← Product Code [HWC](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HWC) · K232251

# APTUS Cannulated Compression Screws Line Extension, APTUS K-Wire System Line Extension (K232251)

_Medartis AG · HWC · Nov 3, 2023 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HWC/K232251

## Device Facts

- **Applicant:** Medartis AG
- **Product Code:** [HWC](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HWC.md)
- **Decision Date:** Nov 3, 2023
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3040
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

APTUS® Cannulated Compression Screws and headed Cannulated Compression Screws are indicated for the treatment of fractures, osteotomies and arthrodesis of bones with the appropriate screw size in the shoulder, elbow, wrist, hand, knee, and the foot and ankle. APTUS® K-Wire System is intended for use in fixation of bone fractures, for bone reconstruction, and as guide pins for insertion of other implants.

## Device Story

Line extension of APTUS Cannulated Compression Screws and K-Wire System; provides additional screw lengths, fully threaded screw options, and longer K-wire versions. Screws feature SpeedTip triangular tip for reduced insertion torque and internal hexalobular instrument interface. K-wires provided in sterile or non-sterile configurations. Used by orthopedic surgeons for bone fixation, reconstruction, and arthrodesis. Screws transform rotational force into compression across bone segments. K-wires serve as guide pins for implant insertion. Devices are implanted during orthopedic procedures; provide mechanical stability to bone fragments to facilitate healing.

## Clinical Evidence

No clinical data provided. Substantial equivalence supported by non-clinical bench testing, including mechanical testing per ASTM F543-17 and ASTM F1839-08(2021).

## Technological Characteristics

Cannulated compression screws: Titanium alloy (ASTM F136). K-wires: Stainless steel (ASTM F138). Features: Triangular SpeedTip, internal hexalobular instrument interface. Sterilization: Moist heat (end-user) or X-ray beam irradiation (terminal).

## Regulatory Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- APTUS® headed Cannulated Compression Screws ([K202589](/device/K202589.md))
- APTUS Cannulated Compression Screws 5.0.7.0 ([K133460](/device/K133460.md))
- APTUS® Cannulated Compression Screws ([K110658](/device/K110658.md))
- APTUS® K-Wire System ([K092038](/device/K092038.md))
- APTUS 3.5 TriLock Straight Plates ([K230971](/device/K230971.md))
- Synthes Sterile 3.5 mm and 4.0 mm Cannulated Screws ([K963192](/device/K963192.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Build Correspondence

November 3, 2023

DRUC

Medartis AG Claudia Santis Head of Regulatory Affairs Hochbergerstrasse 60E Basel, Basel-Stadt CH-4057 Switzerland

Re: K232251

Trade/Device Name: APTUS Cannulated Compression Screws Line Extension, APTUS K-Wire System Line Extension Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: HWC. HTY Dated: July 28, 2023 Received: July 28, 2023

Dear Claudia Santis:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

### K232251 - Claudia Santis

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product: and 21 CFR 820.100. Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products: and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

# Shumaya Ali -S

Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health 

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

## Indications for Use

510(k) Number (if known) K232251

Device Name

APTUS® Cannulated Compression Screws Line Extension

Indications for Use (Describe)

APTUS® Cannulated Compression Screws and headed Cannulated Compression Screws are indicated for the treatment of fractures, osteotomies and arthrodesis of bones with the appropriate screw size in the shoulder, elbow, wrist, hand, knee, and the foot and ankle.

| Type of Use (Select one or both, as applicable) | <span> <span style="padding-right: 20px;"> <span style="vertical-align: middle;">☑</span>     Prescription Use (Part 21 CFR 801 Subpart D)    </span> <span> <span style="vertical-align: middle;">☐</span>     Over-The-Counter Use (21 CFR 801 Subpart C)    </span> </span> |
|-------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

 Department of Health and Human Services  

 Food and Drug Administration  

 Office of Chief Information Officer  

 Paperwork Reduction Act (PRA) Staff  

[PRAStaff@fda.hhs.gov](mailto:PRAStaff@fda.hhs.gov)

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

## Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.

