C-Wires are single use devices. They are intended to be used by trained professionals for the purpose of bone fracture fixation. The C-Wires should only be used with wire drivers that are designed to drive wires having longitudinal flats.
Device Story
C-Wire Double Ended Orthopedic Wires are single-use metallic rods used for bone fracture fixation. Operated by trained professionals in operating rooms or ambulatory surgical centers; wires are inserted into bone using powered handpieces (wire drivers) designed to engage longitudinal flats on the wire. Device provides mechanical stability during natural bone healing. Output is physical fixation of bone fragments. Benefits include stabilization of small bone trauma. Device is a passive mechanical implant; no software or algorithm involved.
Clinical Evidence
No clinical data. Safety and effectiveness established via bench testing (transportation, biocompatibility, sterilization, shelf-life), 38-year history of use with low complaint rates, and comparison to predicate device.
Technological Characteristics
Material: 316LVM stainless steel. Form factor: 5" (127mm) length; diameters 0.7mm, 0.89mm, 1.14mm, 1.57mm; spade or trocar tips. Energy source: Manual/powered handpiece insertion. Sterilization: Gamma radiation (SAL 10^-6). Connectivity: None.
Indications for Use
Indicated for bone fracture fixation in patients requiring surgical repair of small bone trauma. Intended for use by trained professionals.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
K150889 — Suture Wires and Wire Loops · Zimmer, Inc. · Jul 1, 2015
K092038 — APTUS K-WIRE SYSTEM · Medartis AG · Sep 23, 2009
K073674 — BONE FIXATION WIRE (KIRSCHNER WIRE) · Signal Medical Corp. · Mar 24, 2008
K131459 — AAP K-WIRE, STEINMANN PIN, CERCLAGE WIRE · Aap Implantate AG · Oct 17, 2013
K161489 — Syntorr K-wire and Pin System · Syntorr, Inc. · Sep 16, 2016
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other. The profiles are simple and abstract, and they are connected by a flowing line.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 12, 2014
CONMED Corporation Ms. Joy Lovett Regulatory Affairs Specialist 11311 Concept Blvd Largo, Florida 33773
Re: K142134
Trade/Device Name: C-Wire Double Ended Orthopedic Wires Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: HTY Dated: July 31, 2014 Received: August 6, 2014
Dear Ms. Lovett:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
Page 2 - Ms. Joy Lovett
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Lori A. Wiggins -S
for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known): K142134
Device Name: C-Wire Double Ended Orthopedic Wires
Intended Use / Indications for Use:
C-Wires are single use devices. They are intended to be used by trained professionals for the purpose of bone fracture fixation. The C-Wires should only be used with wire drivers that are designed to drive wires having longitudinal flats.
#### Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
# Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for CONMED Corporation. The logo consists of a stylized letter "C" on the left, followed by the word "CONMED" in bold, sans-serif font. Below the word "CONMED" is the word "CORPORATION" in a smaller, sans-serif font. The logo is simple and modern, and the use of bold font gives it a strong and professional look.
# 510(k) Summarv
In accordance with the requirements of the Safe Medical Device Act of 1990 and 21 CFR 807.92, CONMED Corporation is hereby submitting the 510(k) Summary of Safety and Effectiveness for 510(k) Number K142134
Date Prepared: 5 September 2014
#### A. Submitter
ConMed Corporation, Largo 11311 Concept Boulevard Largo, FL Establishment Registration Number:
- B. Company Contact
Joy Lovett Requlatory Affairs Specialist Phone: 727-399-5137 Fax: 727-399-5264
- C. Device Name
| Trade Name: | C-Wire Double Ended Orthopedic Wires |
|------------------------|--------------------------------------|
| Common Name: | Fixation Wire |
| Classification Name: | Pin, Fixation, Smooth |
| Proposed Class:Device: | Class II |
| Product Codes: | HTY |
| Regulation: | 21 CFR Part 888.3040 |
- D. Predicate/Legally Marketed Devices
| Device Name: | SMT Schilling Kirschner/Guide Wires, K100736 |
|---------------|----------------------------------------------|
| Company Name: | SMT Schilling Metalltechnik GmbH |
| 510(k) #: | K100736 |
#### E. Device Description
The C-Wires are single use devices. They are intended to be used by trained professionals for the purpose of bone fracture fixation. The C-Wires should only be used with wire drivers that are designed to drive wires having longitudinal flats.
The intended use / indications statement for the C-Wires materially differs from that of the Schilling Kirschner/Guide Wires in that the C-Wire statement does not include a guidance function for insertion of implants into the skeletal system. CONMED manufactures a separate line of quidewires for this function. The
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Conmed Corporation. The logo features a stylized "C" shape on the left, followed by the word "CONMED" in a bold, sans-serif font. Below the word "CONMED" is the word "CORPORATION" in a smaller, sans-serif font.
separation of these functions does not have an impact on the safety and effectiveness of the device.
- Intended Use / Indications F.
The C-Wires are single use devices. They are intended to be used by trained professionals for the purpose of bone fracture fixation. The C-Wires should only be used with wire drivers that are designed to drive wires having longitudinal flats.
