← Product Code [JDC](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JDC) · K973357

# COONRAD/MORREY TOTAL ELBOW, NEW HINGE PIN (K973357)

_Zimmer, Inc. · JDC · Mar 2, 1998 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JDC/K973357

## Device Facts

- **Applicant:** Zimmer, Inc.
- **Product Code:** [JDC](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JDC.md)
- **Decision Date:** Mar 2, 1998
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3150
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Coonrad/Morrey Total Elbow is indicated for: - Post-traumatic lesions or bone loss contributing to elbow instability . - Ankylosed joints, especially in cases of bilateral ankylosis from causes other . than sepsis - Advanced rheumatoid or degenerative arthritis with incapacitating pain . - Instability or loss of motion when the degree of joint damage precludes less . radical procedures The candidate for total elbow arthroplasty should exhibit joint destruction which significantly compromises the activities of daily living. Patients with single-joint involvement (generally those with traumatic or degenerative arthritis) or significant lower extremity disability which require walking aids are less amenable to treatment than patients with advanced and predominantly upper extremity involvement. If possible, elbow replacement should be done after hip or knee surgery to avoid excessive stress to the prosthesis required by crutch walking during total hip or knee rehabilitation.

## Device Story

The Coonrad/Morrey Total Elbow is a prosthetic device designed for total elbow arthroplasty. It features a modified hinge pin assembly compared to the predicate Coonrad III Total Elbow. The device is intended for surgical implantation by an orthopedic surgeon to replace a damaged elbow joint. It functions as a mechanical hinge to restore joint stability and motion in patients with severe joint destruction. The primary clinical benefit is the reduction of pain and restoration of function for patients whose condition precludes less radical procedures. The device is used in a hospital/surgical setting.

## Clinical Evidence

Bench testing only. Performance testing was conducted to determine the force required to unlock the hinge pin assembly, confirming the device is safe and effective.

## Technological Characteristics

The device is a constrained, cemented elbow prosthesis. It utilizes a hinge pin assembly mechanism. No specific materials or software are described.

## Regulatory Identification

An elbow joint metal/polymer constrained cemented prosthesis is a device intended to be implanted to replace an elbow joint. It is made of alloys, such as cobalt-chromium-molybdenum, or of these alloys and of an ultra-high molecular weight polyethylene bushing. The device prevents dislocation in more than one anatomic plane and consists of two components that are linked together. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).

## Special Controls

*Classification.* Class II. The special controls for this device are:(1) FDA's:
(i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ”
(ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),”
(iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,”
(iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,”
(v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,”
(2) International Organization for Standardization's (ISO):
(i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-Aluminum 4-Vandium Alloy,”
(ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-Chromium-Molybdenum Casting Alloy,”
(iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-Chromium-Molybdenum Alloy,”
(iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,”
(v) ISO 5834-2:1998 “Implants for Surgery—Ultra High Molecular Weight Polyethylene—Part 2: Moulded Forms,”
(vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,”
(vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and
(viii) ISO 14630:1997 “Non-active Surgical Implants—General Requirements,”
(3) American Society for Testing and Materials':
(i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,”
(ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,”
(iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,”
(iv) F 981-93 “Practice for Assessment of Compatibility of Biomaterials (Nonporous) for Surgical Implant with Respect to Effect of Material on Muscle and Bone,”
(v) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,”
(vi) F 1108-97 “Specification for Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,”
(vii) F 1147-95 “Test Method for Tension Testing of Porous Metal Coatings, ” and
(viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”

## Predicate Devices

- Coonrad III Total Elbow ([K883665](/device/K883665.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

P.O. Box 708 Warsaw, IN 46581-0708 219 267-6131

Image /page/0/Picture/1 description: The image shows the logo for Zimmer. The logo consists of a large, bold letter "Z" inside of a circle. Below the circle, the word "zimmer" is written in a lowercase, sans-serif font.

