ACCLAIM TOTAL ELBOW SYSTEM

K060696 · DePuy Orthopaedics, Inc. · JDC · Apr 14, 2006 · Orthopedic

Device Facts

Record IDK060696
Device NameACCLAIM TOTAL ELBOW SYSTEM
ApplicantDePuy Orthopaedics, Inc.
Product CodeJDC · Orthopedic
Decision DateApr 14, 2006
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3150
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Acclaim™ Total Elbow System is indicated to reduce pain and improve the function and mobility of the affected joint in patients with a painful arthritic joint due to osteoarthritis, theumatoid arthritis, or post traumatic arthritis and pathological fractures of the distal humerus in which adequate bone stock exists for the fixation of prosthetic components. Total Elbow replacement may be considered for younger patients, if, in the opinion of the surgeon, an unequivocal indication for elbow replacement outweighs the risks associated with the age of the patient, and if limited demands regarding activity and elbow joint loading can be assured. This included patients for whom an immediate gain of elbow mobility may lead to an expectation of significant improvement in the quality of their lives. The Acclaim™ Total Elbow System is intended for cemented use only.

Device Story

The Acclaim™ Total Elbow System is a replacement hinge pin assembly for total elbow arthroplasty. It addresses hinge pin disassociation in linked (constrained) systems. The assembly consists of a polyethylene humeral yoke and locking sleeve, a metal locking pin, ulnar bearing, ulnar bearing screw, washer, and wire. It is used in conjunction with existing well-fixed humeral and ulnar components. The device incorporates a washer and cross-pin locking mechanism. It is implanted by an orthopedic surgeon in a clinical/surgical setting. The device provides mechanical stability to the elbow joint, aiming to reduce pain and restore mobility in patients with arthritis or distal humerus fractures. It is intended for cemented use only.

Clinical Evidence

No clinical data provided; substantial equivalence is based on design, material, and performance standard conformance.

Technological Characteristics

Materials include polyethylene (humeral yoke, sleeve, ulnar bearing) and metal (locking pin, screw, washer, wire). Features a cross-pin locking mechanism. Designed for cemented fixation. Sterilization and packaging methods are consistent with predicate devices.

Indications for Use

Indicated for patients with painful arthritic joints (osteoarthritis, rheumatoid arthritis, post-traumatic arthritis) or pathological fractures of the distal humerus with adequate bone stock for fixation. May be considered for younger patients if benefits outweigh risks and activity demands are limited.

Regulatory Classification

Identification

An elbow joint metal/polymer constrained cemented prosthesis is a device intended to be implanted to replace an elbow joint. It is made of alloys, such as cobalt-chromium-molybdenum, or of these alloys and of an ultra-high molecular weight polyethylene bushing. The device prevents dislocation in more than one anatomic plane and consists of two components that are linked together. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).

Special Controls

*Classification.* Class II. The special controls for this device are:(1) FDA's: (i) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Devices—Part I: Evaluation and Testing,’ ” (ii) “510(k) Sterility Review Guidance of 2/12/90 (K90-1),” (iii) “Guidance Document for Testing Orthopedic Implants with Modified Metallic Surfaces Apposing Bone or Bone Cement,” (iv) “Guidance Document for the Preparation of Premarket Notification (510(k)) Application for Orthopedic Devices,” (v) “Guidance Document for Testing Non-articulating, ‘Mechanically Locked’ Modular Implant Components,” (2) International Organization for Standardization's (ISO): (i) ISO 5832-3:1996 “Implants for Surgery—Metallic Materials—Part 3: Wrought Titanium 6-Aluminum 4-Vandium Alloy,” (ii) ISO 5832-4:1996 “Implants for Surgery—Metallic Materials—Part 4: Cobalt-Chromium-Molybdenum Casting Alloy,” (iii) ISO 5832-12:1996 “Implants for Surgery—Metallic Materials—Part 12: Wrought Cobalt-Chromium-Molybdenum Alloy,” (iv) ISO 5833:1992 “Implants for Surgery—Acrylic Resin Cements,” (v) ISO 5834-2:1998 “Implants for Surgery—Ultra High Molecular Weight Polyethylene—Part 2: Moulded Forms,” (vi) ISO 6018:1987 “Orthopaedic Implants—General Requirements for Marking, Packaging, and Labeling,” (vii) ISO 9001:1994 “Quality Systems—Model for Quality Assurance in Design/Development, Production, Installation, and Servicing,” and (viii) ISO 14630:1997 “Non-active Surgical Implants—General Requirements,” (3) American Society for Testing and Materials': (i) F 75-92 “Specification for Cast Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implant Material,” (ii) F 648-98 “Specification for Ultra-High-Molecular-Weight Polyethylene Powder and Fabricated Form for Surgical Implants,” (iii) F 799-96 “Specification for Cobalt-28 Chromium-6 Molybdenum Alloy Forgings for Surgical Implants,” (iv) F 981-93 “Practice for Assessment of Compatibility of Biomaterials (Nonporous) for Surgical Implant with Respect to Effect of Material on Muscle and Bone,” (v) F 1044-95 “Test Method for Shear Testing of Porous Metal Coatings,” (vi) F 1108-97 “Specification for Titanium-6 Aluminum-4 Vanadium Alloy Castings for Surgical Implants,” (vii) F 1147-95 “Test Method for Tension Testing of Porous Metal Coatings, ” and (viii) F 1537-94 “Specification for Wrought Cobalt-28 Chromium-6 Molybdenum Alloy for Surgical Implants.”

