (NIDJD) or its composite diagnoses of osteoarthritis, avascular necrosis, traumatic arthritis, slipped capital epiphysis, fused hip, fracture of the pelvis, and diastrophic variant.

### Summary of Technological Characteristics:

The rationale for substantial equivalence is based on consideration of the following characteristics:

- Intended Use: Identical to predicates ●
- Indications for Use: Similar to predicates ●
- Materials: Substantially equivalent to predicates ●
- . Design Features: Substantially equivalent to predicates
- . Sterilization: Identical to predicates

## Summary of Performance Data (Nonclinical and/or Clinical)

#### ● Non-Clinical Tests:

- Performance Evaluation Performance testing O was completed on the line extension to determine equivalence to legally marketed devices.
- Shelf Life Accelerated and real time aging O testing conducted shows the sterile devices included in this submission have a shelf life of eight vears.
- Compatibility Functional Relationship O Analysis, wear performance, range of motion, and lever-out pull-through strength evaluation conducted shows the implants in this submission are compatible with femoral heads.
- 0 Clinical Tests:
	- Clinical test data is not provided for the subject o device.

Substantial Equivalence Conclusion

The data presented in this submission show that the changes do not affect the safety and/or effectiveness of the subject devices and that the subject devices will perform

{5}------------------------------------------------

in a substantially equivalent manner to the legally marketed predicate devices.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JDI/K191449](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JDI/K191449)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
