Clavicle Fixation System

K254288 · Skeletal Dynamics, Inc. · HRS · Mar 23, 2026 · Orthopedic

Device Facts

Record IDK254288
Device NameClavicle Fixation System
ApplicantSkeletal Dynamics, Inc.
Product CodeHRS · Orthopedic
Decision DateMar 23, 2026
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 888.3030
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Skeletal Dynamics Clavicle Fixation System is intended for fractures, osteotomies, and non-unions of the Clavicle including osteopenic bone.

Device Story

The Clavicle Fixation System is a metallic bone fixation system used by orthopedic surgeons in a clinical/surgical setting to stabilize clavicle fractures, osteotomies, and non-unions. The system consists of titanium bone plates and titanium or cobalt-chrome bone screws of varying diameters (2.5mm, 2.7mm, 3.5mm). It also includes pegs that interface with targeting guide instruments to facilitate orthogonal screw placement. The device is provided non-sterile and requires sterilization at the user's facility. By providing rigid internal fixation, the system stabilizes bone segments to promote healing and restore structural integrity, benefiting patients by facilitating recovery from clavicle injuries.

Clinical Evidence

Bench testing only. Static and dynamic plate testing conducted per ASTM F382-24. Screw testing (driving torque, torsional yield strength, axial pullout strength) conducted per ASTM F543-23. Additional non-standard mechanical testing performed on the peg and screw interface to assess torsional strength and axial pullout force. Results confirm the construct performs as intended and is consistent with the predicate device.

Technological Characteristics

Materials: Titanium (Ti 6Al-4V ELI) plates and screws; Cobalt-Chrome (CoCrMo) screws. Standards: ASTM F382-24 (plates), ASTM F543-23 (screws). Components: Bone plates, screws (2.5mm, 2.7mm, 3.5mm), pegs, and targeting guide instruments. Sterilization: Non-sterile, user-facility sterilized.

Indications for Use

Indicated for patients with clavicle fractures, osteotomies, and non-unions, including those with osteopenic bone.

Regulatory Classification

Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION March 23, 2026 Skeletal Dynamics, Inc. Alexandra Rodriguez Rojas Regulatory Affairs Manager 7300 N. Kendall Dr. Suite 800 Miami, Florida 33156 Re: K254288 Trade/Device Name: Clavicle Fixation System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HRS Dated: December 30, 2025 Received: December 31, 2025 Dear Alexandra Rodriguez Rojas: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K254288 - Alexandra Rodriguez Rojas Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See {2} K254288 - Alexandra Rodriguez Rojas Page 3 the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Thomas Mcnamara -S For: Christopher Ferreira, M.S. Assistant Director DHT6C: Division of Restorative, Repair, and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} | Indications for Use | | | | --- | --- | --- | | Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K254288 | ? | | Please provide the device trade name(s). | | ? | | Clavicle Fixation System | | | | Please provide your Indications for Use below. | | ? | | The Skeletal Dynamics Clavicle Fixation System is indicated for fractures, osteotomies, and non-unions of the Clavicle including osteopenic bone. | | | | Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? | {4} skeletal dynamics K254288 510(k) SUMMARY Skeletal Dynamic's Clavicle Fixation System ## Submitter Skeletal Dynamics, Inc. 7300 N. Kendall Drive, Suite 800 Miami, FL 33156 Phone: 305-596-7585 E-mail: arodriguez@skeletaldynamics.com Contact Person: Alexandra Rodriguez Rojas Date Prepared: December 30th, 2025 ## Name and Classification Trade Name: Clavicle Fixation System Common Name: Plate, Fixation, Bone. Classification Name: Single/multiple component metallic bone fixation appliances and Accessories. Classification Number: 21 CFR §888.3030 Regulatory Class: Class II Product Code: HRS ## Predicate Devices - A.L.P.S Clavicle Plating System (K173767) ## Reference Devices - Proximal Humerus Fixation System (K242436) - Protean Fragment Plating System (K241815) - Geminus Volar Plating System (K182492) ## Device Description The Clavicle Fixation System features titanium (Ti 6Al-4V ELI) bone plates, offered in various sizes for the repair of the superior midshaft and superior lateral regions, as well as the anterior midshaft and anterior lateral areas. Included in the Clavicle Fixation system are titanium (Ti 6Al-4V ELI) and Cobalt-Chrome (CoCrMo) bone screws offered in 2.5mm, 2.7mm and 3.5mm diameters. Each plate is designed to accept specific compatible screw sizes, depending on the plate configuration. In addition, the system offers pegs that accept orthogonal screws, K-wires, and specialized instrumentation. The system is provided non-sterile and is sterilized at the user's facility. ## Indications for Use The Clavicle Fixation System is intended for fractures, osteotomies, and non-unions of the Clavicle including osteopenic bone. Page 1 of 2 {5} skeletal dynamics # Substantial Equivalence The Clavicle Fixation System is substantially equivalent to the A.L.P.S. Clavicle Plating System (K173767). Minor differences in the technological characteristics and indications for use do not present any concerns related to safety or effectiveness of the device. # Comparison of Technological Characteristics with the Predicate Device The subject Clavicle Fixation System is substantially equivalent to the predicate A.L.P.S. Clavicle Plating System. Differences between the subject device and the predicate device include variations in plate geometry and screw design. In addition, the subject device incorporates pegs used in conjunction with targeting guide instruments to facilitate orthogonal screw placement. These technological differences do not raise new questions of safety or effectiveness, as demonstrated by mechanical performance testing and supportive justification showing that the subject device meets or exceeds the performance of the predicate device, and that screw performance testing is consistent with applicable FDA guidance. Therefore, the subject device is substantially equivalent to the predicate device. # Performance Testing Mechanical testing demonstrated that the Skeletal Dynamics Clavicle Fixation System is equivalent to the predicate devices currently marketed. Static and dynamic testing of the plates was conducted in accordance with ASTM F382-24, and screw testing, including driving torque, torsional yield strength, and axial pullout strength, was conducted in accordance with ASTM F543-23. Additional non-standard testing was performed to evaluate the mechanical performance of the peg and screw interface. This test included torsional strength and axial pullout force assessments to characterize the pegs' ability to maintain fixation under rotational and axial loading conditions. Results demonstrated that the peg/screw construct provides mechanical performance consistent with the intended function of the system and does not represent a worst-case scenario relative to the overall construct. Based on these results, the subject device performs as intended and is as safe and effective as the legally marketed predicate device. # Conclusions The Skeletal Dynamics' Clavicle Fixation System is substantially equivalent to the predicate device identified in this premarket notification. Page 2 of 2
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...