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 060696 p. APR 1 4 2006 # Summary Acclaim™ Total Elbow System | 510(k) SUMMARY | | |--------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | NAME OF FIRM: | DePuy Orthopaedics, Inc.<br>700 Orthopaedic Drive<br>Warsaw, IN 46580 | | 510(K) CONTACT: | Natalie S. Heck<br>Manager, Regulatory Affairs<br>DePuy Orthopaedics, Inc.<br>PO Box 988<br>700 Orthopaedic Drive<br>Warsaw, IN 46581-0988 | | TRADE NAME: | Acclaim™ Total Elbow System | | COMMON NAME: | Elbow Prosthesis | | CLASSIFICATION: | When used as a semi-constrained (unlinked elbow), it is<br>a Class II device per 21 CFR §888.3160<br>When used as a constrained (linked), it is a Class II<br>device per 21 CFR §888.3150 | | DEVICE PRODUCT CODE: | 87 JDB - Prosthesis, Elbow, Semi-Constrained,<br>Cemented (Class II)<br>87 JDC - Prosthesis, Elbow, Constrained, Cemented<br>(Class II) | | SUBSTANTIALLY EQUIVALENT<br>DEVICES: | Acclaim™ Total Elbow System - K992656<br>(formerly cleared as DePuy Total Elbow System)<br>Mark II Elbow System - K872084 | #### DEVICE DESCRIPTION: #### C. Indications for Use: The Acclaim™ Total Elbow System is indicated to reduce pain and improve the function and mobility of the affected joint in patients with a painful arthritic joint due to osteoarthritis, theumatoid arthritis, or post traumatic arthritis and pathological fractures of the distal humerus in which adequate bone stock exists for the fixation of prosthetic components. Total Elbow replacement may be considered for younger patients, if, in the opinion of the surgeon, an unequivocal indication for elbow replacement outweighs the risks associated with the age of the patient, and if limited demands regarding activity and elbow joint loading can be assured. This included patients for whom an immediate gain of elbow ## 0000006 {1}------------------------------------------------ mobility may lead to an expectation of significant improvement in the quality of their lives. KO66666 The Acclaim™ Total Elbow System is intended for cemented use only. ### D. Device Description: The Acclaim™ Total Elbow System replacement hinge pin assembly is designed as a replacement of a linked (constrained) elbow hinge pin assembly due to hinge pin disassociation. When the Acclaim Total Elbow System is implanted as a linked system. it is held together with the linked ulnar component and pin assembly, and is used when there is poor bone stock. The Acclaim™ Total Elbow System replacement hinge pin assembly is comprised of a polyethylene humeral yoke and locking sleeve, with a metal locking pin, ulnar bearing, ulnar bearing screw, washer, and wire. The replacement hinge pin assembly is used in conjunction with well-fixed humeral and ulnar components. Acclaim™ Total Elbow System replacement hinge pin assembly design includes modifications to the ulnar bearing, polyethylene sleeve, and locking pin. In addition, a washer and cross-pin locking mechanism has been included to the assembly using the same design as the Mark II Total Elbow System locking mechanism, previously cleared in K872084 dated June 25, 1987. ### E. Substantial Equivalence: The substantial equivalence of the Acclaim™ Total Elbow System replacement hinge pin assembly is substantiated by its similarity in indications for use, design, materials, sterilization and packaging to the current Acclaim™ Total Elbow (formerly cleared as DePuy Total Elbow System, K992656), DePuy Mark II Elbow (K872084) hinge pin assemblies. The determination of substantial equivalence for this device was based on a detailed device description, and conformance with voluntary performance standards. {2}------------------------------------------------ Image /page/2/Picture/2 description: The image shows the seal of the Department of Health and Human Services (HHS). The seal is circular and contains the words "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" around the perimeter. In the center of the seal is a stylized image of three human figures, representing the department's mission to protect the health of all Americans and provide essential human services. APR 1 4 2006 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 DePuy Orthopaedics, Inc. c/o Ms. Natalie S. Heck Manager, Regulatory Affairs P.O. Box 988 700 Orthopaedic Drive Warsaw, Indiana 46581-0988 Re: K060696 Trade/Device Name: Acclaim Total Elbow System Regulation Number: 21 CFR 888.3150 Regulation Name: Elbow joint metal/polymer constrained cemented prosthesis Regulatory Class: Class II Product Codes: JDC, JDB Dated: March 15, 2006 Received: March 16, 2006 Dear Ms. Heck: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ Page 2 - Ms. Natalie S. Heck This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours. . Melli Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indications for Use 510(k) Number (if known): K0606 Device Name: Acclaim™ Total Elbow System Indications for Use: The Acclaim™ Total Elbow System is indicated to reduce pain and improve the function and mobility of the affected joint in patients with a painful arthritic joint due to osteoarthritis, theumatoid arthritis, or post traumatic arthritis and pathological fractures of the distal humerus in which adequate bone stock exists for the fixation of prosthetic components. Total Elbow replacement may be considered for younger patients, if, in the opinion of the surgeon, an unequivocal indication for elbow replacement outweighs the risks associated with the age of the patient , and if limited demands regarding activity and elbow joint loading can be assured. This included patients for whom an immediate gain of elbow mobility may lead to an expectation of significant improvement in the quality of their lives. The Acclaim™ Total Elbow System is intended for cemented use only. | Prescription Use | <b>X</b> (Part 21 CFR 801 Subpart D) | |------------------|--------------------------------------| |------------------|--------------------------------------| AND/OR | Over-The-Counter Use | __________ (21 CFR 807 Subpart C) | |----------------------|-----------------------------------| |----------------------|-----------------------------------| (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) el mal (Division Sign-Off) Division of General, Restorative, and Neurological Devices | Page | ____ of ____ | |------|--------------| |------|--------------| 510(k) Number_ko60696 0000004
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...