IRENE Cannulated Screw System

K170056 · Tianjin Zhengtian Medical Instrument Co., Ltd. · HWC · Oct 23, 2017 · Orthopedic

Device Facts

Record IDK170056
Device NameIRENE Cannulated Screw System
ApplicantTianjin Zhengtian Medical Instrument Co., Ltd.
Product CodeHWC · Orthopedic
Decision DateOct 23, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3040
Device ClassClass 2
AttributesTherapeutic

Indications for Use

IRENE Cannulated Screw System is intended for fracture fixation, fusion and osteotomies of bones appropriate for the size of the device. For specific screw indications please see below. Acutec Headless Compression Screw: Used for fusion, fractures, or osteotomies of the clavicle, humerus, radius, ulna, ilium, femur, patella, fibula, tibia, talus, malleolus, calcaneus and other small bones. Cannulated Screw: Intended for bone fracture fixation and bone fragment fixation. Washers may be used with the Cannulated Screw in cases where the patient has poor bone quality.

Device Story

IRENE Cannulated Screw System consists of Acutec Headless Compression Screws and Cannulated Screws with optional washers. Device used for bone fracture fixation, fusion, and osteotomies. Screws are cannulated to allow placement over guide wires. Operated by orthopedic surgeons in clinical settings. Output is mechanical fixation of bone fragments. Benefits include stabilization of fractures and support for bone healing. Device is non-sterile and requires steam sterilization before use.

Clinical Evidence

No clinical data is presented in this submission. Bench testing only.

Technological Characteristics

Materials: Titanium Alloy (ASTM F 136, ISO 5832-3). Design: Cannulated, headless or headed bone screws. Energy source: N/A (mechanical). Connectivity: N/A. Sterilization: Steam sterilization. Software: None.

Indications for Use

Indicated for fracture fixation, fusion, and osteotomies of bones appropriate for device size, including clavicle, humerus, radius, ulna, ilium, femur, patella, fibula, tibia, talus, malleolus, calcaneus, and other small bones. Cannulated screws indicated for bone fracture and fragment fixation; washers indicated for patients with poor bone quality.

