Akros Fibulink Syndesmosis Repair Kit

K162805 · Akros Medical · HTN · Jan 10, 2017 · Orthopedic

Device Facts

Record IDK162805
Device NameAkros Fibulink Syndesmosis Repair Kit
ApplicantAkros Medical
Product CodeHTN · Orthopedic
Decision DateJan 10, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3030
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Akros FibuLink™ Syndesmosis Repair Kit is intended as an adjunct in fracture repair involving metaphyseal and periarticular small bone fragments where screws are not indicated and as an adjunct to fixation systems involving plates, with fracture braces and casting. Specifically, the Akros FibuLink™ Syndesmosis Repair Kit is intended to provide fixation during the healing process following a syndesmotic trauma, such as fixation of syndesmosis (syndesmosis disruptions) in connection with Weber B and C ankle fractures.

Device Story

Multiple-anchor orthopedic fixation system; stainless steel or titanium versions. Used for tibia-fibula fixation during syndesmosis healing. Designed for use with fracture plates accepting 3.5mm cortical screws. Anchors provided pre-threaded/pre-loaded on installation tools in single-use, pre-sterilized kit; includes Kirschner wire and drill bit. Operated by surgeons in clinical settings. Provides mechanical stabilization during healing; allows for fixation through existing fracture repair plates. Benefits patient by providing secure syndesmotic fixation without requiring traditional screw fixation.

Clinical Evidence

No clinical data. Substantial equivalence demonstrated via non-clinical testing including simulated use, lateral pull-to-failure, offset load-to-failure, insertion torque-to-failure, fatigue, fretting, corrosion, MRI compatibility, and biocompatibility (cytotoxicity, pyrogenicity, endotoxin).

Technological Characteristics

Multiple-anchor fixation system; stainless steel or titanium. Compatible with 3.5mm cortical screws per ASTM F543-13. Single-use, pre-sterilized kit. Mechanical fixation principle.

Indications for Use

Indicated for patients requiring fixation of syndesmosis disruptions, including Weber B and C ankle fractures, and as an adjunct in fracture repair of metaphyseal and periarticular small bone fragments where screws are not indicated.

Regulatory Classification

Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of three human profiles facing right, layered on top of each other. These profiles are connected to three flowing shapes below, resembling fabric or ribbons. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 January 10, 2017 Akros Medical Mr. Charles Horrell CEO and Co-Founder 3503 Pleasant Green Rd Durham, North Carolina 27705 Re: K162805 Trade/Device Name: Akros FibuLink Syndesmosis Repair Kit Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: HTN Dated: December 9, 2016 Received: December 12, 2016 Dear Mr. Horrell: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in {1}------------------------------------------------ the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely, # Mark N. Melkerson -S Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ #### Indications for Use K162805 510(k) Number (if known): _ Device Name: Akros FibuLink™ Syndesmosis Repair Kit Indications for Use: The Akros FibuLink™ Syndesmosis Repair Kit is intended as an adjunct in fracture repair involving metaphyseal and periarticular small bone fragments where screws are not indicated and as an adjunct to fixation systems involving plates, with fracture braces and casting. Specifically, the Akros FibuLink™ Syndesmosis Repair Kit is intended to provide fixation during the healing process following a syndesmotic trauma, such as fixation of syndesmosis (syndesmosis disruptions) in connection with Weber B and C ankle fractures. Prescription Use __X_ (Part 21 CFR 801 Subpart D) Over-The-Counter Use _ AND/OR (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Page 1 of 1 {3}------------------------------------------------ 510(k) Summary | Prepared: | September 30, 2016 | |-----------------------|----------------------------------------------------------------------------------------------------------| | Submitter: | Akros Medical<br>3503 Pleasant Green Rd<br>Durham, NC 27705 | | Contact: | Charles Horrell<br>Chief Executive Officer and Co-Founder<br>248.259.5535<br>chuck@akrosmedical.com | | Proprietary Name: | Akros FibuLink™ Syndesmosis Repair Kit | | Common Name: | Syndesmosis Repair Kit | | Regulation and Class: | 21 CFR 888.3030 Single/multiple component metallic bone<br>fixation appliances and accessories; Class II | | Product Code: | 87/HTN | | Predicate Device: | • Arthrex TightRope™ Syndesmosis Device, K043248 | #### Device Description: The Akros FibuLink™ Syndesmosis Repair Kit is a multiple-anchor orthopedic fixation device system, offered in both a stainless steel and a titanium version. The system is designed as an adjunct in repair of unstable joints, specifically as a means to provide fixation between the tibia and fibula during the healing process following a disruption of the ankle syndesmosis. If a fracture of the fibula is present, the design of the fibula anchor permits placement through any fibula fracture repair plate that can accept a 3.5mm cortical screw conforming to ASTM F543-13. The anchors are provided pre-threaded and pre-loaded on their installation tools in a one-time-use pre-sterilized kit, with the Kirschner wire and drill bit needed for site preparation included. #### Intended Use / Indications: The Akros FibuLink™ Syndesmosis Repair Kit is intended as an adjunct in fracture repair involving metaphyseal and periarticular small bone fragments where screws are not indicated and as an adjunct to fixation systems involving plates, with fracture braces and casting. Specifically, the Akros FibuLink™ Syndesmosis Repair Kit is intended to provide fixation during the healing process {4}------------------------------------------------ following a syndesmotic trauma, such as fixation of syndesmosis disruptions) in connection with Weber B and C ankle fractures. ## Summary of Technologies/Substantial Equivalence: The FibuLink™ Syndesmosis Repair Kit is substantially equivalent to the predicate device in terms of its intended use and indications, materials, basic design and mechanical performance. Any noted differences do not raise new types of safety and effectiveness questions, nor are there new technological issues. ## Non-Clinical Testing: Simulated use, lateral pull-to-failure, offset load-to-failure, insertion torque-to-failure, fatigue with pull-to-failure, fretting, pitting corrosion, magnetic resonance, cytotoxicity, pyrogenicity and endotoxin testing were conducted. The results of these tests indicate that the performance of the FibuLink™ device is adequate for its intended use. Endotoxin results demonstrated that the FibuLink™ device met the recommended limit of <20 EU/device. ## Clinical Testing: Clinical testing was not necessary to demonstrate substantial equivalence of the FibuLink™ Syndesmosis Repair Kit to the predicate device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...