The Wright Jones Fracture System is indicated for fixation of bone fractures or for bone reconstruction of the 5th Metatarsal. Examples include: - Fixation of malunions and nonunions - Acute fractures - Avulsion fractures - Repetitive stress fractures - Jones Fractures - Malleolar Fractures - Talus Fractures - Greater Tuberosity Fractures
Device Story
Wright Jones Fracture System consists of partially threaded, solid titanium alloy bone screws; available in 4.5, 5.5, and 6.5 mm diameters; used by orthopedic surgeons for internal fixation of bone fractures or reconstruction of the 5th Metatarsal; provides mechanical stabilization of bone fragments to facilitate healing; device implanted during surgical procedures; performance characteristics include insertion, removal, pull-out, and torque strength.
Clinical Evidence
No clinical data provided; substantial equivalence supported by non-clinical bench testing.
Indicated for fixation of bone fractures or bone reconstruction of the 5th Metatarsal, including malunions, nonunions, acute, avulsion, repetitive stress, Jones, malleolar, talus, and greater tuberosity fractures.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
CHARLOTTE™ CAROLINA™ Jones Fracture System Screw (K140952)
K201319 — Medline UNITE Jones Fracture Screw System · Medline Industires, Inc. · Jun 17, 2020
K131620 — JONES-FX FRACTURE SYSTEM · Instratek, Inc. · Jul 25, 2013
K202657 — Javelin Tailor's Bunion Fixation System · Nvision Biomedical Technologies, Inc. · Dec 14, 2020
K201235 — Arthrex MaxForce MTP Compression Plates and Screws · Arthrex, Inc. · Jul 6, 2020
K961485 — FOREFOOT RECONSTRUCTION SYSTEM · Howmedica Corp. · Jun 1, 1996
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
March 13, 2019
Wright Medical Technology. Inc. Antonio Ayala Regulatory Affairs Specialist I 1023 Cherry Road Memphis. Tennessee 38117
Re: K183026
Trade/Device Name: Wright Jones Fracture System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: HWC Dated: February 8, 2019 Received: February 11, 2019
Dear Antonio Ayala:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrl/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you. however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Laurence D. Coyne -S
For Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K183026
Device Name Wright Jones Fracture System
#### Indications for Use (Describe)
The Wright Jones Fracture System is indicated for fixation of bone fractures or for bone reconstruction of the 5th Metatarsal. Examples include:
- Fixation of malunions and nonunions
- Acute fractures
- Avulsion fractures
- · Repetitive stress fractures
- · Jones Fractures
- Malleolar Fractures
- · Talus Fractures
- · Greater Tuberosity Fractures
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K183026 Page 1 of 2
1023 Cherry Road Memphis, TN 38117 USA
901 867 9971 800 238 7117 wright.com
Image /page/3/Picture/4 description: The image shows the Wright logo. The logo consists of two overlapping trapezoids, one red and one orange, to the left of the word "WRIGHT" in red. Below the word "WRIGHT" is the phrase "FOCUSED EXCELLENCE" in a smaller, gray font.
#### 510(K) SUMMARY
In accordance with the Food and Drug Administration Rule to implement provisions of the Safe Medical Devices Act of 1990 and in conformance with 21 CFR 807.92, this information serves as a Summary of Safety and Effectiveness for the use of the Wright Jones Fracture System.
| (a)(1). Submitted By: | Wright Medical Technology, Inc.<br>1023 Cherry Road<br>Memphis, TN 38117 |
|-------------------------------------|--------------------------------------------------------------------------------------------------------|
| Date: | October 31, 2018 |
| Contact Person: | Antonio Ayala<br>Regulatory Affairs Specialist I<br>Phone - (901) 290-5640<br>Fax – (901) 867-4190 |
| (a)(2). Proprietary Name: | Wright Jones Fracture System |
| Common Name: | Bone Screw |
| Classification Name and Regulation: | Smooth or threaded metallic bone fixation<br>fastener, 21 CFR 888.3040 – Class II |
| Device Product Code, Device Panel: | HWC - Orthopedic |
| (a)(3). Predicate Device: | K140952: CHARLOTTE™ CAROLINA™ Jones<br>Fracture System Screw<br>K053136: 5th Metatarsal Fracture Screw |
# (a)(4). Device Description
The Wright Jones Fracture System consists of partially threaded, solid screws offered in 4.5, 5.5, and 6.5 mm diameters and various lengths. The screws are manufactured from titanium alloy (ASTM F136).
{4}------------------------------------------------
# (a)(5). INTENDED USE
The Wright Jones Fracture System is indicated for fixation of bone fractures or for bone reconstruction of the 5th Metatarsal. Examples include:
- Fixation of malunions and nonunions
- Acute fractures
- · Avulsion fractures
- Repetitive stress fractures
- Jones Fractures
- Malleolar Fractures
- Talus Fractures
- · Greater Tuberosity Fractures
# (a)(6). Technological Characteristics Comparison
The Wright Jones Fracture System Screw and the legally marketed predicates have identical indications, similar design features, the same sterilization method, and similar performance characteristics. The modifications to the system include additional sizes, chamfered and headless screw options, and an updated driver interface.
# (b)(1). Substantial Equivalence - Non-Clinical Evidence
Performance testing and analysis demonstrated substantial equivalence to the predicate device in insertion torque, removal torque, pull out, ultimate torque, and yield torque strength.
# (b)(2). Substantial Equivalence - Clinical Evidence
N/A
# (b)(3). Substantial Equivalence - Conclusions
The design characteristics of the subject system do not raise any new types of questions of safety or effectiveness. From the evidence submitted in this 510(k), the subject devices can be expected to perform at least as well as the predicate device.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.