K063578 · Plus Orthopedics AG · HSD · Mar 2, 2007 · Orthopedic
Device Facts
Record ID
K063578
Device Name
PROMOS MODULAR SHOULDER SYSTEM
Applicant
Plus Orthopedics AG
Product Code
HSD · Orthopedic
Decision Date
Mar 2, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3690
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The PROMOS® Modular Shoulder System is indicated for: - Advanced degeneration of the shoulder joint as a result of degenerative, post-traumatic or inflammatory arthritis - Avascular necrosis of the humeral head - Complex fractures of the proximal Humerus - Functional impairment especially in the case of post-traumatic loss of the joint configuration The humeral component is intended for cemented or cementless use. The glenoid component is for use with bone cement only.
Device Story
Modular shoulder prosthesis system; restores shoulder motion and eliminates pain. Components: glenoid (UHMWPe) and modular humeral (titanium alloy stem/body, CoCrMo head). Humeral assembly uses Morse taper; includes size 3.5 cementless stem, Monoblock stems (sizes 30, 35, 40mm), and varied humeral heads (14 sizes/eccentricities). Features face gear interface between inclination set and body. Used by orthopedic surgeons in clinical settings for total or hemi-shoulder arthroplasty. Provides structural replacement for damaged joint surfaces; improves patient mobility and reduces pain.
Clinical Evidence
Bench testing only. No clinical data provided. Design verification and validation performed per 21 CFR 820.30. Materials conform to ISO 5834-2, ISO 5832-3, and ISO 5832-12 standards.
Technological Characteristics
Modular shoulder prosthesis. Materials: UHMWPe (ISO 5834-2) for glenoid; Ti6Al4V (ISO 5832-3) for stems/bodies; CoCrMo (ISO 5832-12) for heads. Features: Morse taper connections, face gear interface, cementless/cemented options. Sterilization: Radiation (ISO 11137).
Indications for Use
Indicated for patients with advanced shoulder joint degeneration (degenerative, post-traumatic, or inflammatory arthritis), avascular necrosis of the humeral head, complex proximal humerus fractures, or functional impairment due to post-traumatic joint loss.
Regulatory Classification
Identification
A shoulder joint humeral (hemi-shoulder) metallic uncemented prosthesis is a device made of alloys, such as cobalt-chromium-molybdenum. It has an intramedullary stem and is intended to be implanted to replace the articular surface of the proximal end of the humerus and to be fixed without bone cement (§ 888.3027). This device is not intended for biological fixation.
K102670 — TOTAL SHOULDER SYSTEM · Shoulder Innovations, LLC · Jan 24, 2011
K241817 — InSet Total Shoulder System · Shoulder Innovations, Inc. · Feb 11, 2025
K100448 — ASCENSION MODULAR TOTAL SHOULDER SYSTEM · Ascension Orthopedics, Inc. · Aug 30, 2010
K193038 — Comprehensive Shoulder System · Biomet Manufacturing Corp · Oct 28, 2020
K210899 — LINK Embrace Shoulder System- Anatomical Configuration · Waldemar Link GmbH & Co. KG · Jun 21, 2021
Submission Summary (Full Text)
{0}------------------------------------------------
063578
### D.0 Premarket Notification 510(k) Summary
## D.1 Submitter Information
MAR 0 2 2007
Company Name and Address:
PLUS ORTHOPEDICS AG Erlenstraße 4a CH-6343 Rotkreuz Switzerland
Contact Name:
Pamela J. Weagraff, Principal Consultant Quintiles Consulting 18 Bridie Lane Norfolk, MA 02056
Telephone: 508-528-1745 Facsimile: 978-752-1225 E-mail: pamela.weagraff(@quintiles.com
Date Prepared: November 27, 2006
### D.2 Name of Device
- 2.1 Trade Name: PROMOS® Modular Shoulder System
- 2.2 Common Name: Hemi- or total shoulder
- Classification Name and Reference: 2.3
Title 21 Code of Federal Regulation (CFR), Part 888.3690, prosthesis, shoulder, hemi-, humeral, metallic uncemented Title 21, CFR, Part 888.3660, prosthesis, shoulder, semi-constrained, metal/polymer cemented
#### E.3 Substantial Equivalence Claimed to Predicate Device
PROMOS® Modular Shoulder System, K032126 3.1
{1}------------------------------------------------
#### Device Description E.4
The modified PROMOS® Modular Shoulder System is intended to aid in the restoration of shoulder motion and elimination of pain. This device consists of two primary components, the glenoid component and the modular humeral component. The changes to the PROMOS® Shoulder that resulted in the modified PROMOS® Shoulder and are subject of this 510(k) include:
- Addition of size 3.5 humeral stem, cementless .
- Addition of six humeral ball heads R21 R26, with 4 mm eccentricity .
- Redesign of body and inclination set to add face gear to area between . inclination set and body
- Addition of three Monoblock stems, cementless, size 02/30, 35 and 40 mm .
- Various labeling changes .
The glenoid component has a spherical articulating surface with four pegs on the inferior surface for attachment to the bone. It is manufactured from ultra high molecular weight polyethylene (UHMWPe) according to ISO 5834-2:1998, Implants for Surgery - Ultra-High-Molecular-Weight Polyethylene - Part 2: Moulded Forms. The glenoid is available in four sizes with each size having three different spherical radii of curvature for a total of twelve glenoid components.
The modular humeral component consists of a distal stem, body, inclination set and humeral head. The distal stem is rectangular in cross-sectional shape, previously available in seven cemented and seven cementless sizes, and now available in seven cemented and eight cementless sizes. It is attached to the body via a Morse type taper.
