METS MODULAR TOTAL FEMUR

K121055 · Stanmore Implants Worldwide , Ltd. · JDI · Sep 19, 2012 · Orthopedic

Device Facts

Record IDK121055
Device NameMETS MODULAR TOTAL FEMUR
ApplicantStanmore Implants Worldwide , Ltd.
Product CodeJDI · Orthopedic
Decision DateSep 19, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3350
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The METS® Modular Total Femur is intended for the replacement of the total femoral bone.

Device Story

METS® Modular Total Femur is a single-use, modular orthopedic prosthesis for total femoral bone replacement. System comprises titanium trochanter section and shaft, femoral components (axle, bushes, circlip), bumper, and SMILES knee (rotating or fixed configuration). Trochanter trunnion interfaces with 28mm/32mm Cobalt Chrome femoral heads. Optional trochanters allow hard tissue attachment via plates, screws, or Ti/CoCr wire. Device is implanted by surgeons in clinical settings to restore joint function following tumor resection, trauma, or failed arthroplasty. Cemented fixation is required. Clinical benefit includes restoration of skeletal structure and joint mobility in patients with severe bone deficiency or disease.

Clinical Evidence

No clinical trials were conducted. Clinical performance conclusions were based on a review of published literature and post-market surveillance data for similar modular femoral systems.

Technological Characteristics

Materials: Titanium (Ti), Cobalt-Chromium-Molybdenum (CoCrMo), and Ultra High Molecular Weight Polyethylene (UHMWPE). Modular design with taper connections. Cemented fixation. Performance testing conducted per ASTM F1800-07 (knee fatigue/wear).

Indications for Use

Indicated for patients requiring limb salvage via radical bone resection/replacement; painful/disabled joints due to avascular necrosis, osteoarthritis, rheumatoid arthritis, or traumatic arthritis; correction of varus, valgus, or post-traumatic deformity; revision of failed osteotomy, arthrodesis, or previous joint replacement; ligament deficiencies; tumor resections; revision of failed total joint arthroplasty; trauma; and treatment of non-unions or femoral neck/trochanteric fractures of the proximal femur with head involvement unmanageable by other techniques. Components are for cemented use only and single use only.

