NON-STERILE KIRSCHNER WIRES AND STEINMANN PINS, MODELS #'S 290-101/160, 290-201/216, 290-060/330, 290-161/220

K983121 · Syntec Scientific Corp. · HTY · Oct 5, 1998 · Orthopedic

Device Facts

Record IDK983121
Device NameNON-STERILE KIRSCHNER WIRES AND STEINMANN PINS, MODELS #'S 290-101/160, 290-201/216, 290-060/330, 290-161/220
ApplicantSyntec Scientific Corp.
Product CodeHTY · Orthopedic
Decision DateOct 5, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3040
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Kirschner Wires and Steinmann Pins are provided non-sterile. The devices are indicated for use in fracture fixation, for healing of facile bone fragments, for osteotomies in the presence of adequate immobilization, as guide pins for insertion of other implants.

Device Story

Non-sterile Kirschner wires and Steinmann pins; used for bone fixation, osteotomies, and as guide pins for other implants. Manufactured from 316 LS stainless steel. Available in various lengths (70-400 mm) and diameters (0.7-4.8 mm) with diamond or trocar tip styles. Intended for use by surgeons in clinical/OR settings. Provided non-sterile; requires steam sterilization by the user prior to implantation. Mechanical device; no electronic or software components.

Clinical Evidence

No clinical data; bench testing only.

Technological Characteristics

Material: 316 LS stainless steel. Form factor: Shank and threaded wires/pins. Dimensions: 0.7-4.8 mm diameter; 70-400 mm length. Tip styles: Diamond, trocar. Sterilization: Steam (vacuum cycle, 132-135°C for 6 minutes).

Indications for Use

Indicated for fracture fixation, healing of bone fragments, osteotomies requiring immobilization, and as guide pins for other implants in patients requiring orthopedic surgical intervention.

Regulatory Classification

Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ #### OCT 5 1998 # 510(k) Summary of Safety and effectiveness - A Sponsor: Syntec-Taichung Medical Instruments Co., Ltd. 2, Kung San Road, Chuan Shing Industrial Zone, Shen Kang, Chang Hua, Taiwan. 509 Phone / FAX: 886-4-7987099 / 886-4-7987077 Contact Person: Ted Y. Shi - Proprietary Name : Non-sterile Kirschner Wires and Steinmann Pins A - A Common Name : Bone fixation fasteners - ア Classification Name : Class II, CFR 888.3040 - A Device Product Code : 87 HTY and 87 JDW - A Material: This device is manufactured from commercially 316 LS stainless steel. - A Indication for Use : The Kirschner Wires and Steinmann Pins are provided non-sterile. The devices are indicated for use in fracture fixation, for healing of facile bone fragments, for osteotomies in the presence of adequate immobilization, as guide pins for insertion of other implants. #### A Description of the Device : This system makes up of non-sterile, shank and threaded Kirschner Wires and Steinmann Pins. The Kirschner Wires ranges in length from 101.6 to 300 mm and ranges in diameter from 0.7 to 1.6 mm. The Steinmann Pins ranges in length from 70 to 400 mm and ranges in diameter from 0.8 to 4.8 mm. They are available with four-point style: diamond, trocar, both ends diamond and both ends trocar. The 9 "(230 mm) shank and threaded Kirschner Wires and Steinmann Pins are usually available with fracture fixation fasteners. {1}------------------------------------------------ ## Appendixes A ### Sterilization Information-Non-sterile Device These devices are provided non-sterile. The recommended sterilization parameters for nonsterile devices are as follows: | Method | Cycle | Time | Temperature | |--------|--------|-------|-------------| | Steam | Vacuum | 6 min | 132~135 ℃ | Image /page/1/Picture/4 description: The image shows a black and white abstract pattern. The pattern is composed of small, irregular shapes and lines. The shapes are concentrated in the upper right corner of the image, and they become more sparse towards the bottom left corner. The overall effect is one of randomness and chaos. 12 {2}------------------------------------------------ Image /page/2/Picture/2 description: The image shows the seal of the U.S. Department of Health & Human Services (HHS). The seal features a stylized eagle with three legs, representing health, hope, and service. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle. 5 1998 OCT Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Mr. Ted Y. Shi Vice-Chairman Syntec-Taichung Medical Instruments Company Limited 2, Kung San Road Chuan Shing Industrial Zone, Shen Kanq Chang Hua, Taiwan 509 Re: K983121 Non-sterile Kirschner Wires and Steinmann Pins Regulatory Class: II Product Codes: HTY and JDI Dated: September 1, 1998 Received: September 8, 1998 Dear Mr. Shi: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration ...... If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. ਉ substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical General regulation (21 CFR Part 820) and that, Devices: through periodic (QS) inspections, the Food and Drug Administration (FDA) will werify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations. {3}------------------------------------------------ Page 2 - Mr. Ted Y. Shi This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other gener Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Page _ _ of __________________________________________________________________________________________________________________________________________________________________ 510(K) Number (if known): __K983121 Device Name: Syntec-Taichung Non-sterile Kirschner Wires and Steinmann Pins Indications for use: The Kirschner Wires and Steinmann Pins are provided non-sterile. The devices are indicated for use in fracture fixation, for healing of facile bone fragments, for osteotomies in the presence of adequate immobilization, as guide pins for insertion of other implants. (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use _ (Per 21 CFR 801.109) OR Over-The-Counter-Use__________________________________________________________________________________________________________________________________________________________ (Optional Format 1-2-96) (Division Sign-Off) Division of General Restorative Devices 510(k) Number K9831
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...