← Product Code [JDI](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JDI) · K153725

# Corin TaperFit Hip Stem (K153725)

_Corin U.S.A. Limited · JDI · Mar 25, 2016 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JDI/K153725

## Device Facts

- **Applicant:** Corin U.S.A. Limited
- **Product Code:** [JDI](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JDI.md)
- **Decision Date:** Mar 25, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3350
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The TaperFit Hip Stem is indicated for the relief of pain and restoration of hip function following the effects of femoral neck fracture, osteo, rheumatoid and inflammatory arthritis, post-traumatic disease effects, avascular necrosis and total hip revision. The TaperFit Hip Stem is indicated for hemi-arthroplasty when used in combination with Corin Bipolar Prosthesis hemi-arthroplasty femoral heads. The TaperFit Hip Stem is indicated for cemented, single use only.

## Device Story

TaperFit Hip Stem is a highly polished, double-tapered, collarless, stainless steel femoral hip stem; features 12/14 tapered male trunnion for modular femoral head assembly. Used in cemented hip arthroplasty and hemi-arthroplasty procedures. Operated by orthopedic surgeons in clinical/hospital settings. Device provides structural support for hip joint replacement; modular design allows compatibility with Corin Eurocone and Trinity CoCrMo femoral heads or Corin Bipolar Femoral Heads. Benefits include pain relief and restoration of hip function for patients with degenerative or traumatic hip conditions.

## Clinical Evidence

No clinical data. Bench testing only, including FEA, mechanical fatigue, static, rotational torque, fretting, corrosion, and range of motion testing, previously submitted in K142761, K992234, and K003666.

## Technological Characteristics

Highly polished, double-tapered, collarless femoral stem. Material: Wrought high nitrogen stainless steel (ISO 5832-9). Includes PMMA stem centralizer. 12/14 tapered male trunnion. Sizes 0-4; offsets 50mm, 45mm, 38mm, and CDH 36mm. Cemented, single-use.

## Regulatory Identification

A hip joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and an acetabular resurfacing component made of ultra-high molecular weight polyethylene and is limited to those prostheses intended for use with bone cement (§ 888.3027).

## Predicate Devices

- Corin TaperFit Hip Stem ([K142761](/device/K142761.md))
- Corin Taper-Fit Hip Stem ([K992234](/device/K992234.md))
- Corin Taper-Fit Hip Stem ([K003666](/device/K003666.md))

## Reference Devices

- Corin Eurocone ([K003666](/device/K003666.md))
- Trinity CoCrMo modular femoral heads ([K093472](/device/K093472.md), [K110087](/device/K110087.md), [K122305](/device/K122305.md), [K123705](/device/K123705.md), [K130128](/device/K130128.md), [K130343](/device/K130343.md), [K131647](/device/K131647.md))
- Corin Bipolar Femoral Head ([K925897](/device/K925897.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized representation of three human profiles facing right, stacked on top of each other. The profiles are rendered in a dark color, creating a sense of depth and unity. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the profiles.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 25, 2016

Corin USA Limited Ms. Diana L. Nader-Martone Regulatory Affairs Associate 5670 West Cypress Street, Suite C Tampa, Florida 33607

Re: K153725

Trade/Device Name: Taperfit Hip Stem Regulation Number: 21 CFR 888.3350 Regulation Name: Hip joint metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: JDI, JDG, KWY Dated: December 23, 2015 Received: December 28, 2015

Dear Ms. Nader-Martone:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

Page 2 - Ms. Diana L. Nader-Martone

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known) K153725

Device Name TaperFit Hip Stem

Indications for Use (Describe)

The TaperFit Hip Stem is indicated for the relief of pain and restoration following the effects of femoral neck fracture, osteo, rheumatoid and inflammatory arthritis, post-traumatic disease effects, avascular necrosis and total hip The TaperFit Hip Stem is indicated for hemi-arthroplasty when used in combination with Corin Bipolar revision. Prosthesis hemi-arthroplasty femoral heads.