The intended use / indications statement for the C-Wires materially differs from that of the Schilling Kirschner/Guide Wires in that the C-Wire statement does not include a quidance function for insertion of implants into the skeletal system. CONMED manufactures a separate line of guide wires for this function. The separation of these functions does not have an impact on the safety and effectiveness of the device.
- G. Summary of Technological Characteristics
The following table represents a summary of the technological characteristics of the ConMed C-Wire Double Ended Orthopedic Wires and the SMT Schilling Kirschner/GuideWires
| | Proposed Device<br>CONMED C-Wire Double<br>Ended Orthopedic Wires | Predicate Device<br>SMT Schilling<br>Kirschner/Guide Wires<br>(K100736) |
|-------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Brief Description | C-Wire Double Ended<br>Orthopedic Wires are single-<br>use devices that are indicated<br>for surgical repair of small<br>bone trauma. The C-Wires<br>provide mechanical stability<br>during the natural bone<br>healing process. Each wire is<br>5" (127mm) long, with each<br>end ground to either a spade<br>or trocar shaped tip. C-Wires<br>are available in four different<br>diameters (0.7mm, 0.89mm,<br>1.14mm and 1.57mm). The<br>wires are made of 316LVM<br>stainless steel. | Orthopaedic fixation pins and<br>wires are metal pins for use in<br>fixation of bone fractures, for<br>bone reconstructions, as guide<br>pins for insertion of other implants<br>or implantation through the skin<br>so that traction may be applied to<br>the skeleton system. To ensure<br>the multi-use of these devices,<br>many different models are<br>available. This differences can<br>be as follows:<br>Diameter: from 0.6 - 6.35mm<br>Length, from 60 up to 500mm<br>Tips: diamond or trocar point,<br>round, flat, with or without 3- or 4-<br>shank ends, with or without<br>spherical shape.<br>Surface: complete or partial<br>smooth and/or threaded, with or<br>without threading cutter. |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for CONMED Corporation. The logo features a stylized, curved shape on the left, followed by the word "CONMED" in a bold, sans-serif font. Below "CONMED" is the word "CORPORATION" in a smaller, sans-serif font.
| | Proposed Device<br>CONMED C-Wire Double<br>Ended Orthopedic Wires | Predicate Device<br>SMT Schilling<br>Kirschner/Guide Wires<br>(K100736) |
|----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended<br>Use/Indications for<br>Use | C-Wires are single use devices. They are intended to be used by trained professionals for the purpose of bone fracture fixation. The C-Wires should only be used with wire drivers that are designed to drive wires having longitudinal flats. | Orthopaedic fixation pins and wires are intended to perform as fixation and stabilization unit of bone fractures or as guidance at insertion of implants into the skeletal system. |
| Material | 316LVM | 316L |
| Conditions of Use | Operating Room in Hospital or<br>Ambulatory Surgical Center for<br>fracture fixation | Operating Room in Hospital or<br>Ambulatory Surgical Center for<br>fracture fixation or guide for<br>implant |
| Scientific<br>Technology | Metal rods inserted using powered handpieces | Metal rods inserted using powered hand pieces |
| Sterilization<br>Method and SAL | Gamma Sterilization at a SAL of 10-6 | Gamma Sterilization at a SAL of 10-6 |
Completed testing includes the following:
- Transportation 0
- Biocompatibility �
- ® Sterilization
- ø Shelf-life
Functional testing on the C-Wires is not required as safety and effectiveness has been established through the following:
- The device has been in continuous use for thirty-eight (38) years with few 1. complaints regarding breakage, insertion or pull-out as detailed in Post Market Surveillance Reports.
- 2. The C-Wire is very similar to other wires cleared by FDA based on a comparison of intended use, engineering drawings, material, method of operation and shape/size. This similarity is documented in the Comparison Chart of Section 13 and summarized below for the predicate device, SMT Schilling Kirschner/Guide Wires (K100736).
- 3. Conformance to FDA standards related to pins and wires as well as other FDA recognized Standards governing aspects of product development and total life cycle.
#### H, Substantial Equivalence
The C-Wire Double Ended Orthopedic Wires have intended use, conditions of use, sterilization method, sterility assurance level and utilizes the same fundamental scientific technology as the predicate device, while raising no new issues of safety or effectiveness. Similarities of the two devices are demonstrated in
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for CONMED Corporation. The logo consists of a stylized letter C on the left, followed by the word "CONMED" in all caps. Below the word "CONMED" is the word "CORPORATION" in smaller, all-caps letters. A horizontal line separates the two words.
a comparison of the engineering drawings as well. Refer to Attachment 14-1. The C-Wire Double Ended Orthopedic Wires are therefore substantially equivalent to the SMT Schilling Kirschner/GuideWires cleared under K100736.
# I. CONCLUSION
Based upon the testing and analysis performed, the C-Wire Double Ended Orthopedic Wire is as safe, as effective, and performs as well as the SMT Schilling Kirschner/Guide Wires (K100736)..
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.