← Product Code [HSN](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HSN) · K133585

# INBONE TOTAL ANKLE (K133585)

_Wrightmedicaltechnologyinc · HSN · Feb 20, 2014 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HSN/K133585

## Device Facts

- **Applicant:** Wrightmedicaltechnologyinc
- **Product Code:** [HSN](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HSN.md)
- **Decision Date:** Feb 20, 2014
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3110
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The INBONE® Total Ankle is indicated for patients with ankle joints damaged by severe rheumatoid, posttraumatic, or degenerative arthritis. The INBONE® Total Ankle is additionally indicated for patients with a failed previous ankle surgery. CAUTION: The ankle prosthesis is intended for cemented use only.

## Device Story

INBONE® Total Ankle System is a modular total ankle replacement (TAR) prosthesis. Components include tibial stem, tibial tray, poly bearing, talar dome, and talar stem. Device is implanted by orthopedic surgeons in a clinical/surgical setting to replace damaged ankle joints. Modularity allows component sizing based on patient anatomy. System requires cement for fixation. Line extension adds smaller components (Size 1 talar dome, Size 1+ poly bearing), thicker poly bearings for revision cases, and tibial stems with plasma coating removed. Device restores joint function and alleviates pain associated with arthritis or failed prior surgery.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

Modular ankle prosthesis; components include tibial stem, tray, poly bearing, talar dome, and talar stem. Features include plasma-coated and non-coated tibial stems. Cemented fixation required. Design modifications include size variations and bearing thickness adjustments. No software or electronic components.

## Regulatory Identification

An ankle joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace an ankle joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces and has no linkage across-the-joint. This generic type of device includes prostheses that have a talar resurfacing component made of alloys, such as cobalt-chromium-molybdenum, and a tibial resurfacing component made of ultra-high molecular weight polyethylene and is limited to those prostheses intended for use with bone cement (§ 888.3027).

## Predicate Devices

- INFINITY™ Total Ankle ([K123954](/device/K123954.md))
- INBONE® II Total Ankle System (K10088)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/2 description: The image shows the word "WRIGHT." in a bold, sans-serif font. Below the word is a stylized graphic that appears to be two overlapping, elongated triangles pointing in opposite directions. The image is black and white and has a slightly distressed or textured appearance.

Wright Medical Technology, Inc. 5677 Airline Road Arlington, TN 38002 www.wmt.com

# 510(K) SUMMARY

In accordance with the Food and Drug Administration Rule to implement provisions of the Safe Medical Devices Act of 1990 and in conformance with 21 CFR 807.92, this information serves as a Summary of Safety and Effectiveness for the INBONE® Total Ankle System Line Extension.

| (a)(1). Submitted By:              | Wright Medical Technology, Inc.<br>5677 Airline Road<br>Arlington, TN 38002                     |
|------------------------------------|-------------------------------------------------------------------------------------------------|
| Date:                              | December 3, 2013                                                                                |
| Contact Person:                    | Val Myles<br>Regulatory Affairs Specialist I<br>Office - (901) 290-5162<br>Fax - (901) 867-4190 |
| (a)(2). Proprietary Name:          | INBONE® Total Ankle System                                                                      |
| Common Name:                       | Ankle Prosthesis                                                                                |
| Classification Name and Reference: | 21 CFR 888.3110 – Class II                                                                      |
| Device Product Code, Device Panel: | HSN: Ankle Prosthesis                                                                           |
| (a)(3). Predicate Device:          | K123954 INFINITY™ Total Ankle & K10088<br>INBONE® II Total Ankle System                         |

# (a)(4). Device Description

The INBONE® implant is a total ankle replacement (TAR) that is made of several specific components. A key concept of the INBONE® device is its modularity. Based on patient anatomy, a number of component sizes can be optioned for best fit. These components include the tibial stem components, the tibial tray, the poly bearing, the talar dome, and the talar stem.

# (a)(5). Intended Use

The INBONE® Total Ankle is indicated for patients with ankle joints damaged by severe rheumatoid, posttraumatic, or degenerative arthritis. The INBONE® Total Ankle is additionally indicated for patients with a failed previous ankle surgery. CAUTION: The ankle prosthesis is intended for cemented use only.

{1}------------------------------------------------

### (a)(6). Technological Characteristics Comparison

The INBONE® Total Ankle Line Extension is technologically substantially equivalent to the predicates. A summary of the changes is shown below:

- Addition of a Size 1 Talar Dome and Size 1 + Poly Bearing .
- . Addition of thicker poly bearings for revisions
- Trials for all of the poly bearings in this line extension and the size 1 talar dome .
- Mid and top tibial stem pieces with the plasma coating removed .

### (b)(1). Substantial Equivalence - Non-Clinical Evidence

Performance testing and engineering analysis related to articular shear stability, contact pressure, and the lock detail supports the equivalence of the subject device and shows that no new worse-case devices are introduced in this system. The safety and effectiveness of the INBONE® Total Ankle Line Extension is adequately supported by the testing rationales, substantial equivalence information, materials information, and comparison of design characteristics provided within this premarket notification.

### (b)(2). Substantial Equivalence - Clinical Evidence

N/A

### (b)(3). Substantial Equivalence - Conclusions

The design characteristics of the subject system do not affect the safety or effectiveness of the system. From the evidence submitted in this 510(k), the subject devices can be expected to perform substantially equivalent to the predicate system.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles an abstract human figure or a caduceus, a symbol often associated with medicine and healthcare.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 20, 2014

Wright Medical Technology, Incorporated Val Myles Regulatory Affairs Specialist I 5677 Airline Road Arlington, Tennessee 38002

Re: K133585

Trade/Device Name: InBone® Total Ankle Line Extension Regulation Number: 21 CFR 888.3110 Regulation Name: Ankle joint metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: HSN Dated: January 16, 2014 Received: January 24, 2014

Dear Val Myles:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register,

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{3}------------------------------------------------

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

# Lori A. Wiggins

- for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

510(k) Number (if known) K133585

# Device Name

INBONE® Total Ankle Line Extension

#### Indications for Use (Describe)

The INBONE® Total Ankle is indicated for patients with ankle joints damaged by severe rheumatic, or degenerative arthritis.

The INBONE® Total Ankle is additionally indicated for patients with a failed previous ankle surgery.

CAUTION: The ankle prosthesis is intended for cemented use only.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

বার বিষয়কে পারে অন্যান্য করে তার করে না করে না করে না করে না করে আর করে আর করে আর করে আর করে আর করে আর করে আর করে আর করে আর করে আর করে আর করে আর করে আর করে আর করে আর করে আর FOR FOR FDA USE ONLY ..: Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

Division

{5}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HSN/K133585](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HSN/K133585)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
