← Product Code [HSX](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HSX) · K181302

# Bodycad Unicompartmental Knee System (K181302)

_Bodycad Laboratories, Inc. · HSX · Jan 17, 2019 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HSX/K181302

## Device Facts

- **Applicant:** Bodycad Laboratories, Inc.
- **Product Code:** [HSX](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HSX.md)
- **Decision Date:** Jan 17, 2019
- **Decision:** SESE
- **Submission Type:** Abbreviated
- **Regulation:** 21 CFR 888.3520
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The patient-specific Bodycad Unicompartmental Knee System (Bodycad UKS) is indicated for unicompartmental knee arthroplasty (UKA) in patients with advanced knee osteoarthritis (OA) of the medial compartment with evidence of adequate healthy bone to support the implanted components. Candidates for unicompartmental knee replacement include those with: · joint impairment due to osteoarthritis or traumatic arthritis of the knee, - · varus deformity of the knee, and - · as an alternative to tibial osteotomy in patients with unicompartmental OA. The patient-specific Bodycad UKS components fit within an envelope of dimensions that are specific to each patient. The Bodycad UKS femoral component and tibial baseplate are intended for cemented fixation. The Bodycad screws must be used for fixation of the femoral and tibial components.

## Device Story

Patient-specific unicompartmental knee prosthesis; replaces medial tibiofemoral compartment. Inputs: patient MRI or CT images; surgeon input. Transformation: Bodycad software generates CAD/CAM files for custom femoral/tibial components and single-use cutting guides. Output: patient-specific implants and surgical instruments. Used in clinical setting by orthopedic surgeons. Implants require cemented fixation with Bodycad screws. Benefits: anatomical fit tailored to patient geometry; facilitates surgical precision via custom guides.

## Clinical Evidence

Bench testing only. Validation performed to compare CT and MRI imaging modalities for 3D digital bone model generation. Results demonstrated substantial equivalence between imaging modalities for the Bodycad Unicompartmental Knee System.

## Technological Characteristics

Patient-specific unicompartmental knee system. Materials: CoCrMo (ASTM F1537-11) femoral component, Ti6Al4V ELI (ASTM F136-13) tibial component, UHMWPE (ASTM F648-14) tibial insert. Cemented fixation. Gamma radiation sterilization. Includes custom implants and single-use cutting guides.

## Regulatory Identification

A knee joint femorotibial metal/polymer non-constrained cemented prosthesis is a device intended to be implanted to replace part of a knee joint. The device limits minimally (less than normal anatomic constraints) translation in one or more planes. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral condylar resurfacing component or components made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components made of ultra-high molecular weight polyethylene and are intended for use with bone cement (§ 888.3027).

## Predicate Devices

- Bodycad Unicompartmental Knee System ([K163700](/device/K163700.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

January 17, 2019

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Bodycad Laboratories Inc. % Robert Poggie President BioVera, Inc. 65 Promenade Saint Louis Notre-Dame-de-L'Ile-Perrot. Ouebec. J7V 7P2. CA

#### Re: K181302

Trade/Device Name: Bodycad Unicompartmental Knee System Regulation Number: 21 CFR 888.3520 Regulation Name: Knee Joint Femorotibial Metal/Polymer Non-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: HSX Dated: December 19, 2018 Received: December 26, 2018

#### Dear Robert Poggie:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Peter G. Allen -Date: 2019.01.17 15:15:52 ഗ

FOR Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K181302

Device Name

Bodycad Unicompartmental Knee System

#### Indications for Use (Describe)

The patient-specific Bodycad Unicompartmental Knee System (Bodycad UKS) is indicated for unicompartmental knee arthroplasty (UKA) in patients with advanced knee osteoarthritis (OA) of the medial compartment with evidence of adequate healthy bone to support the implanted components. Candidates for unicompartmental knee replacement include those with:

· joint impairment due to osteoarthritis or traumatic arthritis of the knee,

- · varus deformity of the knee, and
- · as an alternative to tibial osteotomy in patients with unicompartmental OA.

The patient-specific Bodycad UKS components fit within an envelope of dimensions that are specific to each patient. The Bodycad UKS femoral component and tibial baseplate are intended for cemented fixation. The Bodycad screws must be used for fixation of the femoral and tibial components.

