Impact PEEK Union Nail System

K250646 · Nvision Biomedical Technologies · HTY · Jun 11, 2025 · Orthopedic

Device Facts

Record IDK250646
Device NameImpact PEEK Union Nail System
ApplicantNvision Biomedical Technologies
Product CodeHTY · Orthopedic
Decision DateJun 11, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3040
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Impact PEEK Union Nail System is indicated for maintenance of alignment and fixation of bone fractures, osteotomies, arthrodeses or bone grafts in the presence of appropriate additional immobilization (e.g. rigid fixation implants, cast, brace).

Device Story

Impact PEEK Union Nail System consists of cannulated pins (3.0mm–4.5mm diameter, 50mm length) used for bone fixation. Implants feature ridges for stability and tantalum pins for radiographic visibility. Used by orthopedic surgeons in clinical settings to maintain alignment of fractures, osteotomies, or arthrodeses. Device is inserted over a k-wire. Provides mechanical stabilization of bone segments; requires supplemental immobilization. Benefits include secure fixation and improved imaging accuracy during placement.

Clinical Evidence

Bench testing only. Performed static and dynamic bending, axial pullout (ASTM F1264-16, ASTM F543-17), and shear testing (ASTM D2344) to demonstrate substantial equivalence.

Technological Characteristics

Materials: HA Enhanced PEEK (ASTM F2026) and tantalum markers. Design: Cannulated pins with ridges. Dimensions: 3.0mm–4.5mm diameter, 50mm length. Principle: Mechanical bone fixation fastener. Sterilization: Not specified.

Indications for Use

Indicated for patients requiring maintenance of alignment and fixation of bone fractures, osteotomies, arthrodeses, or bone grafts, provided additional immobilization (e.g., rigid fixation, cast, brace) is used.

Regulatory Classification

Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION June 11, 2025 Nvision Biomedical Technologies Erin Lansdale Senior Project Engineer 4590 Lockhill Selma Rd San Antonio, Texas 78249 Re: K250646 Trade/Device Name: Impact PEEK Union Nail System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: HTY Dated: May 12, 2025 Received: May 12, 2025 Dear Erin Lansdale: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K250646 - Erin Lansdale Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act, or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K250646 - Erin Lansdale Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, CHRISTOPHER FERREIRA -S Christopher Ferreira, MS Assistant Director DHT6C: Division of Restorative, Repair, and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. Submission Number (if known) K250646 Device Name Impact PEEK Union Nail System Indications for Use (Describe) The Impact PEEK Union Nail System is indicated for maintenance of alignment and fixation of bone fractures, osteotomies, arthrodeses or bone grafts in the presence of appropriate additional immobilization (e.g. rigid fixation implants, cast, brace). Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) **CONTINUE ON A SEPARATE PAGE IF NEEDED.** This section applies only to requirements of the Paperwork Reduction Act of 1995. **DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.** The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4} K250646 Page 1/2 510(k) Summary DATE PREPARED January 30, 2025 MANUFACTURER AND 510(k) OWNER Nvision Biomedical Technologies, INC 4590 Lockhill Selma San Antonio, TX 78249, USA Telephone: (210) 545-3713 Fax: (866) 764-1139 Official Contact: Erin Lansdale, Senior Project Engineer Telephone: (210) 598-5657 Email: erinlansdale@nvisionbiomed.com PROPRIETARY NAME OF SUBJECT DEVICE Impact PEEK Union Nail System COMMON NAME PEEK, Pin, Fixation DEVICE CLASSIFICATION Smooth or threaded metallic bone fixation (Classification Regulations: 21 CFR 888.3040, Product Codes: HTY, Class: II) PREMARKET REVIEW OHT6/DHT6 - Division of Research, Research, and Training Devices Orthopedic Panel INDICATIONS FOR USE The Impact PEEK Union Nail System is indicated for maintenance of alignment and fixation of bone fractures, osteotomies, arthrodeses or bone grafts in the presence of appropriate additional immobilization (e.g. rigid fixation implants, cast, brace). DEVICE DESCRIPTION The Impact PEEK Union Nail System contains 3.0mm - 4.5mm diameter, 50mm long pins manufactured from HA Enhanced PEEK (ASTM F2026). The implants are designed with strategically placed ridges to improve initial stability and cannulated structure to simplify insertion over a k-wire. Additionally, it was designed to include tantalum pins to ensure optimal imaging visibility for device placement accuracy. {5} K250646 Page 2/2 # PREDICATE DEVICE IDENTIFICATION The Impact PEEK Union Nail System is substantially equivalent to the following predicates: | 510(k) Number | Predicate Device Name / Manufacturer | Primary Predicate | | --- | --- | --- | | K133932 | Inion FreedomPin | ✓ | | K181180 | OSSIO Pin Product Family | | # SUMMARY OF NON-CLINICAL TESTING No FDA performance standards have been established for the Impact PEEK Union Nail System. The following tests were performed to demonstrate substantial equivalence based on recognized consensus standards and current industry practice: - Static and Dynamic Bending, Axial Pullout (ASTM F1264-16, ASTM F543-17) - Shear Testing (ASTM D2344) The results of these tests, as well as engineering analysis of device characteristics, indicate that the Impact PEEK Union Nail System is substantially equivalent to the predicate devices. # EQUIVALENCE TO PREDICATE DEVICES Nvision believes that the Impact PEEK Union Nail System is substantially equivalent to the predicate devices based on the information summarized here: The subject device has a similar design and similar dimensions as the devices cleared in K133932 and K181180 but differs because it is manufactured from HA Enhanced PEEK. The subject device also has the same intended use, as well as similar technological characteristics as these predicates. The Indications for Use are equivalent and any minor differences in wording choices are insignificant. These technological characteristics have undergone testing and engineering analysis to ensure the device is as safe and effective as the predicates. # CONCLUSION Based on the testing performed, including bending, axial pullout, and shear testing, as well as engineering analysis of device characteristics, it can be concluded that the subject device The Impact PEEK Union Nail System does not raise new issues of safety or efficacy compared to the predicate devices. The similar indications for use, technological characteristics, and performance characteristics for the proposed are assessed to be substantially equivalent to the predicate devices.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...