# EVOLVE EPS SYSTEM (K081373)

_Wrightmedicaltechnologyinc · HRS · Jul 3, 2008 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS/K081373

## Device Facts

- **Applicant:** Wrightmedicaltechnologyinc
- **Product Code:** [HRS](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS.md)
- **Decision Date:** Jul 3, 2008
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The EVOLVE® EPS System is intended for fixation of fractures, osteotomics and nonunions of the clavicle, scapula, olecranon, humerus, radius, ulna, pelvis, distal tibia and fibula.

## Device Story

EVOLVE® EPS System is a bone plate fixation system consisting of pre-contoured stainless steel plates and locking/non-locking screws. Used by orthopedic surgeons in clinical settings to stabilize bone fractures, osteotomies, and nonunions. Plates feature compression slots and locking screw holes; screws available in two diameters and eighteen lengths. Device provides mechanical stabilization to facilitate bone healing. No software or electronic components.

## Clinical Evidence

No clinical data provided; substantial equivalence based on design and material characteristics.

## Technological Characteristics

Materials: Stainless steel. Components: Pre-contoured plates with compression slots and locking screw holes; locking and non-locking screws (2 diameters, 18 lengths). Mechanical fixation device.

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K081373 page 42

JUL 30 2008

## 510(K) SUMMARY OF SAFETY AMD EFFECTIVENESS

In accordance with the Food and Drug Administration Rule to implement provisions of the S Medical Devices Act of 1990 and in conformance with 21 CRF 807, this information serves : Summary of Safety and Effectiveness for the use of the EVOLVE® EPS System.

| Submitted By:                       | Wright Medical Technology, Inc.                   |
|-------------------------------------|---------------------------------------------------|
| Date:                               | April 23, 2008                                    |
| Contact Person:                     | Theresa Leister                                   |
|                                     | Senior Regulatory Affairs Specialist              |
| Proprietary Name:                   | EVOLVE® EPS System                                |
| Common Name:                        | Bone Plate System                                 |
| Classification Name and Reference:  | 21 CFR 888.3030 Plate, Fixation, Bone -- Class II |
| Device Product Code and Panel Code: | Orthopedics/87/HRS                                |

#### DEVICE INFORMATION

#### A. INTENDED USE

The EVOLVE® EPS System is intended for fixation of fractures, osteotomics and nonunions of the clavicle, scapula, olecranon, humerus, radius, ulna, pelvis, distal tibia and fibula.

### B. DEVICE DESCRIPTION

The design features of the EVOLVE® EPS System are described below.

- Consists of a variety of pre-contoured plate geometrics .
- Plates feature compression slots and locking screw holes ●
- Manufactured from Stainless Steel ●
- Screws are available in both locking ar d non-locking designs .
- Screws are available in 2 diameters and. 18 lengths .

The design features of the EVOLVE® EPS System are substantially equivalent to the desig features of other devices previously cleared for market.

{1}------------------------------------------------

KO8 |373

# C. SUBSTANTIAL EQUIVALENCE INFORMATION

The design features, material, and indications for use of the EVOLVE® EPS System The design features, material, and markeasted predicate devices. The safety substantaly cqurratent to prevents System is adequately supported by the substan - l equivalence information, materials information, and analysis data provided within Premarket Notification.

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal features a stylized eagle-like symbol with three curved lines representing wings or feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

'JUL 30 2008

Wright Medical Technology, Inc. % Ms. Theresa Leister 5677 Airline Road Arlington, TN 38002

Re: K081373 Trade/Device Name: Evolve EPS System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: HRS Dated: May 14, 2008 Received: May 16, 2008

Dear Ms. Leister:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 - Ms. Theresa Leister

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

Mark M Millman

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

K081373

## Indications for Use

510(k) Number (if known): K081373

Device Name: EVOLVE® EPS System

Indications For Use:

The EVOLVE® EPS System is intended for fixation of fractures, osteotomies and nonunions of the clavicle, scapula, olecranon, humerus, radius, ulna, pelvis, distal tibia and fibula.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Mark A. Milken

(Division Sign-Off) Division of General, Restorative, and Neurological Devices

510(k) Number K08 1373

1 of 1

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS/K081373](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS/K081373)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).