Regulatory Classification

Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with a symbolic representation of a parent cradling a child. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 Tianjin ZhengTian Medical Instrument Co., Ltd Wang Qi Regulatory Affairs Specialist No.318, Jingyi Road, Airport Economic Zone Tianjin, CN 100082 P.R. China October 23, 2017 Re: K170056 Trade/Device Name: IRENE Cannulated Screw System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: HWC Dated: September 26, 2017 Received: 09/26/2017 Dear Wang Qi: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in {1}------------------------------------------------ the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely, Mark N. Melkerson -S Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below. 510(k) Number (if known) K170056 Device Name IRENE Cannulated Screw System Indications for Use (Describe) IRENE Cannulated Screw System is intended for fracture fixation, fusion and osteotomies of bones appropriate for the size of the device. For specific screw indications please see below. Acutec Headless Compression Screw: Used for fusion, fractures, or osteotomies of the clavicle, humerus, radius, ulna, ilium, femur, patella, fibula, tibia, talus, malleolus, calcaneus and other small bones. Cannulated Screw: Intended for bone fracture fixation and bone fragment fixation. Washers may be used with the Cannulated Screw in cases where the patient has poor bone quality. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995, #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." FORM FDA 3881 (8/14) PSC Publishing Services (301) 443-6740 EF {3}------------------------------------------------ # 510(k) Summary In accordance with 21 CFR\$807.92 and the Safe Medical Devices Act of 1990, the following information is provided for IRENE Cannulated Screw System 510(k) premarket notification. The submission was prepared in accordance with the FDA guidance document, "Format for Traditional and Abbreviated 510(k)", issued on August 12, 2005. # 1. 510(k) Submitter and Owner Tianjin ZhengTian Medical Instrument Co., Ltd Address: No.318, Jingyi Road, Airport Economic Zone, Tianjin, P.R. China Official Correspondent: Xiang, Zhaojun Position: Regulatory Affairs Manager Tel: +86 10 82292929 Fax: +86 10 82290740 E-mail: xiangzhaojun@naton.cn Date of preparation of the Summary: Oct 17, 2017 # 2. Submission Contact Wang, Qi Position: Regulatory Affairs Specialist Tel: +86 10 82292929 E-mail: wangqi@naton.cn # 3. Proposed Device Trade Name: IRENE Cannulated Screw System Common Name: Screw, Fixation, Bone Class: Class 2 Regulation: 21 CFR 888.3040 Product Code: HWC (Bone Fixation Screw) Review Panel: Orthopedic # Device description: IRENE Cannulated Screw System includes two kinds of screws, Acutec Headless Compression Screw and Cannulated Screw. Acutec Headless Compression Screw has six different diameters in various lengths and is {4}------------------------------------------------ manufactured from Titanium Alloy (ASTM F 136 and ISO 5832-3). Cannulated Screw has four different diameters in various lengths and contains two optional washers. Cannulated Screw and washers are manufactured from Titanium Alloy (ISO 5832-3). Indication for use: IRENE Cannulated Screw System is intended for fracture fixation, fusion and osteotomies of bones appropriate for the size of the device. For specific screw indications please see below. Acutec Headless Compression Screw: Used for fusion, fractures, or osteotomies of the clavicle, humerus, radius, ulna, ilium, femur, patella, fibula, tibia, talus, malleolus, calcaneus and other small bones. Cannulated Screw: Intended for bone fracture fixation and bone fragment fixation. Washers may be used with the Cannulated Screw in cases where the patient has poor bone quality. | | Predicate Device 1 | Predicate Device 2 | Reference Device | |--------------------------|-------------------------------------------------|------------------------------------|----------------------------------------------| | 510(k)<br>Number | k063298 | k100359 | k151508 | | Predicate<br>Device Name | OsteoMed Headless<br>Cannulated Screw<br>System | DARCO® Headed<br>Cannulated Screws | IRENE Locking<br>Compression Plate<br>System | | Manufacture | OsteoMed | Wright | Zhengtian | | Classification | 888.3040(HWC) | 888.3040(HWC) | 888.3040(HWC)<br>888.3030(HRS) | #### 4. Predicate Device and Reference Device #### 5. Preclinical Data Mechanical testing is conducted to demonstrate that differences in technological characteristics of the proposed device to predicate device do not raise new questions regarding safety and effectiveness, and Substantially Equivalent to the predicate device is thus concluded. The testing including torsional properties, driving torque test and axial pullout test incorporates the following standard as test method to evaluate the performance of both proposed and predicate device: ASTM F543-13, Standard {5}------------------------------------------------ Specification and Test Methods for Metallic Medical Bone Screws. # 6. Clinical Data No clinical data is presented in this submission. # 7. Substantial Equivalence Comparison Table 1: comparison between Acutec Headless Compression Screw and predicate devices 1 | Element | Proposed Device<br>Acutec Headless Compression<br>Screw | Predicate Device 1<br>OsteoMed Headless Cannulated<br>Screw System | |-----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Class | 2 | 2 | | Regulation<br>number | 21 CFR 888.3040 | 21 CFR 888.3040 | | Indication<br>for use | Used for fusion, fractures, or<br>osteotomies of the clavicle,<br>humerus, radius, ulna, ilium,<br>femur, patella, fibula, tibia, talus,<br>malleolus, calcaneus and other<br>small bones. | The OsteoMed Headless Screw<br>System is indicated for use in<br>bone reconstruction, osteotomy,<br>arthrodesis, joint fusion, fracture<br>repair, and fracture fixation of<br>bone appropriate for the size of<br>the device. Screws are intended<br>for single use only. | | Material | Ti6Al4V | Ti6Al4V | | Design | Fully threaded, headless,<br>cannulated | Headless, cannulated | | Diameter | 2.5~7.5mm | 2.0~6.5mm | | Length | 8~120mm | 10~120mm | | Supplied | Non-sterile | Non-sterile | | Single-use<br>only | Yes | Yes | | Sterilization | Steam sterilization | Steam sterilization | Table 2: comparison between Cannulated Screw and predicate devices 2 | Element | Proposed Device | Predicate Device 2 | |-------------------|------------------|----------------------| | | Cannulated Screw | Wright DARCO® Headed | | | | Cannulated Screws | | Class | 2 | 2 | | Regulation number | 21 CFR 888.3040 | 21 CFR 888.3040 | {6}------------------------------------------------ | Indication for use | | | | |--------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------| | | Intended for bone fracture<br>fixation and bone fragment<br>fixation. Washers may be<br>used with the Cannulated<br>Screw in cases where the<br>patient has poor bone<br>quality. | Intended for use over a guide pin<br>or wire for bone fracture fixation<br>and bone fragment fixation.<br>Washers may be used with the<br>screws in cases where the<br>patients has poor bone quality. | | | | Material | Ti6Al4V | Ti6Al4V | | | Design | self-tapping,<br>cannulated | self-tapping/self-drilling,<br>cannulated | | | Diameter | 3.5~7.3mm | 2.7~7.5mm | | | Length | 20~110mm | 8~120mm | | | Supplied | Non-sterile | Non-sterile | | | Single-use only | Yes | Yes | | Sterilization | Steam sterilization | Steam sterilization | | #### 8. Substantial Equivalence Conclusion The proposed device is found to possess similar technological characteristics on design under the premises of sharing same intended use. Device performance is compared through the testing of mechanical properties of the proposed and predicate device by methods stipulated in published standards in order to evaluate whether differences existing in the design raise safety issue by compromising performance. Results of testing indicates no difference between the level of performance of proposed device and the predicate device, through which the differences are deemed not relevant to device performance and has been justified, and the conclusion of Substantial Equivalence is reached.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...