Regulatory Classification

Identification

A hip joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and an acetabular resurfacing component made of ultra-high molecular weight polyethylene and is limited to those prostheses intended for use with bone cement (§ 888.3027).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K121055 (1/2) ### SECTION 5: 510(k) SUMMARY SECTION 5: **510(k) SUMMARY** Preparation Date: September 10, 2012 Trade Name: METS® MODULAR TOTAL FEMUR Common Name: Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented Classification Name: Hip joint metal/polymer semi-constrained cemented prosthesis (21 CFR 888.3350, Product Code JDI) Knee joint femorotibial metal/polymer constrained cemented prosthesis (21 CFR 888.3510, Product Code KRO) Applicant/Sponsor: Stanmore Implants Worldwide Ltd 210 Centennial Avenue Centennial Park Elstree WD6 3SJ Phone: +44 (0) 20 8238 6503 Facsimile: +44 (0) 20 8954 0351 Contact Person: Nancy MacDonald Manager of Regulatory Affairs Health Policy Associates Inc. Email: nmacdonald@healthpolicyassociates.com Tel: (781) 329-2993 Fax: (781) 329-2958 Equivalent to: JTS Extendible Implant, Stanmore Implants (K092138) Orthopaedic Salvage System (OSS); Biomet (K002757) Global Modular Replacement System (GMRS); Howmedica (Stryker) (K023087); Exactec Inc All Poly Acetabular Cup (K963313) and the Repiphysis Limb Salvage System Wright Medical (K021489). Device Description: The single use METS® Modular Total Femur is a standard modular system that is intended for the replacement of diseased or deficient bone in the femur. The system is intended for cemented use only and comprises titanium (Ti) components including a trochanter section, shaft. The trochanter trunnion is made to interchange with Stanmore Implants Worldwide Limited's, 28mm and 32mm Ø Cobalt Chrome femoral heads. The METS® Modular Total Femur is offered with an optional set of trochanters which are only to be used for hard tissue attachment using a plate and two screws, or Ti or cobalt chromium (CoCr) wire. A range of shafts, femoral component (including axle, bushes and circlip), bumper and the SMILES knee (available in 3 types of arrangements and in a rotating or fixed configuration). SEP 17 2012 {1}------------------------------------------------ | | The materials used in the manufacture of the systems include<br>titanium (Ti), cobalt-chromium-molybdenum (CoCrMo) and ultra<br>high molecular weight polyethylene (UHMWPE). | |-----------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Intended Use: | The METS® Modular Total Femur is intended for the replacement<br>of the total femoral bone. | | Indications for Use: | Limb salvage procedures where radical resection and replacement<br>of the bone is required<br>Painful and disabled joint resulting from avascular necrosis<br>osteoarthritis, rheumatoid arthritis or traumatic arthritis<br>Correction of varus, valgus or post traumatic deformity<br>Correction of revision of unsuccessful osteotomy, arthrodesis, or<br>previous joint replacement<br>Ligament deficiencies<br>Tumor resections<br>Revision of previously failed total joint arthroplasty<br>Trauma<br>Treatment of non-unions, femoral neck and trochanteric fracture of<br>the proximal femur with head involvement, unmanageable using<br>other techniques<br>METS® Modular Total Femur tibial and acetabular components are<br>for cemented use only.<br>METS® Modular Total Femur and their components are for single<br>use only. | | Performance Data:<br>(non-clinical and<br>clinical) | Non Clinical Testing<br>The results of the non-clinical performance testing demonstrate<br>that the device is safe and effective and substantially equivalent<br>to the predicate devices. The Performance testing included: knee<br>fatigue and wear test, disassembly force testing for the taper<br>connections, ASTM F1800-07. | | | Clinical Performance Conclusions<br>Clinical evaluation was carried out based upon published papers<br>and post market surveillance. | | Substantial<br>Equivalence: | The METS® Modular Total Femur is equivalent to the following<br>predicate devices: the JTS Extendible Implant (K092138); the<br>Biomet OSS (K002757); the Howmedica (Stryker) GMRS<br>(K023087); the Exactec All Poly Acetabular Cup (K963313) and<br>the Repiphysis Limb Salvage System (K021489). The<br>determination of substantial equivalence is based on the<br>similarity of the intended use, indications for use, design /<br>technological characteristics, materials of composition, method<br>of sterilization, performance data and clinical evaluation. | 2 . {2}------------------------------------------------ # DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is a stylized image of an eagle with its wings spread, which is a common symbol of the United States. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Stanmore Implants Worldwide, Limited % Health Policy Associates, Incorporated Ms. Nancy C. McDonald Manager, Regulatory Affairs 690 Canton Street, Suite 302 Westwood, Massachusetts 02090 SEP 17 2012 Re: K121055 Trade/Device Name: METS® MODULAR TOTAL FEMUR Regulation Number: 21 CFR 888.3350 Regulation Name: Hip joint metal/composite semi-constrained cemented prothesis Regulatory Class: Class II Product Code: JDI, KRO Dated: August 17, 2012 Received: August 17, 2012 Dear Ms. McDonald: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical {3}------------------------------------------------ #### Page 2 - Ms. Nancy C. McDonald device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Mark McMillan Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # SECTION 4: INDICATIONS FOR USE STATEMENT #### 510(k) Number (if known): ## METS® MODULAR TOTAL FEMUR. Device Name: Indications for Use: The METS® Modular Total Femur is intended for the replacement of the total femoral bone. 121055 Limb salvage procedures where radical resection and replacement of the bone is required Painful and disabled joint resulting from avascular necrosis osteoarthritis, rheumatoid arthritis or traumatic arthritis Correction of varus, valgus or post traumatic deformity Correction of revision of unsuccessful osteotomy, arthrodesis, or previous joint replacement Ligament deficiencies Tumor resections Revision of previously failed total joint arthroplasty Trauma Treatment of non-unions, femoral neck and trochanteric fracture of the proximal femur with head involvement, unmanageable using other techniques All of the METS® Modular Total Femur tibial and acetabular components are for cemented use only All of the METS® Modular Total Femur and their components are for single use only Prescription Use X (Per 21 CFR 801 Subpart D) AND/ OR Over-The-Counter Use (Per 21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE BAPinds for Division Sign-Division of Surgical, Orthopedic, and Restorative Devices 510(k) Number K121055
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...