Arthrosurface Bone Screws

K172383 · Arthrosurface, Inc. · HWC · Dec 22, 2017 · Orthopedic

Device Facts

Record IDK172383
Device NameArthrosurface Bone Screws
ApplicantArthrosurface, Inc.
Product CodeHWC · Orthopedic
Decision DateDec 22, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3040
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Arthrosurface Bone Screws (2.0-3.0 mm solid and cannulated) are intended to be used as stand-alone bone screws for internal bone fixation for bone fractures, fusions, osteotomies, and non-unions in the ankle, foot, hand, and wrist. The Arthrosurface Bone Screws (3.5 mm and larger, solid and cannulated) are intended to be used as stand-alone bone screws for internal bone fixation for bone fractures, fusions, osteotomies, and non-unions in the ankle, foot, hand, wrist, shoulder, clavicle, scapula, olecranon, humerus, radius, ulna, tibia, calcaneous, femur, and fibula.

Device Story

Arthrosurface Bone Screws are metallic fixation fasteners; diameters 2.0-4.0 mm; lengths 8-50 mm. Features include solid or cannulated designs; partially or fully threaded; self-tapping or self-drilling. Used by orthopedic surgeons for internal bone fixation in fractures, fusions, osteotomies, and non-unions. Device provides mechanical stabilization of bone segments to facilitate healing. Operates via standard surgical implantation techniques.

Clinical Evidence

Bench testing only. Performance testing included torque to failure, insertion/removal torque, and axial pullout force per ASTM F543. Biocompatibility/sterility supported by Kinetic Chromogenic LAL testing per USP <85>, USP <161>, and AAMI ST72:2002/R2010.

Technological Characteristics

Materials: Implant-grade stainless steel or titanium alloy. Design: 2.0-4.0 mm diameter, 8-50 mm length; solid or cannulated; self-tapping or self-drilling. Mechanical testing per ASTM F543. Sterilization: Standard processes used for metallic implants.

Indications for Use

Indicated for internal bone fixation of fractures, fusions, osteotomies, and non-unions in various anatomical sites including ankle, foot, hand, wrist, shoulder, clavicle, scapula, olecranon, humerus, radius, ulna, tibia, calcaneous, femur, and fibula. Applicable to patients requiring bone fixation.

Regulatory Classification

Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. December 22, 2017 Arthrosurface, Inc. Dawn Wilson VP, Quality & Regulatory 28 Forge Parkway Franklin, Massachusetts 02038 Re: K172383 Trade/Device Name: Arthrosurface Bone Screws Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: HWC Dated: November 10, 2017 Received: November 14, 2017 Dear Ms. Wilson: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Mark N. Melkerson -S Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K172383 Device Name Arthrosurface Bone Screws Indications for Use (Describe) The Arthrosurface Bone Screws (2.0-3.0 mm solid and cannulated) are intended to be used as stand-alone bone screws for internal bone fixation for bone fractures, fusions, osteotomies, and non-unions in the ankle, foot, hand, and wrist. The Arthrosurface Bone Screws (3.5 mm and larger, solid and cannulated) are intended to be used as stand-alone bone screws for internal bone fractures, fusions, osteotomies, and non-unions in the ankle, foot, hand, wrist, shoulder, clavicle, scapula, olecranon, humerus, radius, ulna, tibia, calcaneous, femur, and fibula. Type of Use (Select one or both, as applicable) | <div> <span style="font-size:14px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | |----------------------------------------------------------------------------------------------------| | <div> <span style="font-size:14px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> | ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." C) {3}------------------------------------------------ #### 510(k) Summary Section 5 | 510(k) Owner: | Arthrosurface, Inc.<br>28 Forge Parkway<br>Franklin, MA 02038<br>Tel: 508.520.3003<br>Fax: 508.528.4604 | |----------------------------|---------------------------------------------------------------------------------------------------------| | Contact: | Dawn Wilson<br>VP, Quality & Regulatory | | Date of Preparation: | December 14, 2017 | | Trade Name: | Arthrosurface Bone Screws | | Common Name: | Screw, Fixation, Bone | | Device: | Smooth or Threaded Metallic Bone Fixation Fastener | | Classification Regulation: | Regulation Number 21 CFR 888.3040 | | Device Class: | Class II | | Review Panel: | Orthopedic | | Product Code: | HWC | ## Device Intended Use The Arthrosurface Bone Screws (2.0-3.0 mm solid and cannulated) are intended to be used as stand-alone bone screws for internal bone fixation for bone fractures, fusions, osteotomies, and non-unions in the ankle, foot, hand, and wrist. The Arthrosurface Bone Screws (3.5 mm and larger, solid and cannulated) are intended to be used as stand-alone bone screws for internal bone fixation for bone fractures, fusions, osteotomies, and non-unions in the ankle, foot, hand, wrist, shoulder, clavicle, scapula, olecranon, humerus, radius, ulna, tibia, calcaneous, femur, and fibula. ### Device Description The Arthrosurface Bone Screws have diameters ranging from 2.0 mm to 4.0 mm and lengths ranging from 8 mm to 50 mm. The screws are either partially or fully threaded, solid or cannulated, self-tapping or self-drilling and are manufactured from implant grade stainless steel or titanium alloy. Depending on their size, these bone screws are intended to be used for different indications for use. {4}------------------------------------------------ ## Substantial Equivalency: The intended use, materials, and application of the Proposed Device are substantially equivalent to the following previously cleared and commercially marketed predicate devices: | Primary Predicate(s) | | |------------------------------|---------| | • Arthrex Low Profile Screws | K143614 | | Additional Predicate(s) | | | • Arthrex Fracture System | K112437 | | • Arthrex Low Profile Screws | K103705 | | • Synthes Cortical Screws | K112583 | | • Arthrosurface Bone Screws | K161539 | The fundamental scientific technology of the proposed device has not changed relative to the predicate devices: - Has similar indications for use ● - Has similar device designs - Uses the same operating principle - Is manufactured using the same Stainless Steel and Titanium implant materials - o Is packaged and sterilized using the same materials and processes The following non-clinical testing was performed on the Arthrosurface Bone Screws: - Torque to failure, insertion/removal torque and axial pullout force tests per ASTM F543 - . A Kinetic Chromogenic LAL Test for Devices that meets the standard limit of 0.5 EU/mL or 20 EU/ Device per United States Pharmacopeia (USP) Chapter <85> Bacterial Endotoxins Test, USP Chapter <161> Transfusion and Infusion Assemblies and Similar Medical Devices, and AAMI ST72:2002/R2010, Bacterial Endotoxins-Test Methodologies, Routine Monitoring, and Alternatives to Batch Testing. The safety and effectiveness of using the Arthrosurface Bone Screws is adequately supported by the substantial equivalence and performance testing information provided within this Premarket Notification. The results have demonstrated that the Arthrosurface Bone Screws are substantially equivalent to the predicate devices.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...