The cementless and cemented stems are fabricated from titanium alloy (Ti6Al4V) according to ISO 5832-3:1996, Implants for surgery -- Metallic materials -- Part 3: Wrought titanium 6-aluminium 4-vanadium alloy.
The body is made of same titanium alloy as the distal stems and is available in three sizes. The body and inclination set have been redesigned to add a face gear to the area between the inclination set and body. As before the modified inclination set consists of three components: an inclination insert, an offset module, and a ball head screw.
The modular humeral heads are manufactured from wrought cobalt-chromiummolybdenum (CoCrMo) alloy according to ISO 5832-12:1996, Implants for surgery -- Metallic materials -- Part 12: Wrought cobalt-chromium-molybdenum alloy, previously available in eight different sizes/eccentricities, and now available in 14 different sizes/eccentricities.
{2}------------------------------------------------
Since cementless or cemented stems in size 01 may be too large for some patients, the company has developed a humeral stem, size 02. The narrow crosssection of a size 02 stem precluded a modular design, i.e., 02 size stem and a selection of bodies. As a result, the company implemented the size 02 as a Monoblock system with three different heights for body sizes 30mm, 35mm, and 40mm. The design of the cross-section is the same as mentioned above in the description of the distal stem and the stem is made of the same titanium allov. The proximal portion has also been redesigned to add face gear to the area between inclination set and body.
#### દર્શા Intended Use
The PROMOS® Modular Shoulder System is indicated for:
- Advanced degeneration of the shoulder joint as a result of degenerative, post-. traumatic or inflammatory arthritis
- . Avascular necrosis of the humeral head
- Complex fractures of the proximal Humerus .
- Functional impairment especially in the case of post-traumatic loss of the . joint configuration
The humeral component is intended for cemented or cementless use. The glenoid component is for use with bone cement only.
### E.6 Predicate Device Comparison of Indications for Use / Intended Use and Technical Characteristics
The comparison of the modified PROMOS® Modular Shoulder System was based on a review of the Design Control documentation for the design changes, relevant aspects of which are included in the company's 510(k) Premarket Notification, and information concerning the predicate device that was available to the company internally, i.e., PROMOS® Modular Shoulder System, K032126. The comparison considered technical characteristics and the indications for use / intended use. Bench testing was assessed.
#### E.7 Performance Data
- 7.1 Performance Standards (Section 514 Compliance): no applicable performance standards have been promulgated under Section 514 of the Food, Drug and Cosmetic Act, for prosthesis, shoulder, hemi-, humeral, metallic uncemented or prosthesis, shoulder, semi-constrained, metal/polymer cemented. The modified PROMOS® Modular Shoulder System does conform to the following FDA recognized standards:
{3}------------------------------------------------
#### 7.1.1 Sterility
AAMI / ANSI / ISO 11137:1994, Sterilization of health care products - Requirements for validation and routine control -radiation sterilization and ANSI/AAMI/ISO 11137:1994 (Amendment 1:2002)
#### 7.1.2 Biocompatibility
ISO 5834-2:1998, Implants for Surgery - Ultra-High-Molecular-Weight Polyethylene - Part 2: Moulded Forms
ISO 5832-3:1996, Implants for surgery -- Metallic materials -- Part 3: Wrought titanium 6-aluminium 4-vanadium alloy
ISO 5832-12:1996, Implants for surgery -- Metallic materials -- Part 12: Wrought cobalt-chromium-molybdenum alloy
- 7.1.3 Performance Testing: Design verification and design validation, e.g., bench testing was performed according to FDA's Design Control Requirements, Title 21 Code of Federal Regulations, Part 820.30.
#### E.8 Conclusion:
The information and data provided in this 510(k) Premarket Notification establish that the modified PROMOS® Modular Shoulder System is substantially equivalent to the afore-mentioned predicate device with respect to indications for use/intended use, and technical characteristics.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized depiction of a human figure with three arms or lines extending upwards, enclosed within a circle. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" is arranged around the upper portion of the circle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Plus Orthopedics AG % Ouintiles Consulting Ms. Pamela J. Weagraff, MBA, RAC Principal Consultant 18 Bridie Lane Norfolk, MA 02056
MAR 0 2 2007
K063578 Re: Trade/Device Name: PROMOS® Modular Shoulder System Regulation Number: 21 CFR 888.3690 Regulation Name: Shoulder joint humeral (hemi-shoulder) metallic uncemented prosthesis. Regulatory Class: II Product Code: HSD, KWS Dated: February 2, 2007 Received: February 5, 2007
Dear Ms. Weagraff:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good
{5}------------------------------------------------
manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark N. Milkerson
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# ODE Indications Statement
5001 510(k) Number (if known): Unknown ¥ 06 3 5 7 8
Device Name: PROMOS® Modular Shoulder System
### Indications for Use:
The PROMOS Modular Shoulder System is indicated for:
- Advanced degeneration of the shoulder joint as a result of degenerative, post-traumatic . or inflammatory arthritis
- Avascular necrosis of the humeral head �
- Complex fractures of the proximal humerus .
- Functional impairment especially in the case of post-traumatic loss of the joint . configuration
The humeral component is intended for cementless use. The glenoid component is for use with bone cement only.
AND/OR
Prescription Use: X (21 CFR 801 Subpart D) Over-the-Counter Use: (21 CFR 807 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Mark M. Miller
Division Sign-Off Division of Genera'. Restorative, and Neurological L . vices
510(k) Number K063578
Traditional 510(k) - PROMOS® Modular Shoulder System November 27, 2006
Company Confidential Page 9 of 70
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.