COPELAND RESURFACING HUMERAL HEADS

K010664 · Biomet Manufacturing, Inc. · HSD · Apr 5, 2001 · Orthopedic

Device Facts

Record IDK010664
Device NameCOPELAND RESURFACING HUMERAL HEADS
ApplicantBiomet Manufacturing, Inc.
Product CodeHSD · Orthopedic
Decision DateApr 5, 2001
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3690
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Copeland Resurfacing Humeral Heads are indicated for the following conditions where the humeral head and neck are of sufficient bone stock and there is presence of an intact or reconstructable rotator cuff which is necessary for proper functioning and dislocation resistance: 1) Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis. 2) Rheumatoid arthritis 3) Correction of functional deformity 4) Reconstructable rotator cuff 5) Treatment of fractures of the humeral head 6) Traumatic arthritis For cemented use only.

Device Story

Copeland Resurfacing Humeral Heads are orthopedic implants for hemi- or total shoulder replacement. Designed to preserve bone stock by replacing only the defective humeral surface; features tapered, fluted stem for stability. Available in four sizes; small, medium, and large share radius of curvature with varying heights; extra-large features larger radius. Used in surgical procedures by orthopedic surgeons. Implants are cemented into the humerus. Provides alternative to traditional total shoulder arthroplasty where more bone is removed. Compatible with Biomet BioModular or Integrated Glenoids (extra-large restricted to BioModular).

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Materials: Cobalt-chrome-molybdenum alloy (ASTM F-75) with MacroBond™ (Ti-Al-4V, ASTM F-1580) coating on non-articulating surfaces. Design: Tapered, fluted stem; four sizes (small, medium, large, extra-large). Energy source: None (mechanical implant). Connectivity: None. Sterilization: Not specified.

Indications for Use

Indicated for patients with sufficient humeral head/neck bone stock and intact/reconstructable rotator cuff requiring shoulder replacement due to non-inflammatory degenerative joint disease (osteoarthritis, avascular necrosis), rheumatoid arthritis, functional deformity, humeral head fractures, or traumatic arthritis. For cemented use only.