510(k) Number (if known)

K232251

Device Name

APTUS® K-Wire System Line Extension

Indications for Use (Describe)

APTUS® K-Wire System is intended for use in fixation of bone reconstruction, and as guide pins for insertion of other implants.

| Type of Use (Select one or both, as applicable) | <span> <span style="padding-right: 20px;"> <span style="vertical-align: middle;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)    </span> <span> <span style="vertical-align: middle;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)    </span> </span> |
|-------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services  

Food and Drug Administration  

Office of Chief Information Officer  

Paperwork Reduction Act (PRA) Staff  

[PRAStaff@fda.hhs.gov](mailto:PRAStaff@fda.hhs.gov)

*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."*

{4}------------------------------------------------

## 510(k) Summary

## APTUS® Cannulated Compression Screws Line Extension APTUS® K-wire System Line Extension October 5, 2023

| ADMINISTRATIVE INFORMATION |                                                                                                                              |
|----------------------------|------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer Name          | Medartis AG<br>Hochbergerstrasse 60E<br>CH-4057 Basel, Switzerland<br>Telephone: +41 61 633 34 34                            |
| Official Contact           | Claudia De Santis<br>Head Regulatory Affairs<br>Email: claudia.desantis@medartis.com<br>Telephone: +41 79 883 52 88          |
| Additional correspondent   | Salvatore Risoli<br>Senior Regulatory Affairs Manager<br>Email: salvatore.risoli@medartis.com<br>Telephone: +41 61 633 37 72 |

#### DEVICE NAME AND CLASSIFICATION Trade/Device Name

| Trade/Device Name    | APTUS® Cannulated Compression Screws Line Extension<br>APTUS® K-Wire System Line Extension                                                     |
|----------------------|------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name          | Screw, Fixation, Bone (Primary)<br>Pin, Fixation, Smooth                                                                                       |
| Regulation Number    | 21 CFR 888.3040 (Primary)<br>21 CFR 888.3030                                                                                                   |
| Regulation Name      | Smooth Or Threaded Metallic Bone Fixation Fastener (Primary)<br>Single/multiple component metallic bone fixation<br>appliances and accessories |
| Regulatory Class     | Class II                                                                                                                                       |
| Product Codes        | HWC (Primary)<br>HTY                                                                                                                           |
| Classification Panel | Orthopedic                                                                                                                                     |

Applicable, recognized standards used in the non-clinical performance data of the subject device:

- Non-Clinical Performance Testing o
	- FDA Guidance "Orthopedic Non-Spinal Metallic Bone Screws and Washers l Performance Criteria for Safety and Performance Based Pathway"
	- ASTM F543-17, Standard Specification and Test Methods for Metallic Medical Bone Screws
	- ASTM F1839-08(2021) Standard Specification for Rigid Polyurethane Foam for Use as a Standard Material for Testing Orthopaedic Devices and Instruments
- O Materials
	- I ASTM F67-13 (2013/R2017) Standard Specification for Unalloyed Titanium for Surgical Implant Applications (UNS R50250, UNS R50400, UNS R50550, UNS R50700)

{5}------------------------------------------------

- ASTM F136-13 Standard Specification for Wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications (UNS R56401)
- . ASTM F138-19 Standard Specification for Wrought 18Chromium-14Nickel-2.5Molybdenum Stainless Steel Bar and Wire for Surgical Implants (UNS S31673)

#### PREDICATE DEVICE INFORMATION

Primary Predicate Device K202589, APTUS® headed Cannulated Compression Screws, APTUS® Cannulated Compression Screws, APTUS® K-Wire System, Medartis AG

Additional Predicate Devices K133460, APTUS Cannulated Compression Screws 5.0.7.0, Medartis AG K110658, APTUS® Cannulated Compression Screws, Medartis AG K092038, APTUS® K-Wire System, Medartis AG K230971, APTUS 3.5 TriLock Straight Plates, Medartis AG K963192, Synthes Sterile 3.5 mm and 4.0 mm Cannulated Screws, Synthese (USA)

#### INDICATIONS FOR USE STATEMENT

APTUS® Cannulated Compression Screws and headed Cannulated Compression Screws are indicated for the treatment of fractures, osteotomies and arthrodesis of bones with the appropriate screw size in the shoulder, elbow, wrist, hand, knee, and the foot and ankle.

APTUS® K-Wire System is intended for use in fixation of bone fractures, for bone reconstruction, and as guide pins for insertion of other implants.

#### SUBJECT DEVICE DESCRIPTION

The purpose of this submission is to obtain marketing clearance for APTUS® Cannulated Compression Screw, and APTUS® K-Wire System designs to expand the range of Medartis APTUS fixation devices previously cleared in K133460. K110658. K092038 and K202589.

This submission includes for APTUS® Cannulated Compression Screws (non-sterile):

- additional thread lengths for previously cleared thread diameters (2.2 mm, 3.0 mm, and 4.0 mm) - fully threaded screws for two previously cleared thread diameters (2.2 mm and 3.0 mm)

Additionally, the purpose of this submission is to obtain marketing clearance for various APTUS® Kwire designs (non-sterile and sterile) to expand the range of the Medartis APTUS® K-Wire System, previously cleared in K092038, K133460 and K202589, to include longer versions in two (2) diameters (0.8, 1.1 mm). The K-wires are compatible with the subject device cannulated compression screws and APTUS® Cannulated Compression Screws previously cleared in K110658, K133460 and K202589.

