Airlock® Ankle Plating System

K260274 · Novastep SAS · HRS · Mar 25, 2026 · Orthopedic

Device Facts

Record IDK260274
Device NameAirlock® Ankle Plating System
ApplicantNovastep SAS
Product CodeHRS · Orthopedic
Decision DateMar 25, 2026
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 888.3030
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Airlock® Ankle Plating System intended for the fixation of arthrodesis of the ankle.

Device Story

Airlock® Ankle Plating System consists of titanium alloy plates and screws for bone fixation; used for ankle arthrodesis; includes three plate variants with anatomical left/right options; features polyaxial locking and non-locking screws; includes visual indicators on plates to assist positioning and tibio-talar screw insertion; provided sterile for single-use in hospital/clinic settings; implanted by surgeons to stabilize bone segments during fusion; benefits patients by providing rigid fixation for joint arthrodesis.

Clinical Evidence

No clinical data. Substantial equivalence based on non-clinical performance testing including static four-point bending (ASTM F382-24), torsional strength, driving torque, and axial pullout strength (ASTM F543-23), sterilization validation (ISO 11137, ISO 11607), and biocompatibility (ISO 10993-1:2018).

Technological Characteristics

Titanium alloy (TA6V, ASTM F136); polyaxial locking and non-locking screws; anatomical plate design; sterile, single-use; mechanical fixation via bone screws.

Indications for Use

Indicated for arthrodesis of the ankle including tibiotalocalcaneal and tibiotalar joints, involving osteotomies of the distal tibia, talus, and calcaneus in adult patients.

