K252758 · Orthonovis, Inc. · HWC · Oct 2, 2025 · Orthopedic
Device Facts
Record ID
K252758
Device Name
Cannulated Screw and Kirschner (K wire) System
Applicant
Orthonovis, Inc.
Product Code
HWC · Orthopedic
Decision Date
Oct 2, 2025
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Intended Use of the Cannulated Screw and Kirschner (Kwire) System is for the treatment and fixation of bone fractures and osteotomies of various bones including the clavicle, acetabulum, pelvis, scapula, humerus, radius, ulna, femur, tibia, phalanges, carpals, metacarpals, tarsals, metatarsals and fibula.
Device Story
System consists of cannulated, self-tapping, self-drilling screws and Kirschner wires (K-wires) used for bone fixation. Screws feature cortico/cancellous or cancellous threads; available in partial or full thread configurations. K-wires are threaded, spaded, or blunt. Device used in clinical/surgical settings by orthopedic surgeons to stabilize bone fractures or osteotomies. K-wires serve as guides for screw placement. Output is mechanical fixation of bone segments. Benefits include stabilization of fractures to facilitate healing.
Clinical Evidence
No clinical data submitted. Bench testing only, consisting of engineering analysis including cross-sectional analysis and axial pullout testing performed in accordance with Chapman et al. (Orthopedic Non-Spinal Metallic Bone Screws and Washers – Performance Criteria for Safety and Performance Based Pathway).
Technological Characteristics
Materials: Titanium Alloy or 316L Stainless Steel. Design: Cannulated, self-tapping, self-drilling screws; threaded, spaded, or blunt K-wires (0.8-1.4mm diameter, 100-200mm length). Principle: Mechanical bone fixation fastener. Energy source: None (manual).
Indications for Use
Indicated for treatment and fixation of bone fractures and osteotomies in patients requiring orthopedic stabilization of the clavicle, acetabulum, pelvis, scapula, humerus, radius, ulna, femur, tibia, phalanges, carpals, metacarpals, tarsals, metatarsals, and fibula.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
Cannulated Screw and Kirschner (K wire) System (K130613)
BPS - Bone Fragment Fixation Plates, Screws, and Washers (K250055)
TITANEX™ MICROBEAM Screw System, TITANEX™ ARTEMIS Screw System (K231504)
Related Devices
K130613 — CANNULATED SCREW AND KIRSCHNER (KWIRE) SYSTEM · Laura Cattabriga · Jan 22, 2014
K161259 — KLS Martin Cannulated Headless Screws · KLS Martin L.P. · Dec 19, 2016
K060428 — CSS CANNULATED SCREW SYSTEM · Orthohelix Surgical Designs, Inc. · Mar 7, 2006
K003496 — CANNULATED SCREW SYSTEM · Pioneer Surgical Technology · Feb 8, 2001
K143460 — Cannulated Screw System · Wrightmedicaltechnologyinc · Jan 23, 2015
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
October 2, 2025
OrthoNovis, Inc.
% Tawney Schwarz
Regulatory Consultant
Simple Path LLC
712 Lyndhurst Street
Unit 321
Dunedin, Florida 34698
Re: K252758
Trade/Device Name: Cannulated Screw and Kirschner (K wire) System
Regulation Number: 21 CFR 888.3040
Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener
Regulatory Class: Class II
Product Code: HWC, HTY, HTN
Dated: September 29, 2025
Received: September 29, 2025
Dear Tawney Schwarz:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252758 - Tawney Schwarz
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K252758 - Tawney Schwarz
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Shumaya Ali -S
Shumaya Ali, M.P.H.
Assistant Director
DHT6C: Division of Restorative,
Repair, and Trauma Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K252758
Device Name
Cannulated Screw and Kirschner (K wire) System
Indications for Use (Describe)
The Intended Use of the Cannulated Screw and Kirschner (Kwire) System is for the treatment and fixation of bone fractures and osteotomies of various bones including the clavicle, acetabulum, pelvis, scapula, humerus, radius, ulna, femur, tibia, phalanges, carpals, metacarpals, tarsals, metatarsals and fibula.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
> "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}
510(k) #: K252758
510(k) Summary
Prepared on: 2025-09-29
# Contact Details
21 CFR 807.92(a)(1)
| Applicant Name | OrthoNovis, Inc. |
| --- | --- |
| Applicant Address | 1 Hargrove Grade Suite 2f Palm Coast FL 32137 United States |
| Applicant Contact Telephone | 352-807-3306 |
| Applicant Contact | Mr. Ken West |
| Applicant Contact Email | kwest@orthonovis.com |
| Correspondent Name | Simple Path LLC |
| Correspondent Address | 712 Lyndhurst St Unit 321 Dunedin FL 34698 United States |
| Correspondent Contact Telephone | 910-515-0918 |
| Correspondent Contact | Mrs. Tawney Schwarz |
| Correspondent Contact Email | tawney@simplepath.solutions |
# Device Name
21 CFR 807.92(a)(2)
| Device Trade Name | Cannulated Screw and Kirschner (K wire) System |
| --- | --- |
| Common Name | Smooth or threaded metallic bone fixation fastener |
| Classification Name | Screw, Fixation, Bone |
| Regulation Number | 888.3040 |
| Product Code(s) | HWC, HTY, HTN |
# Legally Marketed Predicate Devices
21 CFR 807.92(a)(3)
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| --- | --- | --- |
| K130613 | Cannulated Screw and Kirschner (K wire) System | HTN |
| K250055 | BPS - Bone Fragment Fixation Plates, Screws, and Washers | HWC |
| K231504 | TITANEX™ MICROBEAM Screw System, TITANEX™ ARTEMIS Screw System | HWC |
# Device Description Summary
21 CFR 807.92(a)(4)
Cannulated Screws and Kwires are a self-tapping, self-drilling screw with a cortico/cancellous or cancellous thread that can be guided into position by Kwire placement. Partial or fully threaded screws are available in various different lengths and diameters to provide fixation in various size bones. The screws are made of Titanium Alloy or Stainless Steel. The Kirschner Wires (Kwires) are threaded, spaded or blunt ranging from 0.8 to 1.4mm in diameter and ranging from 100 to 200mm in length and made of 316L Stainless Steel.
Page 1 of 2
{5}
Page 2 of 2
| Intended Use/Indications for Use | 21 CFR 807.92(a)(5) |
| --- | --- |
| The Intended Use of the Cannulated Screw and Kirschner (Kwire) System is for the treatment and fixation of bone fractures and osteotomies of various bones including the clavicle, acetabulum, pelvis, scapula, humerus, radius, ulna, femur, tibia, phalanges, carpals, metacarpals, tarsals, metatarsals and fibula. | |
| Indications for Use Comparison | 21 CFR 807.92(a)(5) |
| Indications for use are similar. | |
| Technological Comparison | 21 CFR 807.92(a)(6) |
| Technological characteristics are similar to the provided primary and reference predicate devices, including material, design, chemical composition and principle of operation. | |
| Non-Clinical and/or Clinical Tests Summary & Conclusions | 21 CFR 807.92(b) |
| The following was provided in an engineering analysis:
-Cross sectional Analysis
-Axial Pullout per Chapman et al. (Orthopedic Non-Spinal Metallic Bone Screws and Washers – Performance Criteria for Safety and Performance Based Pathway)
No clinical tests were submitted.
Based on the engineering analysis of device characteristics, it can be concluded that the subject device does not raise different questions of safety and effectiveness compared to the predicate devices. The similar indications for use, technological characteristics, and performance characteristics for the proposed are assessed to be substantially equivalent to the predicate devices. | |
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.