Type of Use (Select one or both, as applicable)

| <span></span> | Prescription Use (Part 21 CFR 801 Subpart D) |
|---------------|----------------------------------------------|
| <span></span> | Over-The-Counter Use (21 CFR 801 Subpart C)  |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) SUMMARY

## Bodycad Unicompartmental Knee System

In accordance with 21 CFR 807.92 of the Federal Code of Regulations, the following information is a summary of safety and effectiveness of the Bodycad Unicompartmental Knee System.

### A. SUBMITTERS INFORMATION

| Submitter Name:     | BioVera, Inc.                                                                    |
|---------------------|----------------------------------------------------------------------------------|
| Submitter Address:  | 65 Promenade Saint-Louis, Notre-Dame-De-L'Ile-Perrot, Québec,<br>J7V 7P2, CANADA |
| Contact Person:     | Robert A Poggie, PhD                                                             |
| Phone Number:       | (514) 901-0796; (514) 349-7226                                                   |
| Fax Number:         | (514) 901-0796                                                                   |
| Date of Submission: | December 19, 2018                                                                |

# B. DEVICE IDENTIFICATION & MANUFACTURER

| Manufacturer Name:    | Bodycad Laboratories, Inc.                                      |
|-----------------------|-----------------------------------------------------------------|
| Manufacturer Address: | 2035 rue du Haut-Bord, Quebec, QC, G1N 4R7, Canada              |
| Registration Number:  | 3012086398                                                      |
| Contact Name:         | Guy Sévigny                                                     |
| Title:                | Director, Regulatory Affairs                                    |
| Device Trade Name:    | Bodycad Unicompartmental Knee System                            |
| Device Common Name:   | Unicondylar knee device                                         |
| Classification Name:  | Knee joint femorotibial metal/polymer non-constrained, cemented |
| Classification Code:  | HSX – Class II                                                  |
| Classification Panel: | Orthopedic                                                      |
| Regulation Number:    | 21 CFR section 888.3520                                         |

## C. PREDICATE DEVICE

K163700

The Bodycad Unicompartmental Knee System (Bodycad UKS)

{4}------------------------------------------------

#### D. DEVICE DESCRIPTION

The Bodycad Unicompartmental Knee System (Bodycad UKS) is a patient-specific prosthesis that consists of femoral and tibial implants for replacement of the medial tibiofemoral compartment of the knee. The patient-specific femoral and tibial components and single-use cutting guides are manufactured from CAD and CAM files generated from validated Bodycad software, which are based on MRI or CT images of the patient's knee and input from the surgeon. The Bodycad UKS is for cemented use only. The Bodycad UKS is sterilized by gamma radiation.

Materials: CoCrMo (ASTM F1537-11) for the femoral component, Ti6Al4V ELI (ASTM F136-13) for the tibial component, UHMWPE (F648-14) for the tibial insert

#### E. INDICATIONS FOR USE

The patient-specific Bodycad Unicompartmental Knee System (Bodycad UKS) is indicated for unicompartmental knee arthroplasty (UKA) in patients with advanced knee osteoarthritis (OA) of the medial compartment with evidence of adequate healthy bone to support the implanted components. Candidates for unicompartmental knee replacement include those with:

- joint impairment due to osteoarthritis or traumatic arthritis of the knee, .
- . varus deformity of the knee, and
- . as an alternative to tibial osteotomy in patients with unicompartmental OA.

The patient-specific Bodycad UKS components fit within an envelope of dimensions that are specific to each patient. The Bodycad UKS femoral component and tibial baseplate are intended for cemented fixation. The Bodycad screws must be used for fixation of the femoral and tibial components.

#### F. TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE

The Bodycad UKS is a patient-specific unicompartmental knee system comprised of two groups of components: (1) prosthetic components / implants, and (2) instrument components. The first group includes a series of devices that are implanted in the patient's body while the second group includes instruments that facilitate implantation of the components. Both groups of components are manufactured to match the patient's anatomy based on MRI or CT images. The implants and instruments are packaged and terminally sterilized with gamma radiation to a sterility assurance level of 10%. The technological characteristics and manufacturing methods of the predicate and subject devices are identical.

#### G. PERFORMANCE DATA

Validation testing was performed for CT and MRI imaging modalities and demonstrated substantial equivalence of the 3D digital bone models being equivalent for data extracted from CT and MR images.

#### H. CONCLUSION

The results of validation testing demonstrated substantial equivalence of CT and MR imaging modalities for the Bodycad Unicompartmental Knee System.

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HSX/K181302](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HSX/K181302)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