All subject device cannulated compression screws are manufactured from titanium alloy conforming to ASTM F136, Standard Specification for Wrought Titanium-6Aluminum- 4 Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications (UNS R56401), the same material for screws previously cleared in K110658, K133460 and K202589.

The subject device K-wires are manufactured from stainless steel conforming to ASTM F138, Standard Specification for Wrought 18Chromium-14Nickel-2.5Molybdenum Stainless Steel Bar and Wire for Surgical Implants (UNS S31673), the same material for K-wires previously cleared in K092038, K133460 and K202589.

#### PERFORMANCE DATA

Non-clinical testing data submitted, referenced, or relied upon to demonstrate substantial equivalence include: Mechanical testing according to ASTM F543-17 Standard Specification and Test Methods for Metallic Medical Bone Screws, ASTM F1839-08(2021) Standard Specification for Rigid Polyurethane Foam for Use as a Standard Material for Testing Orthopaedic Devices and Instruments and FDA Guidance "Orthopedic Non-Spinal Metallic Bone Screws and Washers – Performance Criteria for

{6}------------------------------------------------

Safety and Performance Based Pathway". Based on the results of the testing, the performance of the subject device was judged to be substantially equivalent to the predicate devices K202589, K110658 and K963192. Clinical data were not provided in this submission.

#### EQUIVALENCE TO MARKETED DEVICES

The subject device APTUS® Cannulated Compression Screws have similar designs to those cleared under K202589, K133460 and K110658. Additional lengths of APTUS® Cannulated Compression Screw thread diameters 2.2 mm, 3.0 mm and 4.0 mm as well as fully threaded versions for thread diameters 2.2 and 3.0 mm are subject devices of this subject device overall lengths and thread lengths are greater for the 2.2, 3.0 and 4.0 mm screws than those of devices previously cleared in K110658 and K202589. The subject device overall lengths are shorter and relative to those of devices previously cleared in K133460. However, the thread geometries are independent of length and, therefore, equivalent to the previously cleared APTUS® Cannulated Compression Screws in K110658, K133460 and K202589. The tip of the APTUS® Cannulated Compression Screws have a triangular SpeedTip® shape designed to improve cutting and reduce insertion torque. All cannulated compression screws have an internal hexalobular instrument interface. These features are equivalent to the devices previously cleared in K110658, K133460 and K202589.

The subject devices APTUS® Cannulated Compression Screws are compatible with the subject device K-wires and Medartis K-wires previously cleared in K202589, K092038 and K133460. Additionally, the subject device K-wires are compatible with the APTUS® Cannulated Compression Screws previously cleared in K202589, K133460 and K110658.

The subject device K-wires have similar designs and dimensions to those of devices cleared in K202589, K133460 and K092038. The predicate K-wires with a diameter of 0.8 mm and 1.1 mm have an overall length of 100 mm. They are provided with either a single-ended trocar (A-5040.xx), singleended bayonet (lancet) (A-5042.xx), or double-ended trocar (A-5043.xx) tips. The subject K-wires with a diameter of 0.8 mm and 1.1 mm have an overall length of 150 mm. They are provided with a single-ended trocar (A -5040.x1) tip. All subject device K-wires are manufactured from identical stainless-steel conforming to ASTM F138 as the K-wires cleared in K202589, K133460 and K092038.

The subject device includes components provided non-sterile in the same packaging and sterilized to a sterility assurance level (SAL) of 10-6 by the end user using the same sterilization method (moist heat) and parameters as devices previously cleared in K202589, K133460, K110658, and K092038. Additionally, the subject device includes components provided sterile to the end user in the same packaging that are terminally sterilized to an SAL of 10-6 using the same sterilization method (X-ray beam irradiation) as the primary predicate K202589. For both sterilizations, the subject devices do not represent a new worst-case.

All subject device cannulated compression screws, K-wires and accessories are manufactured in the same facilities using the same materials and manufacturing processes as used for the Medartis APTUS® devices previously cleared in predicates K202589, K110658, K133460, and K092038. Therefore, no new biocompatibility testing has been performed, as the subject device is substantially equivalent to these devices regarding materials and processing.

#### CONCLUSION

Overall, the subject device has the following similarities to the predicate devices:

- has the similar intended use,
- . uses the same operating principle,
- incorporates the same basic design,
- incorporates the same materials, and
- has similar packaging and is sterilized or is to be end-user sterilized using the same materials and processes.

The data included in this submission demonstrate substantial equivalence to the predicate devices listed above.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HWC/K232251](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HWC/K232251)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
