QGQ · Bone Indentation Device

Orthopedic · 21 CFR 888.1600 · Class 2

Overview

Product CodeQGQ
Device NameBone Indentation Device
Regulation21 CFR 888.1600
Device ClassClass 2
Review PanelOrthopedic

Identification

A bone indentation device is a device that measures resistance to indentation in bone.

Classification Rationale

Class II (special controls). The special controls for this device are:

Special Controls

In combination with the general controls of the FD&C Act, the bone indentation device is subject to the following special controls: - (1) In vivo performance testing must demonstrate that the device performs as intended under anticipated conditions of use. Testing must evaluate the risk of bone fracture, soft tissue damage, pain, discomfort, bruising, or bleeding. - (2) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use, including an evaluation of the accuracy and precision of the device with respect to resistance to bone indentation. - (3) Human factors testing must demonstrate that the intended user(s) can correctly use the device, based on the instructions for use. - The patient-contacting components of the device must be demonstrated to be (4) biocompatible. - (ર) Performance testing must demonstrate: - The sterility of the patient-contacting components of the device; and (i) - (ii) Validation of reprocessing instructions for any reusable components of the device. - Performance data must support the shelf life of the device by demonstrating continued (6) sterility and device functionality over the identified shelf life. - Software verification, validation, and hazard analysis must be performed. (7) - (8) Performance data must be provided to demonstrate the electromagnetic compatibility (EMC) and electrical safety of the device. - (9) Labeling must include: - (i) Instructions for use: - (ii) Validated methods and instructions for reprocessing of any reusable components; - (iii) A shelf life for any sterile components; - (iv) Information regarding limitations of the clinical significance of the device output; and - A detailed summary of the accuracy and precision of the device. (v)

*Classification.* Class II (special controls). The special controls for this device are:(1) In vivo performance testing must demonstrate that the device performs as intended under anticipated conditions of use. Testing must evaluate the risk of bone fracture, soft tissue damage, pain, discomfort, bruising, or bleeding. (2) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use, including an evaluation of the accuracy and precision of the device with respect to resistance to bone indentation. (3) Human factors testing must demonstrate that the intended user(s) can correctly use the device, based on the instructions for use. (4) The patient-contacting components of the device must be demonstrated to be biocompatible. (5) Performance testing must demonstrate: (i) The sterility of the patient-contacting components of the device; and (ii) Validation of reprocessing instructions for any reusable components of the device. (6) Performance data must support the shelf life of the device by demonstrating continued sterility and device functionality over the identified shelf life. (7) Software verification, validation, and hazard analysis must be performed. (8) Performance data must be provided to demonstrate the electromagnetic compatibility (EMC) and electrical safety of the device. (9) Labeling must include: (i) Instructions for use; (ii) Validated methods and instructions for reprocessing of any reusable components; (iii) A shelf life for any sterile components; (iv) Information regarding limitations of the clinical significance of the device output; and (v) A detailed summary of the accuracy and precision of the device.

Recent Cleared Devices (3 of 3)

RecordDevice NameApplicantDecision DateDecision
K250216OsteoProbe (OP-100)Active Life Scientific, Inc.Sep 11, 2025SESE
K221195OsteoProbeActive Life Scientific, Inc.Jun 28, 2022SESE
DEN210013OsteoProbeActive Life Scientific, Inc.Aug 19, 2021DENG

Top Applicants

Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...