The TaperFit Hip Stem is indicated for cemented, single use only.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

| 1. Applicant/Sponsor:<br>Distributor | Corin USA<br>5670 W. Cypress Street<br>Suite C<br>Tampa, Florida 33607<br>Establishment Registration No.: 1056629                                                                                                                                    |
|--------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 2. Contact Person:                   | Diana L. Nader-Martone<br>Regulatory Affairs Associate<br>Corin USA<br>813-977-4469<br>diana.nader-martone@coringroup.com<br><br>Kathy Trier<br>VP Global Regulatory and Clinical Affairs<br>Corin USA<br>813-977-4469<br>kathy.trier@coringroup.com |
| 3. Date:                             | March 24, 2016                                                                                                                                                                                                                                       |
| 4. Proprietary Name:                 | TaperFit Hip Stem                                                                                                                                                                                                                                    |
| 5. Common Name:                      | Hip Prosthesis                                                                                                                                                                                                                                       |
| 6. Product Code(s):                  | JDI, JDG, KWY                                                                                                                                                                                                                                        |

## 7. Classification Name:

21CFR 888.3350 – Hip joint metal/polymer semi-constrained cemented prosthesis 21CFR 888.3390 - Hip joint femoral (hemi-hip) metal/polymer cemented or uncemented prosthesis 21CFR 888.3360 – Hip joint femoral (hemi-hip) metallic cemented or uncemented prosthesis

## 8. Legally Marketed Devices to which Substantial Equivalence is claimed:

- Corin TaperFit Hip Stem (K142761) ●
- Corin Taper-Fit Hip Stem (K992234, K003666) .

{4}------------------------------------------------

## 9. Device Description:

The TaperFit Hip Stem is a highly polished, double tapered, collarless, stainless steel femoral hip stem featuring a 12/14 tapered male trunnion for assembly with modular femoral head components. The stem is manufactured from stainless steel in accordance with ISO 5832-9 -Implants for surgery -- Metallic materials -- Part 9: Wrought high nitrogen stainless steel and is provided with a polymethylmethacrylate (PMMA) Stem Centralizer. The stem is designed to be used in conjunction with Corin Eurocone (E100.XXX cleared in K003666) and Trinity CoCrMo modular femoral heads (E321.XXX cleared in K093472, K110087, K122305, K123705, K130128, K130343 and K131647). The stem is available in sizes, 0 through 4 with 3 offsets 50mm, 45mm, and 38mm, as well as a CDH option with a 36mm offset.

The TaperFit Hip Stem was originally cleared in K142761. This submission is being made to modify the indications for use to include hemi-arthroplasty when used with previously cleared Corin hemi-arthroplasty femoral heads as compatible components. The Corin CoCr Modular 26mm heads (E100.X26) have been cleared as compatible components of the Taper-Fit Hip stem (K003666). The Corin CoCr Modular 26mm heads (E100.X26) are part of the construct of the bipolar heads. Included in this submission is the Corin Bipolar Femoral Head (K925897) as a compatible component, which is indicated for hemi-arthroplasty.

## 10. Intended Use / Indications:

The TaperFit Hip Stem is indicated for the relief of pain and restoration of hip function following the effects of femoral neck fracture, osteo, rheumatoid and inflammatory arthritis, post-traumatic disease effects, avascular necrosis and total hip revision. The TaperFit Hip Stem is indicated for hemi-arthroplasty when used in combination with Corin Bipolar Prosthesis hemi-arthroplasty femoral heads.

The TaperFit Hip Stem is indicated for cemented, single use only.

# 11. Summary of Technologies / Substantial Equivalence:

The TaperFit Hip Stem, subject of this submission, and the predicate TaperFit Hip Stem (K142761) are identical in design, materials, coating, and sizes and similar in terms of intended use/indications for use. Additionally, the TaperFit Hip Stem, subject of this submission, and the predicates, Taper-Fit Hip Stem (K992234 and K003666) are identical in materials, and coating, and similar in terms of design, sizes, and intended use/indications for use. Based on these similarities, Corin believes that the TaperFit Hip Stem is substantially equivalent to the predicate devices.

# 12. Non-Clinical Testing:

{5}------------------------------------------------

Non-clinical testing and analysis included FEA, mechanical fatigue testing, static testing, rotational torque testing, fretting and corrosion testing, and range of motion testing. This testing was performed and submitted in the previous TaperFit and Taper-Fit Hip Stem submissions (K142761, K992234, and K003666).

# 13. Clinical Testing:

Clinical testing was not necessary in this Traditional 510(k).

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JDI/K153725](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JDI/K153725)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
