Estremo Fibular Nail

K250197 · Citieffe S.R.L. · HSB · Oct 17, 2025 · Orthopedic

Device Facts

Record IDK250197
Device NameEstremo Fibular Nail
ApplicantCitieffe S.R.L.
Product CodeHSB · Orthopedic
Decision DateOct 17, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3020
Device ClassClass 2
AttributesTherapeutic

Intended Use

Estremo Fibular Nail is intended for use in the fixation of fibula fractures and osteotomies.

Device Story

Estremo Fibular Nail is an intramedullary fixation system for fibula fractures and osteotomies. The system comprises nails, screws, and end caps. A unique 'Diapason wings deployment' mechanism at the proximal end of the rod facilitates proximal fixation via a screw. The device is used by surgeons in clinical settings to stabilize bone segments. By providing intramedullary fixation, the device supports bone healing and fracture alignment. The system is provided sterile and is intended for prescription use.

Clinical Evidence

No clinical data. Substantial equivalence is supported by bench testing, including static torsion, driving torque, pull-out, self-tapping performance, bending fatigue, and static four-point bend tests performed in accordance with ASTM F1264 and ASTM F543 standards.

Technological Characteristics

Materials: 316L stainless steel (ASTM F138). Design: Angled intramedullary nail with 'Diapason wings deployment' proximal fixation mechanism. Standards: ASTM F1264 (IMFD), ASTM F543 (screws). Sterilization: Provided sterile. Connectivity: None.

Indications for Use

Indicated for the fixation of fibula fractures and osteotomies in patients requiring intramedullary stabilization.

Regulatory Classification

Identification

An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION October 17, 2025 Citieffe s.r.l. Federica Giovetti Quality and Regulatory Manager Via Armaroli 21 Calderara di Reno, BO 40012 Italy Re: K250197 Trade/Device Name: Estremo Fibular Nail Regulation Number: 21 CFR 888.3020 Regulation Name: Intramedullary fixation rod Regulatory Class: Class II Product Code: HSB Dated: September 17, 2025 Received: September 17, 2025 Dear Federica Giovetti: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K250197 - Federica Giovetti Page 2 (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). {2} K250197 - Federica Giovetti Page 3 Sincerely, Joseph P. Russell - S Digitally signed by Joseph P. Russell-S Date: 2025.10.17 09:04:10 -04'00" for: Farzana Sharmin, PhD Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. Submission Number (if known) K250197 Device Name Estremo Fibular Nail Indications for Use (Describe) Estremo Fibular Nail is intended for use in the fixation of fibula fractures and osteotomies. Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) **CONTINUE ON A SEPARATE PAGE IF NEEDED.** This section applies only to requirements of the Paperwork Reduction Act of 1995. **DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.** The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov “An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.” {4} K250197 Page 1 of 2 | 510(k) #: K250197 | 510(k) Summary | Prepared on: 2025-10-15 | | --- | --- | --- | | Contact Details | | 21 CFR 807.92(a)(1) | | Applicant Name | Citieffe s.r.l. | | | Applicant Address | Via Armaroli 21, Calderara di Reno BO 40012 - Italy | | | Applicant Contact Telephone | 00390510543232 | | | Applicant Contact | Mrs. Federica Giovetti | | | Applicant Contact Email | giovetti@citieffe.com | | | Device Name | | 21 CFR 807.92(a)(2) | | Device Trade Name | Estremo Fibular Nail | | | Common Name | Rod, Fixation, Intramedullary And Accessories | | | Classification Name | Intramedullary fixation rod | | | Regulation Number | 888.3020 | | | Product Code(s) | HSB | | | Legally Marketed Predicate Devices | | 21 CFR 807.92(a)(3) | | Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code | | K173656 | Arthrex FibuLock Nail | HSB | | K192710 | The Simple Locking Intramedullary (Slim) System | HSB | | Device Description Summary | | 21 CFR 807.92(a)(4) | | Estremo Fibular Nail includes all implants and instruments required for the fixation of fibula fractures and osteotomies. Estremo Fibular Nail includes the nails, screws, end caps and related instruments. Estremo Fibular Nail Nail differs from traditional nails or rods as it utilizes a "Diapason wings deployment" mechanism at the proximal end of the rod to allow for proximal fixation with a screw. The implants are composed of 316L stainless steel per ASTM F138 and are provided sterile. | | | | Intended Use/Indications for Use | | 21 CFR 807.92(a)(5) | | Estremo Fibular Nail is intended for use in the fixation of fibula fractures and osteotomies. | | | | Indications for Use Comparison | | 21 CFR 807.92(a)(5) | | Estremo Fibular Nail has the same indications for use of the primary predicate device Arthrex FibuLock Nail (K173656) | | | | Technological Comparison | | 21 CFR 807.92(a)(6) | {5} K250197 Page 2 of 2 Estremo Fibular Nail is similar to the legally marketed primary predicate device with respect to intended use and design. The technological characteristics between the subject device, primary predicate and additional predicate are similar. The primary predicate device and the subject device are both angled nails and both devices allow the use of one or two syndesmotic screws. Moreover, both have proximal locking mechanism and distal locking with screws. However, the lengths of the subject device nails are not completely within the range of the primary predicate; therefore, an additional predicate that features a wider range of lengths has been introduced for an exhaustive mechanical comparison. Estremo Fibular Nail differs from traditional nails or rods as it utilizes a "Diapason wings deployment" mechanism at the proximal end of the rod to allow for proximal fixation with a screw. ## Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b) All necessary testing has been performed on representative Estremo Fibular Nail components (nails and screws) to assure substantial equivalence to its predicate and demonstrate the subject device performs as intended. All testing was performed on finished devices. The Estremo Fibular Nail has been tested in accordance with ASTM F1264, the screws have been tested in accordance with ASTM F543. A list of the tests carried for this application can be found below: - Static torsion mechanical testing of IMFD - Static torsion mechanical testing of IMFD Locking Screws - Driving torque mechanical testing of IMFD Locking Screws - Pull-out mechanical testing of IMFD Locking Screws - Self-tapping performance mechanical testing of IMFD Locking Screws - Bending Fatigue mechanical testing of IMFD Locking Screws - Static Four-Point Bend Test of IMFD - Bending Fatigue mechanical testing of IMFD - Fatigue testing of IMFD Assembly with corrosion assessment Clinical data are not needed to support the safety and effectiveness of the subject device. Based on the same intended use, materials, similar technological characteristics, and supportive performance testing, the Estremo Fibular Nail system has been established as substantially equivalent to the predicate device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...