K973357

MAR - 2 1998

# Summary of Safety and Effectiveness - Summary of Safety and Elbow, New Hinge Pin

- Submitted by: .
Zimmer, Inc. P.O. Box 708 Warsaw, IN 46581-0708

- Prepared by: .
Charlene Brumbaugh Specialist Global Regulatory Affairs Telephone: 219-372-4962 Telefax: 219-372-4605

- Date: .
September 4, 1997

- Trade Name: .
Coonrad/Morrey Total Elbow

### Common Name: .

Elbow Prosthesis

## Classification Name: .

constrained, Cemented Prosthesis, Elbow,

### Predicate Devices: .

Coonrad III Total Elbow, marketed by Zimmer

Image /page/0/Picture/20 description: The image shows a geometric design. The design is composed of six triangles arranged around a central point. Each triangle is further divided into three smaller triangles, creating a star-like pattern within each larger triangle. The overall shape of the design is hexagonal.

{1}------------------------------------------------

# Summary of Safety and Effectiveness Coonrad/Morrey Total Elbow, New Hinge Pin (Continued)

### Device Description .

The Coonrad/Morrey Total Elbow is closely based on the Coonrad III Total Elbow (K883665) cleared by FDA on February 3, 1989, with several exceptions.

#### Intended Use .

The Coonrad/Morrey Total Elbow is indicated for:

- Post-traumatic lesions or bone loss contributing to elbow instability .
- Ankylosed joints, especially in cases of bilateral ankylosis from causes other . than sepsis
- Advanced rheumatoid or degenerative arthritis with incapacitating pain .
- Instability or loss of motion when the degree of joint damage precludes less . radical procedures

The candidate for total elbow arthroplasty should exhibit joint destruction which significantly compromises the activities of daily living. Patients with single-joint involvement (generally those with traumatic or degenerative arthritis) or significant lower extremity disability which require walking aids are less amenable to treatment than patients with advanced and predominantly upper extremity involvement. If possible, elbow replacement should be done after hip or knee surgery to avoid excessive stress to the prosthesis required by crutch walking during total hip or knee rehabilitation.

#### Performance Data .

Performance testing was conducted to determine force required to unlock the hinge pin assembly. Results indicate the product is safe and effective.

{2}------------------------------------------------

Image /page/2/Picture/2 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features the department's name in a circular arrangement around the perimeter, reading "DEPARTMENT OF HEALTH & HUM. SERVICES - USA". In the center is a stylized emblem consisting of three overlapping, curved shapes that resemble a caduceus without the snake. The emblem is rendered in black, contrasting with the white background of the seal.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAR - 2 1998

Ms. Charlene Brumbaugh ·Specialist Global Regulatory Affairs Zimmer P.O. Box 708 46581-0708 Warsaw, Indiana

Re: K973357 Coonrad/Morrey Total Elbow, New Hinge Pin Trade Name: Regulatory Class: III Product Code: JDC Dated: September 4, 1997 Received: "September 8, 1997

Dear Ms. Brumbaugh:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A-------substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical General regulation (21 CFR Part 820) and that, Devices: through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. Please note: concerning your device in the Federal Register. this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.

{3}------------------------------------------------

Page 2 - Ms. Charlene Brumbaugh

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to
premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Colfa M. Witton, Ph.D., M

Witten, Ph.D., M.D. M. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

510(k) Number (if known): K973357

Device Name: Coonrad/Morrey Total Elbow, New Hinge Pin

Indications For Use:

- Post-traumatic lesions or bone loss contributing to elbow instability ●
- Ankylosed joints, especially in cases of bilateral ankylosis from causes other than . sepsis
- Advanced rheumatoid or degenerative arthritis with incapacitating pain
- . Instability or loss of motion when the degree of joint damage precludes less radical procedures

The candidate for total elbow arthroplasty should exhibit joint destruction which significantly compromises the activities of daily living. Patients with single joint involvement (generally those with traumatic or degenerative arthritis) or significant lower extremity disability which require walking aids are less amenable to treatment than patients with advanced and predominantly upper extremity involvement. If possible, elbow replacement should be done after hip or knee surgery to avoid excessive stress to the prosthesis required by crutch walking during total hip or knee rehabilitation.

# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

| Division Sign-Off                       |         |
|-----------------------------------------|---------|
| Division of General Restorative Devices |         |
| 510(k) Number                           | K973357 |

| Prescription Use | X (Per 21 CFR 801.109) |
|------------------|------------------------|
|------------------|------------------------|

OR

| Over-The-Counter Use | (Optional Format 1-2-96) |
|----------------------|--------------------------|
|----------------------|--------------------------|

RA08702K.510
ਪ੍ਰੀ

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JDC/K973357](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/JDC/K973357)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