Regulatory Classification

Identification

A shoulder joint humeral (hemi-shoulder) metallic uncemented prosthesis is a device made of alloys, such as cobalt-chromium-molybdenum. It has an intramedullary stem and is intended to be implanted to replace the articular surface of the proximal end of the humerus and to be fixed without bone cement (§ 888.3027). This device is not intended for biological fixation.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # APR - 5 2001 ### SUMMARY OF SAFETY AND EFFECTIVENESS KO10664 1000 | Sponsor: | Biomet, Inc.<br>Airport Industrial Park<br>Warsaw, Indiana 46580 | |----------------------|-------------------------------------------------------------------| | Contact Person: | Sara A. Bailey<br>(219) 372-1568 | | Device: | Copeland Resurfacing Humeral Heads | | Classification Name: | Shoulder joint metal/polymer/ non-constrained cemented prosthesis | | Device Product Code: | KWT (21 CFR 888.3650) | #### Intended Use: The Copeland Resurfacing Humeral Heads are indicated for the following conditions where the humeral head and neck are of sufficient bone stock and there is presence of an intact or reconstructable rotator cuff which is necessary for proper functioning and dislocation resistance: - 1) Non-inflammatory degenerative joint disease including osteoarthritis and avascular necrosis. - 2) Rheumatoid arthritis - 3) Correction of functional deformity - 4) Reconstructable rotator cuff - 5) Treatment of fractures of the humeral head - 6) Traumatic arthritis For cemented use only. #### Device Description: These devices are cemented and are designed to maintain maximum bone stock by removing minimal bone and replacing only the defective surface. Copeland Resurfacing Humeral Heads can be used in hemi- or total shoulder replacement surgical procedures. By preserving the bone stock, this device gives a patient an alternative to other total shoulder devices where more bone is removed. The humeral head components are available in four sizes - small, medium, large, and extra-large. The small, medium, and large humeral heads have the same radius of curvature, but the heights differ to cater for the range of anatomical sizes and offsets. The extra-large head has a larger radius of curvature to accommodate a patient with larger bone stock. The stem is tapered and fluted to provide maximum stability in the ## 00072 {1}------------------------------------------------ humerus. The components are manufactured from cobalt-chrome-molybdenum alloy (ASTM F-75) and coated with MacroBond™ (Ti-Al-4V, ASTM F-1580) on the nonarticulating surface. KO106664 2017 The small, medium, and large humeral head components can be used with Biomet's BioModular Glendoids or Integrated Glenoids. The extra-large head can only be used with Biomet's BioModular Glenoids due to the larger spherical radius. Potential Risks: The potential risks associated with this device are the same as with any joint replacement device. These include, but are not limited to: - l ) Material sensitivity reactions. Implantation of foreign material in tissues can result in histological reactions involving various sizes of macrophages and fibroblasts. The clinical significance of this effect is uncertain, as similar changes may occur as a precursor to or during the healing process. Particulate wear debris and discoloration from metallic and polyethylene components of joint implants may be present in adjacent tissue or fluid. It has been reported that wear debris may initiate a cellular response resulting in osteolysis or osteolysis may be a result of loosening of the implant. - 2) Early or late postoperative infection and allergic reaction. - 3) Intraonerative bone perforation or fracture may occur, particularly in the presence of poor bone stock caused by osteoporosis, bone defects from previous surgery, bone resorption, or while inserting the device. - 4) Loosening or migration of the implants can occur due to loss of fixation, trauma, malalignment, bone resorption, excessive activity. - 5) Prearticular calcification or ossification, with or without impediment of joint mobility. - 6) Inadequate range of motion due to improper selection or positioning of components. - 7) Undesirable shortening of limb. - 8) Dislocation and subluxation due to inadequate fixation and improper positioning. Muscle and fibrous tissue laxity can also contribute to these conditions. - 9) Fatigue fracture of component can occur as a result of loss of fixation, strenuous activity, malalignment, trauma, non-union, or excessive weight. - 10) Fretting and crevice corrosion can occur at interfaces between components. - 11) Wear and/or deformation of articulating surfaces. - 12) Accelerated wear of glenoid articular cartilage. - 13) Postoperative bone fracture and pain. Predicate Device(s): Copeland Resurfacing Heads, K003044 ## 00073 {2}------------------------------------------------ ### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized human figure or a bird in flight, composed of three curved lines. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 APR = 5 2001 Ms. Sara A. Bailey Regulatory Affairs Specialist Biomet, Inc. Airport Industrial Park P.O. Box 587 Warsaw, Indiana 46581-0587 Re: K010664 Trade Name: Copeland Resurfacing Humeral Heads Regulatory Class: II Product Code: HSD and KWT Dated: March 2, 2001 Received: March 6, 2001 Dear Ms. Bailey: We have reviewed your Section 510(k) notification of intent to market the device referenced we have reviewed your Section > rotification of clice in the midications for use to May 28, 1076, the above and we nave delemined the devices marketed in interstate commerce prior to May 28, 1976, the stated in the enclosure) to device Amendments, or to devices that have been reclassified in enactment date of the Medical Dovice Fullendments, or og, and Cosmetic Act (Act). You mays accordance with the provisions of the reachers' room a controls provisions of the Act of devices therefore, market the device, subject to the genirements for annual registration, listing of devices, controls provisions of the Act mende requirentibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III If your device is classified (sec above) into claim additional (casting major regulations). (Premarket Approval), it may be subject to such additional (Title 21). Register ( (Premarket Apployal), it may of subject to ater a Regulations, Title 21, Parts 800 to 895. affecting your device can be found in the Code of Federal Regulations, Title 21, Pa allecting your device our of round ation assumes compliance with the current Good A Substantially Squrvalent decemmant, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to inspections, the rood and Drug Nammount in regulatory action. In addition, FDA may publish Contiply with the GMI Yegataton may roundevice in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation your night have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. This letter will allow you to begin marketing your device as described in your 510(k) prematket notification. The FDA finding of substantial equivalence of your device to a legally marketed nonfication. The I Drice results in a classification for your device and thus, permits your device to proceed to the market. {3}------------------------------------------------ Page 2 - Ms. Sara A. Bailey If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at additionally 607.10 101 in 10 - 10 questions on the promotion and advertising of your device, (2017) 594-4639. Nationally 10. quee at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small monmation on your responsibility free number (800) 638-2041 or at (301) 443-6597, or at its İnternet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Mark N Millican Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Page 1 of 1 510 (k) Number (if known) : __________________________________________________________________________________________________________________________________________________ Device Name: Copeland Resurfacing Humeral Heads Indications For Use: The Copeland Resurfacing Humeral Heads are indicated for the following conditions where the humeral head and neck are of sufficient bone stock and there is presence of an intact or reconstructable rotator cuff which is necessary for proper functioning and dislocation resistance: KOID664 - 1) Non-inflammatory degenerative joint disease including - osteoarthritis and avascular necrosis. - 2) Rheumatoid arthritis - 3) Correction of functional deformity - 4) Reconstructable rotator cuff - 5) Treatment of fractures of the humeral head - 6) Traumatic arthritis For cemented use only. Mark N Mulkeim (Division Sign-Off) Division of General, Resto tive and Neurological Devices 510(k) Number ***_***__ K010664 (PLEASE DO NOT WRITE BELOW THIS LINE. CONTINUE ON ANOTHER PAGE IF NEEDED.) 000008 Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) OR Over-The-Counter-Use (Optional Format 1-2-96)
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...