Regulatory Classification

Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION March 25, 2026 Novastep SAS Manon Dupleichs RA manager 2, Allée Jacques Frimot Rennes, 35000 France Re: K260274 Trade/Device Name: Airlock® Ankle Plating System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HRS, HWC Dated: January 29, 2026 Received: January 29, 2026 Dear Manon Dupleichs: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K260274 - Manon Dupleichs Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K260274 - Manon Dupleichs Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, CHRISTOPHER FERREIRA -S Christopher Ferreira, M.S. Assistant Director DHT6C: Division of Restorative, Repair, and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} | Indications for Use | | | | --- | --- | --- | | Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K260274 | ? | | Please provide the device trade name(s). | | ? | | Airlock® Ankle Plating System | | | | Please provide your Indications for Use below. | | ? | | Airlock® Ankle Plating System is indicated for arthrodesis of the ankle including tibiotalocalcaneal and tibiotalar joints, involving osteotomies of the distal tibia, talus, and calcaneus. The system can be used in adult patients. | | | | Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? | {4} novastep 510(k) summary K260274 # 510(k) summary In accordance with the Food and Drug Administration Rule to implement provisions of the Safe Medical Devices Act of 1990 and in conformance with 21 CFR 807.92, this information serves as a Summary of Safety and Effectiveness for the use of the Airlock® Ankle Plating System. # SUBMITTER/510(K) HOLDER Novastep® S.A.S. 2, Allée Jacques Frimot 35000 Rennes France Phone: +33-2-99-33-86-50 Contact: Manon Dupleichs, Regulatory Affairs Manager Contact Phone: +33-2-99-33-86-50 Contact Email: manon.dupleichs@enovis.com Date Prepared: January 29, 2026 # DEVICE NAME Device name: Airlock® Ankle Plating System Device Classification Name: Plate, Fixation, Bone Regulation Number: 888.3030 Classification Product Code: HRS Subsequent Product Codes: HWC Device Class: Class II # PREDICATE DEVICE (PRIMARY) Manufacturer: Wright medical technology inc Device name: ORTHOLOC 3DI ANKLE FUSION PLATING SYSTEM Device Classification Name: Plate, Fixation, Bone 510(k) Number: K121425 Regulation Number: 888.3030 Classification Product Code: HRS Subsequent Product Codes: HWC Device Class: Class II Page 2 / 4 {5} novastep 510(k) summary K260274 # PREDICATE DEVICE (ADDITIONAL) Manufacturer: Paragon 28 (Zimmer Biomet Company) Device name: Silverback™ TT/TTC Plating System Device Classification Name: Plate, Fixation, Bone 510(k) Number: K250952 Regulation Number: 888.3030 Classification Product Code: HRS Subsequent Product Codes: HTN, HWC Device Class: Class II # PREDICATE DEVICE (ADDITIONAL) Manufacturer: Novastep S.A.S. Device name: Airlock® osteosynthesis plate systems Device Classification Name: Plate, Fixation, Bone 510(k) Number: K143523 Regulation Number: 888.3030 Classification Product Code: HRS Device Class: Class II # DEVICE DESCRIPTION Airlock® Ankle Plating System is a range of implantable devices, made of titanium alloy, designed to treat ankle-related indications. Each implant is provided as a single-use sterile device. The Airlock® Ankle Plating system range is constituted by 3 plates variants with holes, available in left and right option for anatomical adaptation. Screws for the fixation and the stabilization of the airlock ankle plates are available in 4 different variants (Locking polyaxial screws Ø4 or Ø4.7 and non-locking screws Ø4 or Ø4.7) and are available in lengths from 14 to 60mm. # INTENDED USE/INDICATION FOR USE The intended use is: - Airlock® Ankle Plating System intended for the fixation of arthrodesis of the ankle. The indication for use is: Airlock® Ankle Plating System is indicated for arthrodesis of the ankle including tibiotalocalcaneal and tibiotalar joints, involving osteotomies of the distal tibia, talus, and calcaneus. The system can be used in adult patients. # SUMMARY OF TECHNOLOGICAL CHARACTERISTICS COMPARED TO THE PREDICATES DEVICES Airlock® Ankle Plating System is substantially equivalent to the predicate devices, ORTHOLOC 3DI ANKLE FUSION PLATING SYSTEM (primary predicate, K121425), Silverback™ TT/TTC Plating System (additional predicate, K250952) and Airlock® osteosynthesis plate systems (additional predicate, K143523). The subject device has similar indications for use, and technological characteristics as the predicate devices. Page 3 / 4 {6} novastep 510(k) summary K260274 The subject device features the same principles of operation for bone fixation. It is manufactured from the same TA6V alloy (ASTM F136) and shares similar features such as polyaxial locking, anatomical design and plate positioning. The subject device features short talar neck for patient anatomical adaptation, visual indications on the plate to facilitate positioning on the patient and the insertion of the dedicated tibio-talar screw. Each single device is provided single-use sterile pack and ready to use for the hospitals/clinics. ## SUMMARY OF NON-CLINICAL PERFORMANCE TESTING AS BASIS FOR SUBSTANTIAL EQUIVALENCE Performance tests have been done according to the FDA guidance's "Orthopedic Fracture Fixation Plates – Performance Criteria for Safety and Performance Based Pathway" (April 11, 2022) and "Orthopedic Non-Spinal Metallic Bone Screws and Washers – Performance Criteria for Safety and Performance Based Pathway" (November 22, 2024): - Static Four-Point Bending on plates (per ASTM F382-24), - Torsional Strength, Driving Torque and Axial Pullout Strength on screws (per ASTM F543-23), - Sterilization validation according to IDO 11137-1, 2 and 3 and ISO 11607-1 and 2, - Biocompatibility of the device assessed using ISO 10993-1:2018. ## SUMMARY OF CLINICAL TESTING AS BASIS FOR SUBSTANTIAL EQUIVALENCE Not applicable. Clinical studies were not required for this submission. ## CONCLUSIONS Airlock® Ankle Plating System is substantially equivalent to the predicate devices, ORTHOLOC 3DI ANKLE FUSION PLATING SYSTEM (primary predicate, K121425) and Silverback™ TT/TTC Plating System (additional predicate, K250952) and Airlock® osteosynthesis plate systems (additional predicate, K143523). Airlock® Ankle Plating System is has also demonstrated its safety and performance according to the FDA guidance's "Orthopedic Fracture Fixation Plates – Performance Criteria for Safety and Performance Based Pathway" and "Orthopedic Non-Spinal Metallic Bone Screws and Washers – Performance Criteria for Safety and Performance Based Pathway". Page 4 / 4
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...