The Impact PEEK Union Nail System is indicated for maintenance of alignment and fixation of bone fractures, osteotomies, arthrodeses or bone grafts in the presence of appropriate additional immobilization (e.g. rigid fixation implants, cast, brace).
Device Story
Impact PEEK Union Nail System is a bone fixation device; constructed from HA-enhanced PEEK; features cannulated design for k-wire guidance; incorporates tantalum pins for radiographic visibility. Used by surgeons in clinical settings for fracture management, osteotomies, arthrodeses, or bone grafts. Device provides mechanical alignment and fixation; requires supplemental immobilization (cast, brace, or rigid implants). Tantalum markers assist surgeons in verifying placement via imaging. Benefits include radiolucency of PEEK material combined with visibility of tantalum markers, facilitating accurate surgical placement and post-operative monitoring.
Clinical Evidence
No clinical data provided. Substantial equivalence demonstrated via engineering analysis and non-clinical bench testing based on recognized consensus standards.
Technological Characteristics
Material: HA-enhanced PEEK; tantalum pins for imaging. Design: Cannulated for k-wire guidance; various lengths and diameters. Energy source: None (mechanical). Connectivity: None. Sterilization: Not specified.
Indications for Use
Indicated for maintenance of alignment and fixation of bone fractures, osteotomies, arthrodeses, or bone grafts requiring additional immobilization (e.g., rigid fixation, cast, brace).
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
K250646 — Impact PEEK Union Nail System · Nvision Biomedical Technologies · Jun 11, 2025
K132774 — PICCOLO COMPOSITE NAILING SYSTEM · CarboFix Orthopedics , Ltd. · Oct 3, 2013
K141905 — PRO-LINK WEDGE SYSTEM · Life Spine · Sep 26, 2014
K160002 — Piccolo Composite(R) Distal Volar Radius Plate System · CarboFix Orthopedics , Ltd. · Mar 9, 2016
K102597 — PICCOLO COMPOSITE PLATE SYSTEM · N.M.B. Medical Applications, Ltd. · Feb 24, 2011
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
September 19, 2025
Nvision Biomedical Technologies INC.
Tom Zink
Senior Vice President of Product Development
4590 Lockhill Selma Rd
San Antonio, Texas 78249
Re: K252657
Trade/Device Name: Impact PEEK Union Nail System
Regulation Number: 21 CFR 888.3040
Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener
Regulatory Class: Class II
Product Code: HTY
Dated: August 13, 2025
Received: August 22, 2025
Dear Tom Zink:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252657 - Tom Zink
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE
{2}
K252657 - Tom Zink
Page 3
by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Shumaya Ali -S
Shumaya Ali, M.P.H.
Assistant Director
DHT6C: Division of Restorative,
Repair, and Trauma Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K252657
Device Name
Impact PEEK Union Nail System
Indications for Use (Describe)
The Impact PEEK Union Nail System is indicated for maintenance of alignment and fixation of bone fractures, osteotomies, arthrodeses or bone grafts in the presence of appropriate additional immobilization (e.g. rigid fixation implants, cast, brace).
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
**CONTINUE ON A SEPARATE PAGE IF NEEDED.**
This section applies only to requirements of the Paperwork Reduction Act of 1995.
**DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.**
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}
K252657
# 510(k) Summary
## DATE PREPARED
August 15, 2025
## MANUFACTURER AND 510(k) OWNER
Nvision Biomedical Technologies, INC
4590 Lockhill Selma
San Antonio, TX 78249, USA
Telephone: (210) 545-3713
Fax: (866) 764-1139
Official Contact: Tom Zink, Senior Vice President of Product Development
Telephone: (210) 870-8311
Email: tomzink@nvisionbiomed.com
## PROPRIETARY NAME OF SUBJECT DEVICE
Impact PEEK Union Nail System
## COMMON NAME
PEEK, Pin, Fixation
## DEVICE CLASSIFICATION
Smooth or threaded metallic bone fixation
(Classification Regulations: 21 CFR 888.3040, Product Codes: HTY, Class: II)
## PREMARKET REVIEW
OHT6/DHT6 - Division of Research, Research, and Training Devices
Orthopedic Panel
## INDICATIONS FOR USE
The Impact PEEK Union Nail System is indicated for maintenance of alignment and fixation of bone fractures, osteotomies, arthrodeses or bone grafts in the presence of appropriate additional immobilization (e.g. rigid fixation implants, cast, brace).
## DEVICE DESCRIPTION
The Impact PEEK Union Nail System represents a cutting-edge fixation device crafted from advanced HA Enhanced PEEK. The system is versatile, offering a range of lengths and diameters to suit various clinical needs, and features a cannulated design for precise k-wire guidance. Additionally, it integrates tantalum pins to ensure optimal imaging visibility, enhancing surgical accuracy and outcomes.
Page 1 of 2
{5}
K252657
# PREDICATE DEVICE IDENTIFICATION
The Impact PEEK Union Nail System is substantially equivalent to the following predicates:
| 510(k) Number | Predicate Device Name / Manufacturer | Primary Predicate |
| --- | --- | --- |
| K250646 | Nvision Impact PEEK Union Nail | ✓ |
| K143618 | Zimmer Kirschner wires and Steinmann pins | |
| K160304 | Phalanx Innovations OsteoBullet Compression Screw | |
# SUMMARY OF NON-CLINICAL TESTING
No FDA performance standards have been established for the Impact PEEK Union Nail System. The following tests were performed to demonstrate safety based on recognized consensus standards and current industry practice:
The results of these tests, as well as engineering analysis of device characteristics, indicate that the Impact PEEK Union Nail System is substantially equivalent to the predicate devices.
# EQUIVALENCE TO PREDICATE DEVICES
Nvision believes that the Impact PEEK Union Nail System is substantially equivalent to the predicate devices based on the information summarized here:
The subject device has a similar design, similar dimensions, and uses similar materials as the devices cleared in K250646, K143618 and K160304. The subject device also has the same intended use, as well as similar technological characteristics as these predicates. The Indications for Use are equivalent and any minor differences in wording choices are insignificant. These technological characteristics have undergone testing and engineering analysis to ensure the device is as safe and effective as the predicates.
# CONCLUSION
Based on the engineering analysis of device characteristics, it can be concluded that the subject device The Impact PEEK Union Nail System does not raise new issues of safety or efficacy compared to the predicate devices. The similar indications for use, technological characteristics, and performance characteristics for the proposed are assessed to be substantially equivalent to the predicate devices.
Page 2 of 2
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.