Phoenix Sinus Tarsi Stent System

K251382 · Astra Orthomed, Inc. · HWC · Aug 27, 2025 · Orthopedic

Device Facts

Record IDK251382
Device NamePhoenix Sinus Tarsi Stent System
ApplicantAstra Orthomed, Inc.
Product CodeHWC · Orthopedic
Decision DateAug 27, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3040
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

Phoenix Sinus Tarsi Stent is an implant stabilization device used in the treatment of talotarsal joint instability in adult and pediatric patients four years of age and older. The stent is designed to stabilize the talus to prevent excessive anterior, and/or medial, and/or plantarflexion of the talus, while allowing normal talotarsal joint motion.

Device Story

Phoenix Sinus Tarsi Stent is an extra-osseous talotarsal stabilization (EOTTS) implant; inserted into sinus tarsi to stabilize talus; prevents excessive anterior, medial, or plantarflexion of talus while maintaining normal joint motion. System includes implant and corresponding insertion instrumentation. Used by surgeons in clinical settings. Device provides mechanical stabilization to treat talotarsal joint instability; benefits patient by correcting abnormal talar motion.

Clinical Evidence

Bench testing only. Testing included screw pullout (ASTM F543), cantilever bending (ASTM F2193), endotoxin (AAMI ST72, USP <85>), sterilization (ISO 11137), cytotoxicity (ISO 10993-5), biocompatibility risk assessment, and packaging shelf-life (ISO 11607-1, ASTM F88, F2096, F1886). All tests met predetermined acceptance criteria.

Technological Characteristics

Material: Ti-6Al-4V ELI per ASTM F136. Form factor: 5 sizes with varying diameters. Manufacturing: Subtractive techniques. Sterilization: ISO 11137-1/2. Non-pyrogenic type IIB EOTTS device.

Indications for Use

Indicated for treatment of talotarsal joint instability in adult and pediatric patients 4 years of age and older.

Regulatory Classification

Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD &amp; DRUG ADMINISTRATION August 27, 2025 Astra OrthoMed, Inc. % Justin Gracyalny Regulatory Affairs Program Manager Secure BioMed Evaluations 7828 Hickory Flat Hwy Woodstock, Georgia 30188 Re: K251382 Trade/Device Name: Phoenix Sinus Tarsi Stent System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: HWC Dated: July 28, 2025 Received: July 28, 2025 Dear Justin Gracyalny: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food &amp; Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K251382 - Justin Gracylny Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K251382 - Justin Graycalny Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Shumaya Ali -S Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair, and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} Phoenix Sinus Tarsi Stent System Page 10 of 49 | Indications for Use | | | | --- | --- | --- | | Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K251382 | ? | | Please provide the device trade name(s). | | ? | | Phoenix Sinus Tarsi Stent System | | | | Please provide your Indications for Use below. | | ? | | Phoenix Sinus Tarsi Stent is an implant stabilization device used in the treatment of talotarsal joint instability in adult and pediatric patients four years of age and older. The stent is designed to stabilize the talus to prevent excessive anterior, and/or medial, and/or plantarflexion of the talus, while allowing normal talotarsal joint motion. | | | | Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? | {4} Page 1 of 2 K251382 510(k) SUMMARY: Phoenix Sinus Tarsi Stent System | Date Prepared | August 21, 2025 | | --- | --- | | Sponsor | Astra OrthoMed 330 Franklin Road Suite 135A-213 Brentwood, TN 37027 586-604-9462 | | 510(k) Contact | Secure BioMed Evaluations Justin Gracylny, MSE 7828 Hickory Flat Highway, Suite 120 Woodstock, GA 30188 770-837-2681 Regulatory@SecureBME.com | | Trade Name | Phoenix Sinus Tarsi Stent System | | Common Name | Sinus Tarsi Implant | | Code –Classification | HWC - Smooth or threaded metallic bone fixation fastener (21 CFR 888.3040) | | Primary Predicate | K142534 GraMedica LLC. HyProCure II | | Reference Device | K042030 GraMedica LLC. HyProCure Subtalar Implant System K111834 Trilliant Surgical LTD Disco Subtalar Implant | | Device Description | The Phoenix Sinus Tarsi Stent is a non-pyrogenic type IIB extra-osseous talotarsal stabilization (EOTTS) device used in the treatment of talotarsal joint instability per Graham Et al. EOTTS classification system. The stent system consists of an implant designed to be inserted into the sinus tarsi and has corresponding instrumentation to facilitate the insertion. All stents are manufactured from Ti-6Al-4V ELI per ASTM F136 and have 5 sizes with varying diameter. | | Indications for Use Statement | Phoenix Sinus Tarsi Stent is an implant stabilization device used in the treatment of talotarsal joint instability. The stent is designed to stabilize the talus to prevent excessive anterior, and/or medial, and/or plantarflexion of the talus, while allowing normal talotarsal joint motion. | ## Comparison of Technological Characteristics The Phoenix Sinus Tarsi Stent System is substantially equivalent to the legally marketed predicate devices in terms of intended use, design, materials of construction, mechanical safety and performance. The subject device indications for use include minor differences for clarification. Any differences in terminology do not create a change in the device and are simply clarification of terminology as compared to predicate device. There are no significant technological differences between the subject and predicate device. The subject device uses the same material as the predicate and reference devices, offers sizes within the ranges offered by the predicate and reference devices, and all devices are manufactured using subtractive techniques. Any minor geometric differences are supported by performance testing. {5} Page 2 of 2 # Non-Clinical Performance Testing Summary Performance testing for Phoenix Sinus Tarsi Stent System include: - Screw Pullout Testing per ASTM F543 - Cantilever Bending Testing per ASTM 2193 - Endotoxin Testing per AAMI ST72 and USP &lt;85&gt; - Sterilization Testing per ISO 11137-1, ISO 11137-2 - Cytotoxicity Testing per ISO 10993-5 - Biocompatibility Risk Assessment - Packaging Shelf-Life Performance Testing per ISO 11607-1, ASTM F88/F88M, ASTM F2096, ASTM F1886/1886M All testing showed the subject device performed as intended. All testing met applicable predetermined acceptance criteria. All testing supports the subject device is substantially equivalent to the predicate device. # Conclusions Based on the similarities of the intended use/indications for use, technological and functional characteristic, and the results of the non-clinical performance testing, the subject device is substantially equivalent to the legally marketed predicate device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...